The largest database of trusted experimental protocols

Bds hypersil c18 hplc column

Manufactured by Thermo Fisher Scientific
Sourced in United States

The BDS Hypersil™ C18 HPLC column is a reversed-phase high-performance liquid chromatography (HPLC) column designed for the separation and analysis of a wide range of organic compounds. It features a bonded C18 stationary phase on a high-purity silica support.

Automatically generated - may contain errors

10 protocols using bds hypersil c18 hplc column

1

HPLC analysis of potassium phosphate buffer

Check if the same lab product or an alternative is used in the 5 most similar protocols
Potassium phosphate monobasic buffer (0.2 M) was prepared by dissolving 27.2 g of KH2PO4 in 800 mL water, adjusting the pH at 2.8 ± 0.05 with 85% H3PO4 and adding water to the total volume of 1000 mL. SOS solution (20 mM) was prepared by dissolving 4.7 g SOS in 1000 mL water. Mobile phases A and B, both consisting of ACN, 20 mM SOS, 0.2 M KH2PO4 buffer at pH 2.8 and water were prepared as specified in the Ph. Int. in the following ratios—mobile phase A: ACN/SOS/KH2PO4 buffer/water = 4/70/10/16 v/v/v/v; and mobile phase B: ACN/SOS/KH2PO4 buffer/water = 17/70/10/3 v/v/v/v.
A Hypersil BDS C18 HPLC column (250 × 4.6 mm; 5 µm) from Thermo Fischer Scientific (San Jose, CA, USA) was used for the analysis. Gradient elution was applied at 45 °C and a flow rate of 1 mL min−1: 0% B (0–16 min), 0%–100%B (16–18 min), 100% B (18–22 min), 100%–0%B (22–24 min), and 0% B (24–30 min). The detection wavelength was set at 219 nm and the injection volume was 50 µL.
+ Open protocol
+ Expand
2

Quantitative Analysis of 2,3-BDO and Intracellular Metabolites

Check if the same lab product or an alternative is used in the 5 most similar protocols
The extracellular 2,3-BDO titer was analyzed by GC-BID (Shimatsu, Japan) using the TG-WAXMS column (Shimatsu). The cells were cultured in 40 ml medium in the shaker flask for 11 days. After harvesting and centrifugation (6,000×g, 15 min), the supernatant was diluted 10-fold and analyzed using GC-BID. Meanwhile, 2,3-BDO standard (Sigma) was diluted to 200, 100, 50, 25, and 10 mg/L and analyzed similarly to establish the standard curve. The 2,3-BDO titers were determined based on the area of the signal and the standard curve.
For the analysis of intracellular metabolites F6P, AcCoA, OAA, and sucrose, the cells were cultured in 40 ml BG-11 medium for 11 days and 1 ml of the supernatant was withdrawn. After centrifugation, the cells were lysed using 100 μL TE buffer containing 2 mg/ml lysozyme in a 37 °C water bath for 20 min, followed by centrifugation (12,000×g, 5 min). Subsequently, 500 μL supernatant was analyzed by HPLC-MS (Shimadzu) using Hypersil™ BDS C18 HPLC Column (5 μm, 10 × 250 mm, Thermo Fisher) and 5% acetonitrile as the mobile phase. The standard F6P, AcCoA, OAA, and sucrose (Sigma) were diluted to 1,000, 500, 100, 10, and 1 mg/L and analyzed similarly to generate the standard curve.
+ Open protocol
+ Expand
3

UPLC-APCI-FT-MS Analysis of Metabolites

Check if the same lab product or an alternative is used in the 5 most similar protocols
LC-APCI-FT-MS was performed using a Hypersil BDS C18 HPLC column (100 x 2.1 mm, 3 μm; Thermo Fisher Scientific) mounted on an Accela UPLC system (Thermo Electron Corporation). The LC system was coupled to a LTQ FT Ultra (Thermo Electron Corporation) via an atmospheric pressure chemical ionization (APCI) system operated in positive mode. The following gradient was run using water/acetonitrile (99:1, v/v) (solvent A) and acetonitrile/water (99:1, v/v) (solvent B): time 0 min, 50% B; 19 min, 99.5% B; 25 min, 99.5% B. The injection volume was 10 μL, flow rate 400 μL/min, and column temperature 40°C. Positive ionization using the APCI source was obtained with the following parameter values: capillary temperature 200°C, APCI probe 350°C, sheath gas 40 (arbitrary units), aux. gas 5 (arbitrary units), spray voltage 6.0°kV, discharge current 5 μA, capillary voltage 4.0 V, and tube lens 70 V. Full MS spectra between m/z 250-1400 were recorded at a resolution of 100,000. Full MS spectra were interchanged with a dependent MS 2 scan event in which the most abundant ion in the previous full MS scan was fragmented, two dependent MS 3 scan events in which the two most abundant daughter ions were fragmented, and a dependent MS 4 scan event in which the most abundant granddaughter ion of the first MS 3 scan event was fragmented. The collision energy was set at 35%.
+ Open protocol
+ Expand
4

