The largest database of trusted experimental protocols

Acquity uplc hss t3 1.8 μm column

Manufactured by Waters Corporation
Sourced in United States

The ACQUITY UPLC HSS T3 1.8 μm column is a high-performance liquid chromatography (HPLC) column designed for use in ultra-high-performance liquid chromatography (UPLC) systems. The column features a particle size of 1.8 micrometers and a C18 stationary phase, which provides high-resolution separation of a wide range of analytes.

Automatically generated - may contain errors

7 protocols using acquity uplc hss t3 1.8 μm column

1

Targeted LC-MS/MS Metabolite Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
Xevo TQ-S (Waters, USA) was utilized for the analysis of all biological samples. Chromatographic separation was conducted on Waters ACQUITY UPLC HSS T3 1.8 μm column (2.1 mm × 100 mm). Column temperature was 45°C and the flow rate was 0.5ml/min. Room temperature was 20 ± 2°C. Phase A consisted of water and 0.1% formic acid (v/v), and phase B consisted of acetonitrile (ACN) 0.1% formic acid (v/v). Column separation was performed by gradient elution: 0-0.5 min, 10% B; 1-2 min, 90% B; 4 min, 10% B.
Electrospray ionization (ESI) and positive ion V mode were used to determine the samples. The main parameters for mass spectrum were set as follows: capillary voltage: 3.5 kv; sample cone: 60 v; source offset: 50 v; source temperature: 150°C; desolvation temperature: 350°C; cone gas flow: 150 L/h; desolvation gas flow: 650 L/h; injection volume: 5 μL.
+ Open protocol
+ Expand
2

Ionization Profiling of Reference Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
An LC-FT/MS system with Waters binary management was configured by connecting an Acquity UPLC Waters system with an Acquity UPLC HSS T3 1.8 μmcolumn (Waters) to an FT/MS system (Bruker). To investigate ionization abilities of these reference compounds without labeling, the compounds were dissolved at a concentration of 2 mM in LC-MS grade water (Fluka), then diluted to yield 10, 100, and 500 pmol/μl samples, which were injected separately by direct infusion at a flow rate of 180 μl/min apart from this, ionization and detectability of dansyl-labeled standard compounds (5, 10, and 50 pmol/run) were evaluated by injecting standards onto the LC-FT/MS system. The LC gradient used for LC-FT/MS analysis of individual specimen was the same as that used for LC-HCT/MS.
+ Open protocol
+ Expand
3

Quantification of Floral Anthocyanins

Check if the same lab product or an alternative is used in the 5 most similar protocols
Tissue samples from the blotched and non-blotched parts of petals at each developmental stage or bud scales were used for anthocyanin analysis as previously described [6 (link)] with minor modifications. Specifically, all non-blotched or blotched parts of each flower were collected as a single biological replicate, and the same amount of tissue, by weight, was taken from each biological repetition to quantify anthocyanin concentration. From each sample, 0.2 g of fresh tissue was treated with 1 mL of 0.2% formic acid/methanol (v/v) solution for 20 minutes, ultrasonically homogenized, then incubated in the dark for 2 h. The resulting extract was centrifuged at 12,000 rpm for 5 min and the supernatant was collected. The above steps were repeated until all anthocyanins had been extracted. The supernatant was filtered through a 0.22 μm filter and stored at −20°C. Anthocyanin types and concentrations were determined using an UPLC-DAD (ACQUITY UPLC® I-Class, Waters, Massachusetts, USA). The analytical column was an ACQUITY UPLC® HSS T3 1.8 μm column (Waters, Massachusetts, USA). Four standards were purchased from Solarbio (Beijing, China), namely Cy3G, Cy3G5G, Pn3G, and Pn3G5G. Among them, Cy3G was used as a standard for quantifying anthocyanin concentration through linear regression.
+ Open protocol
+ Expand
4

