The largest database of trusted experimental protocols

Silicone tubing

Manufactured by Cole-Parmer
Sourced in United States, Israel

Silicone tubing is a flexible, clear, and durable material used for a variety of laboratory applications. It is resistant to chemicals, heat, and extreme temperatures, making it suitable for use in various laboratory settings. Silicone tubing is designed to provide a reliable and efficient means of transferring liquids and gases in laboratory equipment and setups.

Automatically generated - may contain errors

10 protocols using silicone tubing

1

Inertial Microfluidic Cell Separation

Check if the same lab product or an alternative is used in the 5 most similar protocols
Inertial microfluidic cell separation was performed using a closed-loop system that included a circulation pump, a microfluidic chip (see above), a pinch valve and a flowmeter. The circulation pump (peristaltic, diaphragm and piezoelectric pump, Cole–Parmer) was connected to the sample and to the microfluidic chip via silicone tubing (Cole–Parmer). To evenly load a sample into each channel, a 0.16 cm barbed five-way connector was used and an equal length of tubing was directly plugged into the inlet holes of the microfluidic chip from the connector. Both the inner-wall and outer-wall outlet tubing were also of equal lengths. The inner-wall outlet was connected to the sample for recirculation and the outer-wall outlet was connected to a pinch valve and flowmeter to adjust the fluidic resistance, before ending in the waste tube (Fig. 1a,b and Supplementary Fig. 1). After preparation, PBS−/− was introduced at a flow rate of 4 ml min−1 to prime the system before sample loading.
+ Open protocol
+ Expand
2

Simulating Interstitial Flow in HUVEC Cultures

Check if the same lab product or an alternative is used in the 5 most similar protocols
Fluid flow was controlled with a programmable syringe pump (Harvard Apparatus). The flow medium was the same as the growth medium (EGM). Before initiation of flow-based experiments, clear polypropylene barbed elbow fittings (1/16 inch, Cole-Parmer) connected to silicone tubing (Saint-Gobain) were inserted into the 1.5-mm diameter inlet/outlet ports of the HUVEC channels. The opposite inlet/outlet ports of the HUVEC channels were connected to luer adapters (Cole-Parmer) which served as fluid reservoirs. To subject endothelial cells solely to interstitial flow (transverse convection), we pulled media through the collagen matrix from the reservoirs connected to the other endothelial-lined channel, while exposing both channels to negligible levels of tangential shear. A pressure gradient across the collagen gel is also generated. In this configuration, fluid extravasates from one HUVEC channel, convects through the collagen and intravasates into the other channel. The flow rate was 7.3 μl/h corresponding to a flow velocity of 60 μm/sec. Devices without imposed flow served as controls (“static”). In this case, the media was changed every day.
+ Open protocol
+ Expand
3

Inertial Microfluidic Cell Separation

Check if the same lab product or an alternative is used in the 5 most similar protocols
Inertial microfluidic cell-separation was performed using a closed-loop system that includes a circulation pump, a microfluidic chip (vide supra), pinch valve, and a flow-meter. The circulation pump (i.e., peristaltic, diaphragm, and piezoelectric pump, Cole-parmer, USA) was connected to the sample and to the microfluidic chip via silicone tubing (Cole-parmer, USA). To evenly load a sample into each channel, a 1/16 inch barbed 5-way connector was used, and equal length of tubing was directly plugged into the inlet holes of the microfluidic chip from the connector. Both inner-wall and outer-wall outlet tubing was also of equal lengths. The inner-wall outlet was connected to the sample for recirculation, and the outer-wall outlet was connected to a pinch valve and flow-meter to adjust fluidic resistance, before ending in the waste tube (Fig. 1AB and Supplementary Fig. 1). After preparation, PBS −/− was introduced at a flow rate of 4 mL/min to prime the system before sample loading.
+ Open protocol
+ Expand
4

Bioreactor Perfusion System Setup

Check if the same lab product or an alternative is used in the 5 most similar protocols
BRECS were placed in a continuous recirculation circuit with 250 mL of supplemented media in a 500 mL sterile perfusion bottle. Custom tubing sets for BRECS consisted of PharMed BPT double-stop pump tubing (Cole-Parmer, Vernon Hills, IL) and silicone tubing (Cole-Parmer). A Masterflex peristaltic pump (Cole-Parmer) was used to maintain the BRECS at a constant perfusion of between 10 to 100 mL/min. Media changes on the BRECS were performed twice per week.
+ Open protocol
+ Expand
5

Pressure-Driven Microfluidic Valve Actuation

Check if the same lab product or an alternative is used in the 5 most similar protocols
There are one flow channel and one control channel in the valve device. The inlet of the flow channel is connected to a bottle containing blue food coloring dye. The bottle is pressurized by a computer-controlled air pressure source (Elveflow OB1 MK3) to drive the dye into the flow channel. The fluid driving pressure is fixed at 0.2 psi. To observe the flow rate, the outlet of the flow channel is connected to the silicone tubing (0.79 mm I.D., Cole Parmer). As for the control channel part, the outlet is sealed with a plastic tube cap and the inlet is also connected to the air pressure source starting from 0 psi. The air pressure is increased in increments of 1 psi in the control channel until the valve closes such that movement of the front end of the fluid in the flow channel stops. Experiments were conducted in triplicate.
+ Open protocol
+ Expand
6