Quantifying Anthocyanins in Flower Petals

Check if the same lab product or an alternative is used in the 5 most similar protocols
For the determination of the anthocyanin content, 10 g of shock-frozen petals were pulverized and mixed with 35 ml 2 M methanolic hydrochloric acid. After shaking for 12 hours at 4°C in an overhead rotator, the suspension was centrifuged for 10 minutes at 19200×g. 10–140 µl of the supernatant was adjusted with 2 M methanolic hydrochloric acid to a final volume of 1000 µl. The absorption at 520 nm was determined on a DU-65 spectrophotometer (Beckman Instruments). The anthocyanin content was calculated as pelargonidin equivalent using a calibration curve obtained with commercially available pelargonidin chloride (Roth, Germany). For acidic hydrolysis of anthocyanins, 20 µl methanolic hydrochloric acid extract were mixed with 180 µl of 4 N HCL and incubated for 60 minutes at 90°C. After cooling for 10 minutes the mixture was centrifuged for 10 minutes at 10000×g. The supernatant was adjusted to 200 µl with 4 N HCL and aliquots were used for HPLC analysis [32] (link) using a Perkin Elmer Series 200 HPLC system equipped with a Perkin Elmer Series 200 diode array detector and Total Chrom Navigator, version 6.3.1 (Perkin Elmer Inc). The column was a BDS Hypersil C18 HPLC column, 5 µm, 250×4.6 mm (Thermo Scientific).
+ Open protocol
+ Expand
5

HPLC Separation of Long-Chain Fatty Acids

Check if the same lab product or an alternative is used in the 5 most similar protocols

Example 2

Sample injection was performed with an Agilent Technologies G1367B Autosampler.

The autosampler system automatically injected an aliquot of the above prepared reconstituted samples into a Thermo Scientific BDS Hypersil C18 HPLC column (3 μm particle size, 100×2.1 mm, from Thermo Scientific). An HPLC gradient was applied to the analytical column, to separate VLCFA and BCFA from other components in the sample. Mobile phase A was 20 mM ammonium acetate and mobile phase B was 82% acetonitrile in methanol. The HPLC gradient started with an 82% solvent B which was ramped to 90% in approximately 1 minute, then ramped up to 95% for another minute, and held at that percentage for approximately 36 seconds, before being ramped back down to 90% over the next one minute and 18 seconds, and then down to 82% over the next 24 seconds. Column flow rate during solvent application was about 0.85 mL/min. Pristanic acid, phytanic acid, docosanoic acid, tetracosanoic acid, and hexacosanoic acid were observed to elute off the column at approximately 1.43 minutes into the gradient profile.

+ Open protocol
+ Expand
6

HPLC Quantification of DPV Compound

Check if the same lab product or an alternative is used in the 5 most similar protocols
A Waters HPLC system (Elstree, UK) consisting of a 1525 binary HPLC pump, a 717 plus autosampler, an in-line degasser unit, a 1500 series column heater, a 2487 dual wavelength absorbance detector, and a 2998 photodiode array detector was used for all analyses. Release and solubility samples (10 µL), appropriately diluted, were injected onto a Thermo Scientific BDS Hypersil™ C18 HPLC column (150 mm × 4.6 mm, 3 µm particle size) maintained at 25 °C and fitted with a guard column. Isocratic elution was performed using a mobile phase comprising 60% HPLC-grade acetonitrile and 40% phosphate buffer (pH 3.0) and a total flow rate of 1.5 mL/min. The run time was 6 min and DPV was detected at 287 nm after ~2.85 min. DPV amounts were determined using a weighted calibration curve generated from three separate sets of dilutions performed independently in IPA/water (Supplementary Information, Fig. S1 and Table S1).
+ Open protocol
+ Expand
7