UHPLC-QTOF-MS/MS for Metabolite Profiling

Check if the same lab product or an alternative is used in the 5 most similar protocols
Analysis was performed using a UHPLC (1290 Infinity LC, Agilent Technologies) coupled to a quadrupole time-of-flight (AB Sciex TripleTOF 6600) at Shanghai Applied Protein Technology Co., Ltd. To achieve HILIC separation, we used a 2.1 × 100 mm ACQUITY UPLC BEH 1.7 μm column (Waters, Ireland). The mobile phases for both ESI+ and ESI modes consisted of A, which was 25 mmol/L ammonium acetate and 25 mmol/L ammonium hydroxide in water, and B, which was acetonitrile. To obtain RPLC separation, we used a 2.1 × 100 mm ACQUITY UPLC HSS T3 1.8-μm column (Waters, Ireland). For ESI+ mode, the mobile phases included A, which was water with 0.1% formic acid, and B, which was acetonitrile with 0.1% formic acid. On the other hand, for ESI mode, the mobile phases comprised A, which was 0.5 mmol/L ammonium fluoride in water, and B, which was acetonitrile. During the MS-only acquisition, the instrument was set to capture data within the m/z range of 60–1,000 Da, with an accumulation time of 0.20 s/spectra for the TOF MS scan. Meanwhile, in auto MS-MS acquisition, the instrument captured data within the m/z range of 25–1,000 Da, with an accumulation time of 0.05 s/spectra for the product ion scan. To conduct the product ion scan, we used information-dependent acquisition, specifically selecting the high-sensitivity mode (11 (link)).
+ Open protocol
+ Expand
5

Kinetic Analysis of Pteroyl-di-L-glutamate Cleavage

Check if the same lab product or an alternative is used in the 5 most similar protocols
Kinetic parameters (KM and kcat) of pteroyl‐di‐l‐glutamate cleavage by Avi‐mGCPII, as well as Ki values for all inhibitors, were determined as previously described 39. Briefly, in a 96‐well plate, Avi‐mGCPII was mixed with 25 mm Bis‐Tris propane, 150 mm NaCl, 0.001% octaethylene glycol monododecyl ether (Affymetrix), pH 7.4 (and tested inhibitor, if used), into a final volume of 90 μL. Reactions were started by adding 10 μL of 4 μm pteroyl‐di‐l‐glutamate and incubated at 37 °C for 20 min. The reactions were stopped with 20 μL of 25 μm 2‐PMPA and subsequently analyzed on an Agilent 1200 Series system using an Acquity UPLC HSS T3 1.8 μm column (2.1 × 100 mm; Waters, Milford, MA, USA).
+ Open protocol
+ Expand
6

UPLC-MS/MS Quantification of Compound X

Check if the same lab product or an alternative is used in the 5 most similar protocols
Chromatographic separation was performed by a gradient elution on the ACQUITY UPLC HSS T3 1.8 μm column (2.1 × 100 mm) with VanGuard pre‐column (Waters Corporation). The mobile phase solvents were 0.1% formic acid in water (A) and 0.1% formic acid in acetonitrile (B) at a flow rate 0.4 ml/min for 5 min. The gradient program mobile phase conditions were 85% of A and 15% of B for the first 2.5 min, then changed linearly to 10% of A and 90% of B from 2.5–4.0 min, then immediately back to 85% of A and 15% of B from 4.0–5.0 min to re‐equilibrate at the initial conditions. The column temperature was 40°C, the autosampler temperature was maintained at 10°C and the injection volume was 8 μl. The positive electrospray ionization mode (ESI [+]) was used with the multiple reactions monitoring (MRM). The tune page source voltages were 0.7 kV and 84 V for the capillary and cone, respectively. The source desolvation temperature was 500°C. The source desolvation gas flow was 800 L/h and the cone gas was 50 L/h. The MS file cone voltage setting was 74 V with collision energy setting of 48 V. Argon was used as the collision gas and nitrogen as the desolvation and cone gases. Quantification was performed using the transition Parent (m/z): 468.34 and Daughter (m/z): 83.77 with a retention time of 2.43 min.
+ Open protocol
+ Expand
7

Derivatization and Retention Optimization for INH

Check if the same lab product or an alternative is used in the 5 most similar protocols
To improve hydrophobicity of INH, we derivatizated INH with cinnamaldehyde. To improve the retention of highly polar compounds in the chromatography column, we added ion-pairing reagent (HFBA) to the sample diluted solution. After optimizing the method for sample preparation, we compared the elution solution (B) of acetonitrile (ACN) and methanol, and solution A of 0.1% FA water, 5 mM NH4FA, 5 mM NH4FA (pH 4.0), 10 mM NH4FA, and 10 mM NH4FA (pH 4.0). Two kinds of spiked plasma standards with concentration of upper limit of quantification (ULOQ) and lower limit of quantification (LLOQ) were used. Chromatographic separation was performed on an Acquity UPLC HSS T3 1.8 μm column (2.1 × 100 mm, Waters) at a flow rate of 0.35 mL/min with HPLC gradient of 0–0.5 min 2% B, 1.5–2.5 min 80% B, and 3.0–4.0 min 2% B. The column temperature was 35 °C. For each sample, 5.0 μL was injected with a total running time of 4.0 min per sample.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!