The largest database of trusted experimental protocols

12 protocols using zorbax nh2

1

Oligosaccharide and Fructose Analysis via HPLC

Check if the same lab product or an alternative is used in the 5 most similar protocols
The oligosaccharide and fructose analysis were performed on the Shimadzu LC-20A HPLC system coupled with RID. Agilent ZORBAX NH2 (5 μm, 4.6 mm × 250 mm, i.d.) and column was used. The column was maintained at 35 °C. Isocratic elution with 70% aqueous acetonitrile was used as mobile phase at a flow rate of 1 mL/min for 45 min.
+ Open protocol
+ Expand
2

Taste Comparison of RB and RA

Check if the same lab product or an alternative is used in the 5 most similar protocols

Example 7

Taste comparison between RB (prepared in Example 1, first part) and RA Purity of RA was 99.0% by HPLC and RB was 99.1% by HPLC.

HPLC Conditions

Instrument: Shimadzu SPD-20A

Mobile Phase: Acetonitrile-Water (Dissolve 25 mg ammonium acetate and 25 μL acetic acid into 200 mL water, filtered through 0.45 μm membrane)=80:20

Column: AgiLent Zorbax NH2 (5 μm, 4.6 mm×150 mm)

Flow: 1 mL/min

Temperature: Ambient

Wavelength: 210 nm

Sample Preparation: Weigh accurately 10 mg of sample into a 10 mL volumetric flask, add 5 mL mobile phase, stir till the solid dissolved, then add the mobile phase to volume.

Injection Volume: 10 μL

RB sample: 400 ppm aqueous solution

RA sample: 400 ppm aqueous solution

Result: 7/10 experts found that RB had a lesser sweetness than RA, but also had less bitterness than RA. RA has a strong bitterness profile.

Generally it was found that RB has a similar taste profile with RA, especially with regard to sweetness. Because of less bitterness, RB tasted better than RA.

+ Open protocol
+ Expand
3

Quantifying Nutrients in Bacillus Cultivation

Check if the same lab product or an alternative is used in the 5 most similar protocols
Residual content of the main nutrients (glycerol, total nitrogen, and total phosphorus) was determined using the biomass-free supernatants obtained by centrifugation of samples of Bacillus IP22 cultivation broth. Glycerol content was determined using the HPLC method. The HPLC system (Thermo Scientific Dionex UltiMate 3000 series; Thermo Fisher Scientific, Waltham, MA, USA) consists of the following components: pump HPG-3200SD/RS, autosampler WPS-3000(T)SL (10 μL injection loop), column Zorbax NH2 (250 mm × 4.6 mm, 5 μm; Agilent Technologies, Santa Clara, CA, USA), and the refractive index detector (ERC RefractoMax520, ERC GmbH, Riemerling, Germany). Mobile phase was 70% (v/v) acetonitrile. Analysis parameters were: mobile phase flow rate 1 mL/min, run time 15 min, column temperature 30 °C, detector temperature 45 °C, injection volume 10 μL.
Residual content of total nitrogen was determined by the Kjeldahl method [34 ], while residual content of total phosphorus was determined using the spectrophotometric method with ascorbic acid [35 (link)].
+ Open protocol
+ Expand
4

HPLC Analysis of 2-KGA and L-Sorbose

Check if the same lab product or an alternative is used in the 5 most similar protocols
The concentration of 2-KGA and L-sorbose in the fermentation broth were analyzed by the High-Performance Liquid Chromatography (HPLC) (Waters Corp., Massachusetts, USA) using an amino column (Agilent Zorbax NH2) with an ultraviolet detector at 210 nm. Cetonitrile-KH2PO4 (5%/95%, v/v) was used as mobile phase with a flow rate of 0.6 mL/min.
+ Open protocol
+ Expand
5

HPLC Analysis of Oligosaccharides and Residual Sugars

Check if the same lab product or an alternative is used in the 5 most similar protocols
The oligosaccharides were analyzed by HPLC using a Symmetry C18 column (4.6 mm× 150 mm) and a Waters 1525 Binary HPLC Pump (Waters, Milford, MA, USA). Oligosaccharides were detected by using a Waters 2414 refractive index detector. The products were identified in the chromatograms as described by Remaud-Simeon and co-workers [19 (link)].
Residual sugars (glucose, xylose, and XOSs in the fermentation broth after fermentation) were determined by HPLC using a Zorbax carbohydrate column (4.6 mm × 150 mm; Agilent, Santa Clara, CA, USA), analytical guard column Zorbax NH2 (4.6 mm× 12.5 mm), and a mobile phase of 75/25 acetonitrile/water. Breeze Chromatography Manager Software (Waters) (v2, Waters, Milford, MA, USA) was used for data treatment.
Proteins were assayed by the method of Lowry [20 (link)].
All the analyses for the determination of sugars and proteins were performed in duplicate from different experiments.
+ Open protocol
+ Expand
6