HPLC Analysis of Drug Release

Check if the same lab product or an alternative is used in the 5 most similar protocols
All in vitro release and content samples were analysed on a Waters HPLC system (Waters Limited, Elstree, UK) consisting of a 1525 Binary HPLC pump, a 717 Plus Autosampler and a 2487 Absorbance Detector. Samples (25 μL) were injected onto a Thermo Scientific BDS Hypersil™ C18 HPLC column (150 mm ⨉ 4.6 mm, 3 μm particle size) fitted with a guard column. The column was held at 25 °C and isocratic elution was performed using a mobile phase comprising 58% HPLC-grade acetonitrile and 42% pH 2.4 phosphate buffer (prepared using 7.7 mM pH 3.0 phosphate buffer modified by addition of 2.2 mL of 85% H3PO4 and 1 mL of 10 M NaOH), a flow rate of 1.2 mL/min and a run time of 5 min. DPV was detected at 210 nm after approximately 2.5 min and LNG at 240 nm after approximately 3.5 min (Dallal Bashi et al., 2019 (link)).
+ Open protocol
+ Expand
8

Quantification of DPV and LNG via HPLC

Check if the same lab product or an alternative is used in the 5 most similar protocols
Quantification of DPV and LNG was performed on a Waters HPLC system (Waters Limited, Elstree, UK) consisting of a 1525 Binary HPLC pump with an in-line degasser AF unit, column heater, 717 Plus Autosampler and a 2487 dual wavelength absorbance detector. Samples (25 μL) were injected onto a Thermo Scientific BDS Hypersil™ C18 HPLC column (150 mm × 4.6 mm, 3 μm particle size) fitted with a guard column. The column was held at 25 °C and isocratic elution was performed using a mobile phase of 58% HPLC-grade acetonitrile and 42% pH 2.4 phosphate buffer (prepared using 7.7 mM, pH 3 phosphate buffer modified by addition of 2.2 mL of 85% phosphoric acid and 1.0 mL of 10 M sodium hydroxide), a flow rate of 1.2 mL/min and a run time of 5 min. DPV was detected at 210 nm at ~2.7 min and LNG at 240 nm at ~3.7 min (Dallal Bashi et al., 2019 (link)).
+ Open protocol
+ Expand
9

HPLC Analysis of DPV Release

Check if the same lab product or an alternative is used in the 5 most similar protocols
A Waters HPLC system (Elstree, UK) consisting of; a 1525 binary HPLC pump, a 717 plus autosampler, an in-line degasser unit, a 1500 series column heater, a 2487 dual wavelength absorbance detector and a 2998 photodiode array detector was used for all analysis.
In vitro release 25 µL of sample was injected on to a Thermo Scientific BDS Hypersil C18 HPLC column (150 x 4.6 mm, 3 µm particle size) fitted with a guard column. The column was held at 45˚C and isocratic elution performed using a mobile phase of 45% acetonitrile and 55% phosphate buffer (pH 3.0, 7.7 mM) with a total flow rate of 1.2 mL/min and a run time of 8 min. DPV was detected at 240 nm after approximately 6.1 min.
+ Open protocol
+ Expand
10

HPLC Analysis of DPV Content and Release

Check if the same lab product or an alternative is used in the 5 most similar protocols
Samples for DPV content analysis in rings were analysed on a Waters HPLC system (Waters Corporation, Dublin, Ireland) consisting of a 1525 Binary HPLC pump with an inline degasser AF unit, 1500 column heater, 717 Plus Autosampler and a 2487 dual wavelength absorbance detector. 10 µL of each content sample was injected onto a Kromasil C18 HPLC column (150 mm x 4.6 mm, 5 µm particle size). Column temperature was maintained at 25 °C and isocratic elution was performed using a mobile phase of 75% HPLC-grade methanol and 25% water with a flow rate of 0.75 mL/min and a run time of 15 min. DPV was detected at 257 nm after approximately 10.8 min.
In vitro release samples (25 µL) were injected onto a Thermo Scientific BDS Hypersil™ C18 HPLC column (150 mm x 4.6 mm, 3 µm particle size) fitted with a guard column. The column was held at 45 °C and isocratic elution was performed using a mobile phase of 45% HPLC-grade acetonitrile and 55% phosphate buffer (pH 3.0; 7.7 mM) with a run time of 8 min. DPV was detected at 240 nm after 6.1 min.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!