Turbidostat-controlled Cell Culture Maintenance

Check if the same lab product or an alternative is used in the 5 most similar protocols
Cell density during experiments was maintained at an OD600 of 0.1 via a turbidostat composed of two peristaltic pumps (Preciflow, Lambda-instruments) with an RS-232 interface that is controlled via a microcontroller. A photodiode (BP104, Vishay Semiconductors) and an infrared LED (LD274-3, Osram Semiconductors) are used to measure culture turbidity. The infrared LED has a peak emission wavelength of 950 nm, which is well-separated from the longest wavelength at which the red-absorbing form of CcaS shows any absorption (∼750 nm) (ref. 48 (link)). The photodiode measurements are sent to a PI controller implemented in the microcontroller, which maintains turbidity at a pre-specified value by operating the peristaltic pumps. A programmable logic controller was used as the microcontroller for the GFP experiments. It was replaced by an Arduino for the growth-control experiments, because of its amenability to being interfaced with a computer. Media is fed and removed from the culture via silicone tubing (1.6 mm Φinternal × 3.2 mm Φexternal, Cole Parmer). Additional details are provided in Supplementary Note 1.
+ Open protocol
+ Expand
7

Microfluidic Filtration System for Cell Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
In a sterile hood (Class II type A/B3, Nuaire), a 30 mL BD syringe (BD Biosciences, Franklin Lakes, NJ) without a plunger was clamped vertically. The bottom half of a filter holder assembly (Millipore, Billerica, MA), was attached to 3 inches of silicone tubing (Cole Parmer, Vernon Hills, IL) and placed into a stopper on top of a 250 mL glass Erlenmeyer flask. The Erlenmeyer arm was then attached to a VacuGene Pump (GE Healthcare Piscataway, NJ) with a regulated pressure gauge (Ashcroft). A fabricated microfilter was washed in 10 mL PBS, placed onto the filter holder and centered 5 mm beneath the 30 mL syringe (ESI, Fig. 1A-C). This “open” system allows for a consistent vacuum exertion, without pressure build up across the membrane.
Three flow rates, and subsequent negative pressures exerted, were determined using both PBS and whole human blood. One filter was placed into a whole Swinnex filter holder (Millipore) and set up on the system. Vacuum pump was adjusted to reach flow rates (subsequent pressures in mbar) of 1 mL min−1 (10 mbar), 5 mL min−1 (15.1 mbar), and 10 mL min−1 (18.5 mbar). Pre-stained samples were prepared and were run in triplicate in 15 mL PBS, 7.5 mL PBS with 7.5 mL fixative, 7.5 mL whole blood with 7.5 mL PBS, and 7.5 mL Blood with 7.5 mL 1× fixative under the 3 flow rates described.
+ Open protocol
+ Expand
8

Clot Diameter Measurement in Microscopic Flow

Check if the same lab product or an alternative is used in the 5 most similar protocols
The clot attached to the suture was placed in a clean micropipette (Drummond Scientific Company, Broomall, PA), and inserted into a U-shaped sample holder composed of hollow luer lock connectors and silicone tubing (Cole Parmer, Vernon Hills, IL; outer diameter 0.125″). The sample holder was placed in an acrylic water tank with a microscope slide at the bottom allowing visualization of the clot diameter using an inverted microscope (Olympus, IX-71) with a charge-coupled device (CCD) camera (Retiga 2000R, QImaging, Surrey, BC, Canada). The CCD camera images were recorded at a rate of 6 images/minute. The average clot width (CW) was calculated, using a computer program written in Matlab 6.5 R13 (Mathworks, Inc., Natwick, MA). The positions of the two clot-plasma interfaces were determined via an edge-detection routine. The width of the clot at each coordinate along the height of the image (z) was calculated, averaged over all z values, corrected for suture width, and normalized to the average of the clot width during the first six frames. The fractional clot loss (FCL) at 30 minutes can be defined as:
FCL=1-CW30/CW0 where,
+ Open protocol
+ Expand
9

Comparative Evaluation of Coronary Stents

Check if the same lab product or an alternative is used in the 5 most similar protocols
Flow loops were assembled from silicone tubing (Cole-Palmer), connected via hose barb connectors (Cole-Parmer) (Fig. 1). Coronary stents were deployed within the tubing based on the manufacturer's suggested nominal pressure to obtain a stent-to-tube ratio of 1.1:1.0. Coated and uncoated COBRA stents were compared with Resolute Onyx stents (n = 4 experimental runs) and with Synergy stents (n = 4 experimental runs). Each flow loop contained only one stent at a time, so one experimental run comparing 3 different types of stents (coated COBRA-PzF vs. uncoated COBRA vs. Resolute Onyx and coated COBRA-PzF vs. uncoated COBRA vs. Synergy) comprised 6 flow loops in a side-by-side experimental design (control vs. cytokine storm conditions, for each stent).

Experimental setup. A) Each flow loop was assembled from silicone tubing. B) Whole blood was circulated for 60 min using a perfusion pump. C) Enlarged view on the stents deployed in silicone tubes.

Fig. 1
+ Open protocol
+ Expand
10

Perfusion Vessel Comparison for Cell Culture

Check if the same lab product or an alternative is used in the 5 most similar protocols
Polystyrene culture flasks with vented caps were used as control standard vessels (Corning). Three bags were herein used: polyolefin/EVA copolymer (Evolve, Origen), custom FEP (American Durafilm), and custom silicone rubber bags were fabricated in house from sheets (Rogers Corporation). Custom silicone bags were fitted with silicone tubing (Cole Parmer), and sealed with adhesive sealant designed for silicone (DAP). Culture flasks and Origen bags came pre-sterilized. FEP and silicone bags/tubing were sterilized through exposure to 0.5M NaOH for a minimum of one hour. NaOH sterilized vessels were subsequently flushed out three times with sterile PBS to remove all trace NaOH. Specific details on vessels can be found in Table 1.
Perfusion with silicone cell culture bags was accomplished through peristaltic pumping (Cole Parmer). Flow rate was set at 100mL/min and run for 1–2 minutes total.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!