The largest database of trusted experimental protocols

8 protocols using series hp 1100

1

HPLC Analysis for Oleuropein Quantification

Check if the same lab product or an alternative is used in the 5 most similar protocols
To determine the OLE composition, high-performance liquid chromatography
(HPLC) analysis was performed using a Hewlett-Packard Series HP 1100
equipped with a diode array detector and a C18 LiChrospher 100 analytical
column (250 × 4 mm). The flow rate was adjusted to 1 mL min–1, and the change of absorbance was displayed at 280
nm wavelength. The mobile phases were used as 2.5/97.5 (v v–1) of acetic acid/water (A) and acetonitrile (B). A linear gradient
was run for 60 min in total, from 95% A–5% B to 75% A–25%
B for 20 min, and then it was changed into 50% A–50% B for
20 min and 20% A–80% B for 10 min, respectively. At last, re-equilibration
was performed for 10 min to have the initial composition.
The
oleuropein level in OLE was also determined by HPLC analysis using
an oleuropein calibration curve (Figure S3). As an internal standard, coumarin was utilized for the measurement
of antioxidant oleuropein.
+ Open protocol
+ Expand
2

HPLC Analysis of Citrolive Bioactive Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
The HPLC equipment that was used was a Hewlett-Packard Series HP 1100 that was equipped with a diode array detector. The stationary phase was a C18LiChrospher 100 analytical column (250 × 4 mm i.d.) with a particle size of 5 nm (Merck, Darmstadt, Germany) thermostated at 25 °C.
For the elucidation and quantification of bioactive compounds in the Citrolive™, the extract was dissolved in dimethylsulfoxide (DMSO) at a ratio of 5 mg/mL and this solution was filtered through a 0.45-nm nylon membrane. The flow rate was 1 mL/min and the absorbance changes were monitored simultaneously at 280 and 340 nm. The mobile phases for chromatographic analysis were as follows: (A) acetic acid/water (2.5:97.5) and (B) acetonitrile. A linear gradient was run from 95% (A) and 5% (B) to 75% (A) and 25% (B) during the first 20 min; which changed to 50% each of (A) and (B) after another 20 min (40 min in total); which, after 10 more minutes, changed to 20% (A) and 80% (B) (50 min in total), and finally equilibrated over the last 10 min (60 min in total) to the initial composition.
+ Open protocol
+ Expand
3

HPLC Analysis of Phenolic Components

Check if the same lab product or an alternative is used in the 5 most similar protocols
(HPLC) Analysis High-performance liquid chromatography (HPLC, Hewlett-Packard Series HP 1100 equipped with a diode array detector) was used for determination of OLE components released from coaxial nanofibers. Samples taken from released medium at the end of 1 month were fed to system after filtering through 0.45 lm syringe filter. The stationary phase was a C18 LiChrospher 100 analytical column (250 3 4 mm i.d.) with a particle size of 5 mm thermostated at 308C. The flow rate was 1 mL/min, and the absorbance changes were monitored at 280 nm. The mobile phases for chromatographic analysis were (A) acetic acid/water (2.5:97.5) and (B) acetonitrile. A linear gradient was run from 95% A and 5% B to 75% A and 25% B during 20 min; it changed to 50% A and B in 20 min (40 min, total time); in 10 min it changed to 20% A and 80% B (50 min, total time), after re-equilibration in 10 min (60 min, total time) to initial composition. Phenolic components in released medium were determined by comparing retention times of each component with those of known standards.
+ Open protocol
+ Expand
4

Stability of Flavonoids in Aqueous Solutions

Check if the same lab product or an alternative is used in the 5 most similar protocols
The stability of naringenin and hesperetin in aqueous solution at different pHs (3.5, 6.5 and 8.5) was studied at 25 ºC. For this purpose, a 1 ml aliquot of naringenin or hesperetin solution (1 mg/mL in 80 % ethanol) was mixed with 5 ml of buffer solution (13 % ethanol) containing (or not) 6 mM of CDs (β-or HP-β-CDs). At different time intervals, the remaining
naringenin or hesperetin concentration was determined by HPLC (Hewlett-Packard Series HP 1100 equipped with a diode array detector). The stationary phase was a C18 LiChrospher 100 analytical column (250 x 4 mm i.d.) with a particle size of 5 µm (Merck, Darmstadt, Germany) thermostatic at 30 ºC. The mobile phase consisted of acetonitrile/water (60/40) with a 1 mL/min flow rate, and the injection volume was 20 µL. Naringenin or hesperetin was detected at 290 or 292 nm, respectively and they were quantified buy using a calibration curve from 0.005 to 0.05 mM prepared in ethanol 80%.
+ Open protocol
+ Expand
5

