The largest database of trusted experimental protocols

Agilent 1290 series liquid chromatograph

Manufactured by Agilent Technologies
Sourced in United States, Germany

The Agilent 1290 series liquid chromatograph is a high-performance liquid chromatography (HPLC) system designed for analytical and preparative applications. It features a modular design that allows for the integration of various components, including pumps, autosamplers, and detectors, to create a customized analytical solution. The 1290 series provides advanced functionality, precise control, and reliable performance to support a wide range of separation and analysis needs.

Automatically generated - may contain errors

7 protocols using agilent 1290 series liquid chromatograph

1

Quantitative Analysis of UAMC-3203 by LC-MS/MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
Quantitative analysis of UAMC-3203 was performed with the LC-MS/MS system that consisted of an Agilent 1290 series liquid chromatograph connected to an Agilent 6495 triple-quadruple mass spectrometer (Agilent Technologies, Inc., USA). The mobile phase consisted of 0.1% formic acid (Fisher chemical, Geel, Belgium), 1 mM ammonium formate (Honeywell, Fluka, Seelze, Germany) in Milli-Q-H2O (eluent A), and acetonitrile (LC-MS Chromasolv, Honeywell, Riedel-de Haen, Seelze, Germany) (eluent B). Gradient elution was used: 0.0 to 3.0 min: 25%→95% B, 3.0 to 3.5 min: 95% B, 3.5 to 3.6 min: 95%→25% B, 3.6 to 5.0 min 25% B. Solvent flow was 0.3 mL/min, and Poroshell 120 SB-C18 column (2.1 mm × 50 mm, 2.7 µm, Agilent Technologies, Inc., USA) was maintained at 60 °C. The injection volume was 2 µL. Nitrogen was used as a drying, nebulizer, and collision gas. The following conditions were used in the electrospray ion source: positive ion mode, sheath gas flow rate 11 L/min at 350 °C, drying gas flow rate 16 L/min at 200 °C, nebulizer gas pressure adjusted to 25 psi, capillary voltage 3500 V (ESI+), fragmentor voltage 380 V, dwell time 150 ms and cell accelerator voltage 5 V. The data were analyzed with Agilent Mass Hunter Quantitative Analyzed software (vB.09.00, build 9.0.647.0, Agilent Technologies, CA, USA). The precursor and fragment ions are presented in Supplementary Table S1.
+ Open protocol
+ Expand
2

Quantitative Analysis of Active Metabolites

Check if the same lab product or an alternative is used in the 5 most similar protocols
Samples were analyzed using an Agilent® 1290 series liquid chromatograph (Agilent Technologies, Santa Clara, CA, USA), coupled with an ABSCIEX® API 5500 mass spectrometer (SCIEX, Framingham, MA, USA), and a SunfireTM C18 chromatographic column (Waters Corp., Milford, MA, USA) of 3.5 μm and 150 × 2.1 mm was used. The analytes were separated chromatographically using a mobile gradient of 0.1% formic acid in water (Phase A) and 0.1% formic acid in methanol (Phase B). The flow rate was adjusted to 0.2 mL/min, the injection volume was 20 µL, the duration was 25.423 min, and the column temperature was set at 35 °C ± 1 °C. The liquid chromatographic pump gradient is shown in Table 1. The scans per peak are shown in Table S1.
The criteria to identify the different AMs and their active metabolites was the monitoring of the masses of the precursor and fragment ions (Table S2). In addition, different parameters were used for the operation of the mass detector (Table 2). The chromatographic integration of the samples was performed using Analyst® software version 1.6.2 (SCIEX, Framingham, MA, USA).
Samples were processed and analyzed at the Veterinary Pharmacology Laboratory (FARMAVET, as per initials in Spanish) of the Faculty of Veterinary and Animal Sciences at the University of Chile, which is accredited under ISO 17,025 standards.
+ Open protocol
+ Expand
3

Multiplex Quantification of Antimicrobials

Check if the same lab product or an alternative is used in the 5 most similar protocols
Samples were analyzed using an Agilent 1290 series Liquid Chromatograph (Santa Clara, CA) coupled to a triple quadrupole tandem mass spectrometer, in multiple reaction monitoring mode via an electrospray interface. In particular, an AB Sciex API 5500 mass spectrometer (AB Sciex LLC, Framingham, MA) was used. This device through polarization switching was operated in positive ionization mode for the analytes TC, 4-epi-TC, OTC, 4-epi-OTC, CTC, 4-epi-CTC, TYL, EFX, CFX, FLU, SCP, SDZ, SMZ13C6, EFX-D5, ETM13C-D3, TC-D6 and in negative ionization mode for the FF and CAF-D5 analytes. The specifications of the mass spectrometer and the liquid chromatograph are described in Supplementary Table 1 while Supplementary Table 2 shows the specific mass spectrometer conditions for the analytes.
+ Open protocol
+ Expand
4

