The largest database of trusted experimental protocols

Acquity liquid chromatography system

Manufactured by Waters Corporation
Sourced in United States

The Acquity liquid chromatography system is a high-performance liquid chromatography (HPLC) instrument designed for analytical and preparative applications. It is capable of separating, identifying, and quantifying components within a sample mixture.

Automatically generated - may contain errors

7 protocols using acquity liquid chromatography system

1

Quantifying Plasma and PBMC Acylcarnitines

Check if the same lab product or an alternative is used in the 5 most similar protocols
The concentrations of acylcarnitines in the plasma and PBMC samples were determined with a UPLC MS/MS method using a Waters ACQUITY liquid chromatography system and a Waters Quattro Micro or Waters Xevo TQ‐S mass spectrometer, as previously described (Liepinsh et al., 2017).
+ Open protocol
+ Expand
2

Time-Resolved Quantification of Sim-L and Sim-H

Check if the same lab product or an alternative is used in the 5 most similar protocols
Sim‐H and Sim‐L were prepared as 10 mM solutions in water and DMSO, respectively, and stored at −78°C, and warmed to room temperature immediately prior to injection; 1 μL of each solution was added to 1 mL of water or Tris/HEPES (250 mM HEPES, 80 mM Tris, 10 μM free Ca2+, pH 7.2) buffer, respectively, and sampling by LC‐MS was begun immediately. Samples were kept at room temperature for the duration of the analysis. Analyses were performed using a Thermo Exactive mass spectrometer equipped with Waters Acquity liquid chromatography system. Instrument control was performed using Thermo Xcalibur Software. Electrospray source conditions were adjusted to maximize sensitivity; 3 μL of each sample was injected onto the system. The column was a 2.1 × 50 mm (3 μm) ACE equivalence C18. A mixed eluent gradient was employed and is described in Supporting Information Table S3. The four sample types were analysed sequentially; the 5 min analysis time for each sample provided a 20 min sampling frequency for each sample. Mass analysis was performed using MestReNova MS software (2016, V11.0), where the peaks were integrated to determine relative percentages of Sim‐L and Sim‐H over time.
+ Open protocol
+ Expand
3

UPLC-MS/MS Quantification of Analyte

Check if the same lab product or an alternative is used in the 5 most similar protocols
Waters Acquity liquid chromatography system coupled with a Waters TQD triple quadrupole mass spectrometer was used. Mass spectrometric detection was carried out using an electrospray interface (ESI) operated in the negative ionization mode with multiple reaction monitoring (MRM) for both IRB and IS. Nitrogen was used as a desolvating gas at a flow rate of 500 L/h. The desolvating temperature was set at 400°C and the source temperature was set at 150°C. The collision gas (argon) flow was set at 0.1 mL/min. The capillary voltage was set at 3.2 kV. The MS analyzer parameters were as follows: LM1 and HM1 resolution 10.0 and 8.0; ion energy 1, 1 V; LM2 and HM2 resolution 15.0 and 10.0, respectively; ion energy 2, 0.1 V, dwell time, 0.146 seconds. The cone voltage and collision energy were optimized in case of each analyte so as to maximize the signal corresponding to the major transition observed in the MS/MS spectra, following the fragmentation of the [M+H]+ ions corresponding to the selected compounds. The Mass Lynx software (Version 4.1, SCN 805, Waters, Milford, MA, USA) was used to control the UPLC-MS/MS system as well as for data acquisition and processing.
+ Open protocol
+ Expand
4

Homocysteine Quantification by LC-MS/MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
Total homocysteine concentrations in plasma were simultaneously measured by the Biochemical Genetics Unit, Department of Clinical Biochemistry, Cambridge University Hospitals NHS Foundation Trust using underivatized liquid chromatography tandem‐mass spectrometry (LC‐MS/MS) (Waters ACQUITY liquid chromatography system and Quattro Premier mass spectrometer) operated in electrospray ionization positive mode as was previously described in detail (Padmanabhan et al. 2013).
+ Open protocol
+ Expand
5

Plasma Acylcarnitine Profiling by UPLC-MS/MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
The concentrations of acylcarnitines in the plasma samples were determined with a UPLC MS/MS method using a Waters Acquity liquid chromatography system and Waters Quattro Micro or Waters Xevo TQ-S mass spectrometer, as previously described (17 (link), 27 (link)).
+ Open protocol
+ Expand
6

Glycan Profiling of Anti-CD20 Antibody

Check if the same lab product or an alternative is used in the 5 most similar protocols
Glycan analysis was performed to study the heterogeneity of the test anti-CD20 antibody sample (P2-CRD-1508) compared to that of the control antibody sample (RS-03-25), in terms of glycosylation, which is a common post-translation modification in recombinant monoclonal antibodies expressed in mammalian cell lines. Sample preparation and glycan labeling was done using the glycan labeling kit (Ludger, UK). Antibody samples were first reduced and alkylated. N-glycans were released from the Fc part of reduced antibody samples after digestion with PNGase enzyme. Released N-glycans were separated by using S-cartridges and concentrated by centrifugal evaporation. The dried N-glycan was derivatized and labeled using 2–AB (amino benzamine) florescent dye. Waters Acquity® liquid chromatography system (Waters, USA) with fluorescence detector (excitation, 330 nm; emission, 420 nm) was employed to separate and monitor 2-AB labeled glycans using Acquity BEH glycan column (Waters, USA) with a run time of 55 min in gradient elution with mobile phase-A (100 mM ammonium formate buffer, pH 4.5), mobile phase-B: (100% acetonitrile), and mobile phase-C (water for injection).
+ Open protocol
+ Expand
7

LC-MS Profiling of Strawberry and Ginger

Check if the same lab product or an alternative is used in the 5 most similar protocols
LC-MS profiling was performed on methanolic extracts of strawberry and ginger according to the method described by Haggag et al. [36 (link),37 (link),38 (link),39 (link)] on an Acquity Liquid Chromatography system coupled to a Synapt G2 HDMS quadrupole time-of-flight hybrid mass spectrometer (Waters, Milford, CT, USA). The database used for the identification of compounds was the Dictionary of Natural Products.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!