HPLC-DAD Analysis of Flavonoid Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
For the HPLC–DAD analysis, a Hewlett—Packard HP series 1100 (Hewlett—Packard, Wilmington, DE, USA) instrument operated with ChemStation software v. A.09.03 and a Discovery C18 column (250 × 4.6 mm, particle size of 5 µm) was used. The sample was eluted with an isocratic mixture of methanol/acetonitrile/water (25:25:50) with a flow rate of 1 mL/min and measured at a wavelength of 260 nm with a complete screen from 200 to 400 nm. Thirty microlitres of a stock solution of 3 mg/mL of extract in HPLC—grade MeOH was injected into the instrument. Additionally, 30 µL of 15 flavonoid standards (vanillin, quercetin, catechin, luteolin, pinocembrin, chrysin, baicalein, myricetin, naringenin, naringin, gallic acid, catechol, apigenin, acacetin and kaempferol; Sigma—Aldrich, St. Louis, MO, USA) were injected at a concentration of 1 mg/mL.
+ Open protocol
+ Expand
6

Purification of Latex Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
Increasing concentrations of methanol and water were used in a Sephadex LH-20 (SIGMA-ALDRICH, USA) column, with 5 mL of latex. Fifty-eight fractions were obtained, and the higher yielding (14) fractions were loaded onto the high-performance liquid chromatography (HPLC) HP Series 1100 separations module from Hewlett-Packard (Wilmington, DE, USA), equipped with an Allphere ODS-1 column of 250×4.6 mm, with a particle size of 5 μm. The mobile phase consisted of methanol: acetonitrile: H2O (25:25:50) for 25 min, flow one mL/min. A diode array detector (DAD) wavelength of 280 nm with a full scan of 200–400 nm was used.
+ Open protocol
+ Expand
7

Quantitative Analysis of Phenolic Acids via HPLC

Check if the same lab product or an alternative is used in the 5 most similar protocols
Phenolic acids were analyzed using a Hewlett Packard (HP) series 1100 chromatograph and detector. The HPLC apparatus was equipped with a degasser (G1322A), a quaternary pump (G1311A), a column thermostat (G1316A), and an autosampler (G1313A). Detection was achieved at 280 nm with diode array detector (DAD) type G1315A. The stationary phase was a C18 Hypersyl ODS column (250 × 4 mm, i.d.) with a particle size of 5 µm and thermostated at 20 °C. The mobile phase consisted of two solvents, A: water/formic acid (99/1, v/v) and B: methanol/formic acid (99/1, v/v). Gradient elution was applied at a flow rate of 1 mL min−1. Solvent B was gradually eluted from 10 to 100% in 25 min. The sample volume injection was 20 µL, and the UV absorbance was determined at 250, 280, 300, and 320 nm.
+ Open protocol
+ Expand
8

RP-HPLC Analysis of Lipoxygenase Products

Check if the same lab product or an alternative is used in the 5 most similar protocols
Reverse-phase high performance liquid chromatography (RP-HPLC) analysis was performed with an HP series 1100 instrument (Hewlett Packard, Waldbronn, Germany) equipped with diode array detector (DAD) and a HP ODS Hypersil column (5 μm, 125 mm × 4 mm). The isocratic mobile phase was acetonitrile:0.01M phosphate buffer (pH 2.8) containing 5% acetonitrile (70:30). Sample size injected was 10 μL and the flow rate was 1 mL/min. Products of LOX activity were detected at 234 nm (Villafuerte Romero and Barrett, 1997 ). Peaks were identified from retention times (Figure 2) and UV-spectra, and were manually integrated using the software HP ChemStation Version 05.01.

Reverse-phase chromatogram of lipoxygenase products detected at 234 nm of a) an active lipoxygenase enzyme and b) an inactivated lipoxygenase enzyme.

Figure 2
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!