The largest database of trusted experimental protocols

Agilent 1150 lc system

Manufactured by Agilent Technologies
Sourced in United States

The Agilent 1150 LC system is a liquid chromatography (LC) system designed for analytical and preparative separation and purification of chemical compounds. It features a quaternary pump, an autosampler, a column compartment, and a UV-Vis detector. The system is capable of performing gradient and isocratic elution.

Automatically generated - may contain errors

4 protocols using agilent 1150 lc system

1

LC-MS Analysis of Protein Extracts

Check if the same lab product or an alternative is used in the 5 most similar protocols
Extracts were resuspended in 5% ACN with 0.2% formic acid to a final concentration of 2 μg/μL. For LC-MS analysis, 40 μg of sample was loaded onto a 150 mm × 2.1 mm i.d., 2 μm particle size Kinetex C18 RPLC column (Phenomenex, Torrance, CA). Analysis was performed using an Agilent 1150 LC system (Agilent, Santa Clara, CA) equipped with a photodiode array and placed in-line with a Q-Exactive mass spectrometer (Thermo Fisher Scientific, Waltham, MA). Chromatography was performed at a flow rate of 200 μL/min using water/0.1% formic acid (solvent A) and acetonitrile/0.1% formic acid (solvent B) with the following gradient: time 0 min., 2% B; 35 min., 60% B; 54 min., 98% B. UV spectra were acquired at a rate of 1 Hz. For every spectrum, the three most intense apices were recorded. The mass spectrometer instrument settings were as follows: capillary temperature 275 °C, sheath gas 8 (arbitrary units), spray voltage 4.2 kV. Full MS spectra were acquired at 35,000 resolution for the mass range m/z 250 to 3750 for all samples. This resulted in an average scan rate of 6 Hz. Following each full MS scan, the top 5 most intense ions were selected for a dependent MS2 scan. MS2 was conducted using HCD with a collisional energy of 25%.
+ Open protocol
+ Expand
2

LC-MS Analysis of Protein Extracts

Check if the same lab product or an alternative is used in the 5 most similar protocols
Extracts were resuspended in 5% ACN with 0.2% formic acid to a final concentration of 2 μg/μL. For LC-MS analysis, 40 μg of sample was loaded onto a 150 mm × 2.1 mm i.d., 2 μm particle size Kinetex C18 RPLC column (Phenomenex, Torrance, CA). Analysis was performed using an Agilent 1150 LC system (Agilent, Santa Clara, CA) equipped with a photodiode array and placed in-line with a Q-Exactive mass spectrometer (Thermo Fisher Scientific, Waltham, MA). Chromatography was performed at a flow rate of 200 μL/min using water/0.1% formic acid (solvent A) and acetonitrile/0.1% formic acid (solvent B) with the following gradient: time 0 min., 2% B; 35 min., 60% B; 54 min., 98% B. UV spectra were acquired at a rate of 1 Hz. For every spectrum, the three most intense apices were recorded. The mass spectrometer instrument settings were as follows: capillary temperature 275 °C, sheath gas 8 (arbitrary units), spray voltage 4.2 kV. Full MS spectra were acquired at 35,000 resolution for the mass range m/z 250 to 3750 for all samples. This resulted in an average scan rate of 6 Hz. Following each full MS scan, the top 5 most intense ions were selected for a dependent MS2 scan. MS2 was conducted using HCD with a collisional energy of 25%.
+ Open protocol
+ Expand
3

Metabolomics of Inactivated GABA-AT

Check if the same lab product or an alternative is used in the 5 most similar protocols

Example 28

Metabolomics of the Inactitaved GABA-AT. A 50 μL portion of ammonium bicarbonate buffer (50 mM, pH 7.4) containing GABA-AT (23 μg, 0.21 nmol), α-ketoglutarate (5 mM), and 17 (43 mM) was protected from light and incubated at rt overnight until the activity of GABA-AT was less than 1%. Another 50 μL of ammonium bicarbonate buffer (50 mM, pH 7.4) containing GABA-AT (23 μg, 0.21 nmol) and α-ketoglutarate (5 mM) was subjected to the same condition as a control. After incubation, formic acid (10 μL) was added to each sample, and 20 μL of each resulting solution was loaded onto a 5-μm Luna C18 column (2 mm i.d.; 150 mm) (Phenomenex, Torrance, Calif., USA). A 30-min LC gradient was employed at a flow rate of 200 μl/min on an Agilent 1150 LC system (Agilent, Santa Clara, Calif., USA). Mass spectrometry was performed on a Q-Exactive mass spectrometer (Thermo Fisher Scientific, Waltham, Mass., USA). Intact MS spectra were acquired at a resolution of 35,000. The top-five most intense ions were selected for fragmentation in a data-dependent acquisition mode. Mass spectra were acquired at a resolution of 17,500.

+ Open protocol
+ Expand
4

Metabolomic Profiling of Inactivated GABA-AT

Check if the same lab product or an alternative is used in the 5 most similar protocols

Example 28

Metabolomics of the Inactitaved GABA-AT.

A 50 μL portion of ammonium bicarbonate buffer (50 mM, pH 7.4) containing GABA-AT (23 μg, 0.21 nmol), α-ketoglutarate (5 mM), and 17 (43 mM) was protected from light and incubated at rt overnight until the activity of GABA-AT was less than 1%. Another 50 ∞L of ammonium bicarbonate buffer (50 mM, pH 7.4) containing GABA-AT (23 μg, 0.21 nmol) and α-ketoglutarate (5 mM) was subjected to the same condition as a control. After incubation, formic acid (10 μL) was added to each sample, and 20 μL of each resulting solution was loaded onto a 5-μm Luna C18 column (2 mm i.d.; 150 mm) (Phenomenex, Torrance, Calif., USA). A 30-min LC gradient was employed at a flow rate of 200 μl/min on an Agilent 1150 LC system (Agilent, Santa Clara, Calif., USA). Mass spectrometry was performed on a Q-Exactive mass spectrometer (Thermo Fisher Scientific, Waltham, Mass., USA). Intact MS spectra were acquired at a resolution of 35,000. The top-five most intense ions were selected for fragmentation in a data-dependent acquisition mode. Mass spectra were acquired at a resolution of 17,500.

+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!