The largest database of trusted experimental protocols

6890n gc fid system

Manufactured by Agilent Technologies
Sourced in United States

The 6890N GC-FID system is a gas chromatograph equipped with a flame ionization detector (FID). It is designed for the separation, identification, and quantification of organic compounds in complex samples. The system provides reliable and sensitive analysis for a wide range of applications.

Automatically generated - may contain errors

6 protocols using 6890n gc fid system

1

GC-FID and GC-MS Analysis of Organic Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
GC was carried out using a Agilent 6890 N GC-FID system, equipped with a flame ionization detector (FID) on a Agilent capillary column, HP-5 (30 m × 0.32 mm; film thickness 0.25 μm). The column temperature was programed from 40 °C to 250 °C for 5 °C/min. The column temperatures of injector and detector were fixed to 250 °C. Helium was used as the carrier, flow rate 1.0 ml/min.
The GC–MS analysis was carried out using a Agilent 6890 N GC-FID system, equipped with a flame ionization detector (FID) on a Agilent capillary column, HP-5 (30 m × 0.32 mm; film thickness 0.25 μm). The column temperature was programed from 40 °C to 250 °C at a rate of 5 °C/min. The column temperatures of injector and detector were 250 °C. Helium was used as the carrier, flow rate 1.0 ml/min. Split ratio was 1:100. The GC–MS analysis was done in the EI mode at 70 eV, inlet temperature was 200 °C and transfer line temperature was 250 °C. The temperature program was the same with that of the GC analysis. The injected volume was 0.2 μl.
+ Open protocol
+ Expand
2

GC-FID Analysis of Compound Concentrations

Check if the same lab product or an alternative is used in the 5 most similar protocols
The gas chromatography (GC) analyses were carried out using an 6890N GC-FID system equipped with DB-5 capillary column (30 m × 0.25 mm; film thickness of 0.25 mm) (Agilent Technologies, Santa Clara, United States) connected to an FID detector. The injector and detector temperatures were set at 280 °C at a rate of 5 °C/min. The carrier gas was helium (>99.2% purity) at a flow rate of 1.3 mL/min. All samples were analyzed in duplicate. Relative component concentrations were calculated based on GC peak areas without using correction factors [18] .
+ Open protocol
+ Expand
3

GC-FID Analysis of M. suaveolens Oil

Check if the same lab product or an alternative is used in the 5 most similar protocols
The gas chromatography (GC) analysis was performed with Agilent Technologies 6890N GC-FID system, equipped with DB-5 capillary column (30 m × 0.32 mm; 0.50 µm) and connected to a FID detector. The thermal programmer was 60 °C (1 min) to 180 °C at 3 °C/min; injector temperature 220 °C; detector temperature 220 °C; split ratio 1:10; carrier gas Helium; flow rate: 1.0 mL/min. The injected volume of M. suaveolens essential oil was 1 μL diluted in chloroform (1:10). Two replicates of sampleswere processed in the same way.
Component relative concentrations were calculated based on GC peak areas without using correction factors (Bezerra et al., 2017) .
+ Open protocol
+ Expand
4

Gas Chromatographic Analysis of Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
The gas chromatography (GC) analyses was performed with Agilent Technologies 6890N GC-FID system, equipped with DB-5 capillary column (30 m x 0.32 mm; 0.50 µm) and connected to an FID detector. The thermal programmer was 60 °C (1min) to 180 °C at 3 °C/min; injector temperature 220 °C; detector temperature 220 °C; split ratio 1:10; carrier gas Helium; flow rate: 1.0 mL/min. The volume injected 1 μL diluted in chloroform (1:10). Two replicates of samples were processed in the same way. Component relative concentrations were calculated based on GC peak areas without using correction factors (Boligon; Feltrin; (link)Athayde, 2013) (link).
+ Open protocol
+ Expand
5

GC-FID Analysis of Total Oil Samples

Check if the same lab product or an alternative is used in the 5 most similar protocols
GC-FID analysis of the total oil samples was performed according to Martins et al. 2017 , with an Agilent 6890N GC-FID system with synthetic air, H 2 , and N 2 as flame gases, an HP-5 capillary column (30 m x 0.32 mm x 0.25 μm), and 5α-androstane (0.02 mg.mL -1 ) applied as the internal standard for n-alkane quantification. n-Alkanes, pristane (Pr), and phytane (Ph) were identified based on the chromatographic profile of reference samples and their retention time.
+ Open protocol
+ Expand
6

GC-FID and GC-MS Analysis of Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
Gas chromatography was performed using a 6890N GC-FID system (AGILENT TECHOLOGIES, Santa Clara, California, USA), equipped with DB-5 capillary column (30m × 0.32 mm; 0.50 mm) and connected to a flame ionization detector (FID). GC-MS analyses were performed on an AutoSystem XL GC-MS system (AGILENT TECHNOLOGIES) operating in the EI mode at 70 eV, equipped with a split/splitless injector (220 °C). The relative amounts of individual components were calculated using their GC peak areas (Boligon et al., 2013) .
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!