HPLC Analysis of Carbohydrates

Check if the same lab product or an alternative is used in the 5 most similar protocols
Carbohydrate analysis was carried out on a Shimadzu (Kyoto, Japan) HPLC system which consisted of an LC-20AD prominence solvent delivery module, a DGU-20A5R degassing unit, an SIL-10 AF automatic sample injector, and an RID-10A refractive index detector. The instrument was coupled with a computer equipped with LabSolutions Lite Version 5.52 software. The analytical column used for carbohydrate (fructose, glucose, sucrose, maltose, melezitose, raffinose and xylose) separation was an Agilent Technologies ZORBAX NH2 (Santa Clara, CA, USA) (4.6 × 250 mm, 5 μm particle size). The mobile phase consisted of HPLC-grade acetonitrile (J. T. Baker, Devnter, The Netherlands) and ultrapure water (70/30, v/v), while the operating conditions were: an injection volume of 10 µL, a mobile phase flow of 1 ml/min and a temperature of 30 °C. Carbohydrates were identified according to their retention times, and quantification was performed via external calibration carried out with carbohydrate standards suitable for HPLC analysis. Fructose, anhydrous glucose, sucrose, raffinose pentahydrate and melezitose hydrate were purchased from Sigma-Aldrich (St. Louis, MO, USA), while xylose and maltose monohydrate were purchased from Kemika (Zagreb, Croatia) [21 ].
+ Open protocol
+ Expand
7

HPLC Analysis of Sugar Concentrations

Check if the same lab product or an alternative is used in the 5 most similar protocols
Sugar concentrations in the solution were determined with high-performance liquid chromatography (HPLC). The concentration of each sample was adjusted using deionized water and filtered through a filter (0.45 μm) before an analysis. Conditions for the HPLC analyses were as follows: column, ZORBAX NH2 (5 µm, 4.6 mm × 150 mm; Agilent, Santa Clara, CA, USA); mobile phase acetonitrile/water (70:30 (v/v)); flow rate, 1.2 mL min−1; temperature, 30 ℃; and refractive index detector 5450 (Hitachi High-Tech Science, Tokyo, Japan).
Sugar concentration was determined from a standard curve plotted using the known concentrations of each sugar. Figure 1 shows a chromatogram of sugar compositions analyzed with HPLC.
+ Open protocol
+ Expand
8

Quantifying Metabolites via HPLC

Check if the same lab product or an alternative is used in the 5 most similar protocols
Glucose, ethanol, and glycerol were measured on an Agilent 1260 high‐performance liquid chromatography (HPLC). The analytical method for glucose and glycerol; column: ZORBAX NH2 (250 mm × 4.6 mm, 5 μm; Agilent); column temperature: 30°C; RID temperature: 35°C; flow rate: 1 mL/min; elution time: 30 min; mobile phase: 90% acetonitrile. The analytical method for alcohol: guard column: MetaCrab 87H (50 mm  × 4.6 mm; Agilent); column: MetaCrab 87H (300 mm × 7.8 mm; Agilent); column temperature: 35°C; RID temperature: 35°C; flow rate: 0.6 mL/min; elution time: 30 min; mobile phase: 0.01 N H2SO4 (0.29 mL concentrated sulfuric acid per liter).
+ Open protocol
+ Expand
9

Oligosaccharide Determination in Black Ginseng

Check if the same lab product or an alternative is used in the 5 most similar protocols
Referring to the oligosaccharide determination method [17 (link)] to determine the oligosaccharide content of black ginseng. An exact weight of 0.5 g of the sample powder was taken, dissolved to 50 mg/mL in purified water while vigorously shaking the mixture, and filtered through a 0.22 m microporous membrane to get the test solution. An Agilent ZORBAX NH2 (4.6 × 150 mm, 5 μm) column was used and maintained at 30 °C. Using acetonitrile (A) and water (B) as mobile phases, gradient elution was optimized as follows: 0 ~ 5 min, 12% ~ 17% (B); 5 ~ 10 min, 17% (B); 10 ~ 20 min, 17% ~ 28% (B); 20 ~ 30 min, 28% ~ 33% (B). The flow rate was 1 mL/min with an injection volume of 10 μL. The detector was ELSD, with a drift tube temperature of 40 °C; the carrier gas was nitrogen; and the evaporative light detector gain was 5 °C.
+ Open protocol
+ Expand
10

HPLC Quantification of Metformin in Blood

Check if the same lab product or an alternative is used in the 5 most similar protocols
HPLC analysis of metformin was conducted with an Agilent Technologies 1200 Series apparatus with a DAD detector (Agilent, Santa Clara, CA, USA). The analysis was performed with the chromatographic column Zorbax NH2 (4.6 × 150 mm, 5.0 μm) by Agilent Technologies. Agilent ChemStation program was used for monitoring the chromatographic system and data acquisition.
Metformin hydrochloride, certified reference material by Sigma-Aldrich (St. Louis, MO, USA) was used. Acetonitrile and methanol were of HPLC and LC-MS grade (Honeywell Riedel-de Haen, Offenbach, Germany).
HPLC analysis of metformin in blood samples was performed using a slightly modified method published by Mary Rebecca et al (13 ). The analysis was carried out at 30 °C on Zorbax-NH2 column. The mobile phase consisted of 100% acetonitrile with a flow rate of 0.725 mL/min. Detection was performed at 232 nm, and quantification was done with calibration curve method. A sample of 200 μL of blood plasma was mixed with 200 μL of acetonitrile. The mixture was vortexed and centrifuged for 10 min at 4 °C and 14 000 rpm. The supernatant was filtered through 0.45 μm syringe filter, and 20 μL was injected on Zorbax-NH2 column.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!