The largest database of trusted experimental protocols

Acquity 1 class uplc system

Manufactured by Thermo Fisher Scientific

The Acquity I-Class UPLC System is a high-performance liquid chromatography (HPLC) instrument designed for analytical separations. It utilizes ultrahigh-pressure liquid chromatography (UPLC) technology to deliver rapid, high-resolution separations with enhanced sensitivity. The system is capable of operating at pressures up to 15,000 psi (1,034 bar) and features advanced fluidics and detector technologies to enable efficient, accurate, and reproducible chromatographic analysis.

Automatically generated - may contain errors

6 protocols using acquity 1 class uplc system

1

Metabolic Profiling of LGG Planktonic Cultures

Check if the same lab product or an alternative is used in the 5 most similar protocols
LGG planktonic cultures grown in 50% TSB, TSB supplemented with (1% w/v) D-(+) - glucose, (1% w/v) D-(+)- raffinose or (1% w/v) D- (+)- mannose for 24 h at 37 °C. To get an equal number of bacteria the OD600 of the bacteria was compared, and after that washed twice with PBS. For polar metabolite analysis in the polar phase samples, the lyophilized pellets were dissolved using 100 µL DDW-methanol (1:1), centrifuged twice (at maximum speed) to remove possible precipitants, and were injected into LC-MS system. Polar analysis in the polar phase was done as following: Analysis was performed using Waters Acquity I class UPLC System combined with a mass spectrometer (Thermo Exactive Plus Orbitrap) operated in a negative ionization mode. The LC separation was done using the SeQuant Zic-pHilic (150 mm × 2.1 mm) with the SeQuant guard column (20 mm × 2.1 mm) (Merck). The Mobile phase B: acetonitrile and Mobile phase A: 20 mM ammonium carbonate with 0.1% ammonia hydroxide in water: acetonitrile (80:20, v/v). The flow rate was kept at 200 μL min−1 and gradient as follows: 0–2 min 75% of B, 17 min 12.5% of B, 17.1 min 25% of B, 19 min 25% of B, 19.1 min 75% of B, 23 min 75% of B.
+ Open protocol
+ Expand
2

Intact Mass Analysis of Antibody Chains

Check if the same lab product or an alternative is used in the 5 most similar protocols
For analysis
of the molecular weights of the light and heavy chains of IRCR201
and cIRCR201, intact mass analysis was performed. To reduce the sample,
TCEP-HCl (50 mM) was added to antibodies followed by incubation for
45 min at 56 °C. The reduced sample was immediately loaded onto
a liquid chromatography–mass spectrometry (LC–MS) column
at a flow rate of 0.2 mL/min. Separation was accomplished using a
Waters ACQUITY I class UPLC system (Milford) with a Thermo MabPac
RP column (2.1 mm × 50 mm, 4 μm particle size) at 65 °C.
Buffers used included eluent A, consisting of 0.1% formic acid in
water, and eluent B, consisting of 0.1% formic acid in 100% acetonitrile.
The gradient was fixed with 25% eluent B for 2 min. A linear gradient
was applied with 25 to 45% eluent B for 8 min. The effluent was analyzed
with a Thermo LTQ Orbitrap system.
+ Open protocol
+ Expand
3

Metabolic Profiling of Polar Phase

Check if the same lab product or an alternative is used in the 5 most similar protocols
Metabolic profiling of the polar phase was performed as described in Zheng et al. (2015) with minor modifications. Briefly, analysis was performed using the Acquity I class UPLC system combined with a mass spectrometer (Thermo Exactive Plus Orbitrap) which was operated in negative ionization mode. The LC separation was performed using SeQuant Zic-pHilic (150 mm×2.1 mm) with the SeQuant guard column (20 mm×2.1 mm) (Merck). Mobile phase A was acetonitrile and mobile phase B was 20 mM ammonium carbonate plus 0.1% ammonia hydroxide in water. The flow rate was kept at 200 μl min−1 and the gradient was: 0–2 min 75% B, 17min 12.5% B, 17.1min 25% B, 19min 25% B, 19.1min 75% B, 19min 75% B.
+ Open protocol
+ Expand
4

Lipid Profiling using UPLC-Orbitrap

Check if the same lab product or an alternative is used in the 5 most similar protocols
Lipid extracts were analyzed using a Waters ACQUITY I Class UPLC System coupled to a mass spectrometer (Thermo Exactive Plus Orbitrap) which was operated in switching positive and negative ionization mode. The analysis was performed using ACQUITY UPLC System combined with chromatographic conditions as described in [6 (link)].
+ Open protocol
+ Expand
5

Polar Phase Metabolic Profiling by UPLC-MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
Metabolic profiling of polar phase was done as previously described99 (link) with minor modifications described below. Briefly, analysis was performed using Acquity I class UPLC System combined with mass spectrometer Q Exactive Plus Orbitrap™ (Thermo Fisher) that was operated in a negative ionization mode. The LC separation was done using the SeQuant Zic-pHilic (150×2.1 mm) with the SeQuant guard column (20 × 2.1 mm; Merck). The composition of mobile phase B and A was acetonitrile, and 20 mM ammonium carbonate with 0.1% ammonium hydroxide in DDW:acetonitrile (80:20, v/v), respectively. The flow rate was kept at 200 μL/min and gradient as follows: 0–2 min 75% of B, 14 min 25% of B, 18 min 25% of B, 19 min 75% of B, for 4 min.
+ Open protocol
+ Expand
6

Metabolic Profiling of Polar Phase

Check if the same lab product or an alternative is used in the 5 most similar protocols
Metabolic profiling of the polar phase was described11 with minor modifications described below. Briefly, analysis was performed using Acquity I class UPLC System combined with mass spectrometer Q Exactive Plus Orbitrap™ (Thermo Fisher Scientific), operated in a negative ionization mode. The LC separation was done using the SeQuant Zic-pHilic (150 mm × 2.1 mm) with the SeQuant guard column (20 mm × 2.1 mm) (Merck). Mobile Phase B: acetonitrile and Mobile Phase A: 20 mM ammonium carbonate with 0.1% ammonia hydroxide in DDW: acetonitrile (80:20, v/v). The flow rate was kept at 200 μl/ min, and the gradient was as follows: 0–2 min 75% of B, 14 min 25% of B, 18 min 25% of B, 19 min 75% of B, for 4 min.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!