Extraction and Analysis of Air Filter Contaminants

Check if the same lab product or an alternative is used in the 5 most similar protocols
The extraction method was referred to the treatment of air lters with some modi cation. Cotton balls with dust were extracted in polypropylene tube by ultrasound equipment. Before extraction, 5 ng IS mixture was spiked. About 15 mL 0.1% ammonia methanol were added in the tube, and then extracted for 15 min. After being centrifuged at a speed of 3500 r/min for 10 min, the super cial liquid was collected. Cotton balls with dust were extracted twice. All extract was collected together, and then concentrated to less than 0.5 mL under a gentle stream of high purity nitrogen. Envi-carb cartridges and nylon member lters were used to remove the impurity. The nal volume of each sample was reduced to 200 µL.
The instrument analysis employed Agilent 1290 series liquid chromatograph interfaced with an Agilent 6460 triple quadrupole mass spectrometer (HPLC-MS/MS, Agilent Technologies, USA). The detailed information of instrument analysis could be found in previous studies (Fang et al., 2018 (link), Zhao et al., 2020) .
+ Open protocol
+ Expand
5

Multi-modal Materials Characterization

Check if the same lab product or an alternative is used in the 5 most similar protocols
Next generation sequencing was performed on Illumina platform. A scanning electron microscope (SEM) with a secondary electron detector operated in high vacuum mode at an accelerating voltage of 10–20 kV (FEI Quanta FEG 250, Thermo Fisher Scientific, Waltham, MA, USA) was used to examine the morphology of the samples. The elements were identified by energy dispersive spectroscopy (EDS) with ApolloX SDD spectrometer (Ametek, Berwin, PA, USA) at an accelerating voltage of 20 kV. A Nicolet iS50 FTIR spectrometer controlled by the OMNIC v9.2 software package (Thermo Fisher Scientific, Waltham, MA, USA) equipped with Specac Quest Total Reflection Diamond (ATR) accessory and a single reflection diamond was used to record infrared transmission spectra. Chromatographic analysis was performed using Agilent liquid chromatograph series 1290 (Agilent Technology, Waldbronn, Germany) consisting of binary pump G4220A, autosampler G4226A, thermostated column compartment G1316C, diode-array detector G1315C, and triple quadrupole mass spectrometer G6460 with AJS electrospray ionization source. The chromatographic system was controlled with Agilent MassHunter software B 06.01.
+ Open protocol
+ Expand
6

Multitechnique Characterization of Materials

Check if the same lab product or an alternative is used in the 5 most similar protocols
A scanning electron microscope (SEM) with a secondary electron detector operated in high vacuum mode at an accelerating voltage of 10–20 kV (FEI Quanta FEG 250, Thermo Fisher Scientific, Waltham, MA, USA) was used to examine the morphology of the samples. The elements were identified by energy dispersive spectroscopy (EDS) using an ApolloX SDD spectrometer (Ametek, Berwin, PA, USA) at an accelerating voltage of 20 kV. Chromatographic analysis was performed using an Agilent Liquid Chromatograph Series 1290 (Agilent Technology, Waldbronn, Germany) containing a high speed binary pump (G7120A), autosampler (G7167B), thermostated column compartment (G7116B), diode-array detector (G1315C), and triple quadrupole mass spectrometer with AJS electrospray ionization source (G6470B). The Agilent MassHunter software (B 06.01) controlled the chromatographic system.
Thermogravimetric analysis (TGA) in two different atmospheres was applied to monitor the thermal degradation process, which is related to mass loss as a function of rising temperature. Thermogravimetric measurements (TGA/DSC1 analyzer, Mettler Toledo, Greifensee, Switzerland) were taken at a linear heating rate of 10 °C min−1 over the temperature range from ambient temperature to 1100 °C, in an air or N2 flow.
+ Open protocol
+ Expand
7

Liquid Chromatography-Mass Spectrometry Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
Chromatographic analysis was performed using Agilent liquid chromatograph series 1290 (Agilent Technology, Waldbronn, Germany) consisting of binary pump G4220A, autosampler G4226A, thermostated column compartment G1316C, diode-array detector G1315C, and triple quadrupole mass spectrometer G6460 with AJS electrospray ionization source. The chromatographic system was controlled with Agilent MassHunter software B 06.01.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!