The largest database of trusted experimental protocols

Synapt g2 si qtof instrument

Manufactured by Waters Corporation
Sourced in United Kingdom

The SYNAPT G2-Si QTOF instrument is a high-performance quadrupole time-of-flight mass spectrometer designed for advanced analytical applications. It provides accurate mass measurements and high-resolution data acquisition for a wide range of analytes. The instrument combines a quadrupole mass analyzer with a time-of-flight analyzer to deliver precise and sensitive mass analysis.

Automatically generated - may contain errors

6 protocols using synapt g2 si qtof instrument

1

Linear TWIM Analysis of Biomolecules

Check if the same lab product or an alternative is used in the 5 most similar protocols
For experiments using linear TWIM, the method outlined
above was performed but with a SYNAPT G2-Si QTOF instrument (Waters
Corporation, Wilmslow). Identical UPLC equipment, buffers, samples,
and columns were used to maximize comparability. Further details regarding
instrument tuning and mobility parameters can be found in the Supporting Information.
+ Open protocol
+ Expand
2

Glycoproteins Analysis by HILIC-MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
An intact protein MS analysis was carried out following a HILIC-based separation on a glycoprotein column. The chromatographic separation of the glycoforms was performed using a Glycoprotein BEH Amide column (Waters (Milford, MA); 300 Å, 1.7 μm, 2.1 mm × 150 mm) on a Waters Acquity UPLC system. The separation of the bioconjugate was achieved with a gradient starting at 25% of 0.1% trifluoroacetic acid (TFA) in water (buffer A) and 75% of 0.1% TFA in in acetonitrile (buffer B) going to 55% buffer A and 45% buffer B in 30 min at a flow rate of 0.2 mL/min. Detection was performed using an UV detector at 215/280. The UPLC was directly connected to a Waters Synapt G2Si Q-TOF instrument for mass determination run in resolution mode. Masses were acquired in the positive ion mode between m/z 500 and 4500 with conditions adapted for intact protein analysis. Lock mass correction was applied. MS spectra were deconvoluted with the MassLynx MaxEnt 1 algorithm (Waters) to determine the average intact masses of the glycoproteins injected. For the deconvolution, spectra belonging to an individual peak were combined, smoothed and background subtracted and the deconvolution carried out individually for each of these peaks.
+ Open protocol
+ Expand
3

MALDI-MSI Metabolite Detection Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
MALDI-MSI was carried out using a Synapt G2-Si QToF instrument (Waters) fitted with a uMALDI ion source57 (link). The samples were coated with 9-AA (Merck Life Science) at 10 mg ml−1 in 80:20 ethanol:water by TM Sprayer (HTX Technologies) with a temperature of 65 °C, flow rate of 0.06 ml min−1, nozzle velocity of 1,200 mm s−1, four passes, 3 ml track spacing and CC pattern alternation. The instrument was calibrated with Red Phosphorus to an RMS mass accuracy of <1 ppm, and the instrument method was set as follows: pixel size of 45 × 45 µm, sensitivity mode, negative ion mode, m/z range of 50–1,200, laser repetition rate of 2,500 Hz and scan time of 0.5 s.
DESI and MALDI-MSI data obtained from APC and APC KRAS tumours were also comprehensively evaluated for potential ion signal corresponding to SAM, SAH, homocysteine, cystathionine and methionine. On-tissue ion signal of sufficient signal-to-noise or accurate mass error was not identified for these ions. Metabolite standards were readily detected by DESI (Supplementary Table 7) but the lack of detection of endogenous ions in tissues derived from genetically engineered mouse models suggests these are below the limit of detection for the systems used in this study.
+ Open protocol
+ Expand
4

ESI-MS Characterization of hIAPP

Check if the same lab product or an alternative is used in the 5 most similar protocols
ESI-MS experiments were carried out with a Synapt G2-Si Q-TOF instrument (Waters, Manchester, United Kingdom) equipped with an ESI interface. Positive ion mode was used. Samples of hIAPP (as previously described for the CE-UV experiments) were injected by direct infusion through a syringe at 5 μL/min. Main MS parameters are detailed the following: the capillary voltage was set at 2.4 kV, with the sampling cone and source offset at 70 V. Source and desolvation temperatures were, respectively, 40 and 75°C. Data were processed with MassLynx.
+ Open protocol
+ Expand
5

High-Resolution Mass Spectrometry Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
ESI-TOF HRMS experiments were performed using a SYNAPT G2-Si Q-TOF instrument hyphenated with the ACQUITY UPLC H-Class system (Waters, Manchester, UK). The mobile phase was composed of ammonium formate 50 mM and formic acid 0.1%, and the flow rate was set at 0.2 mL/min. One microliter of sample was injected, and the acquisition method run time was 2 min.
The ESI source was operating in positive ionization mode using a capillary voltage of 2.5 kV and the following conditions: cone voltage, 120 V; source offset, 20 V; source temperature, 120 °C; desolvation gas temperature, 450 °C; desolvation gas flow, 800 L/h, and cone gas flow, 50 L/h. Nitrogen (>99.5%) was employed as the desolvation gas. Mass calibration was carried out using a sodium formate solution, and lock mass correction was applied for accurate mass measurements using the [M−H] ions (m/z 554.2615) obtained from a Leu-enkephalin solution. The scan range was m/z 50–2500 at 0.25 s/scan. The TOF was operated in sensitivity mode, providing an average resolving power of 20,000 (FWHM). All spectra were recorded in continuum mode. Data acquisition was performed with MassLynx software (V4.1, Waters).
+ Open protocol
+ Expand
6

Characterizing Compounds via ESI-MS and ESI-HR-MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
Electrospray ionization mass spectra (ESI-MS) were recorded on a Thermo Finnigan LCQ Advantage spectrometer (Hemel Hempstead, Hertfordshire, UK). Mass spectra acquisition was performed in positive and negative ion mode in a mass range of 50–2000 m/z. Source parameters: T, 200 °C; sheath gas flow rate, 20 arb; aux gas flow rate, 10 arb; spray voltage, 5.50 kV; capillary temperature, 275 °C; capillary voltage, 38.00 V; tube lens, 65.00 V.
Electrospray ionization high-resolution mass spectrometry (ESI-HR-MS) analyses were performed on a Synapt G2-Si QTof instrument (Waters, Milford, MA, USA) equipped with a ZsprayTM ESI-probe for electrospray ionization (Waters, Milford, MA, USA). The sample was dissolved in water–0.1% formic acid in methanol and analyzed by direct infusion. The ion source and interface conditions were as follows: capillary voltage 1.3 kV, sampling cone 120 V, source temperature 120 °C, desolvation temperature 150 °C and desolvation gas flow rate 600 L h−1. Mass spectra were acquired over the m/z range 50–1000 in a positive ion Full Scan mode. Leucine enkephalin was used as a lock-mass compound. Data were processed with MassLynxTM V4.2 software (Waters Milford, MA, USA).
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!