The largest database of trusted experimental protocols

Afg3022b

Manufactured by Tektronix
Sourced in United States, United Kingdom

The AFG3022B is a dual-channel arbitrary function generator produced by Tektronix. It generates waveforms with a frequency range of up to 25 MHz and amplitude up to 10 Vpp. The instrument offers multiple standard waveform types and the ability to create custom waveforms.

Automatically generated - may contain errors

20 protocols using afg3022b

1

Biasing Contact Characterization Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
The chip carrier socket was soldered to a printed circuit board and directly used for biasing the contacts by a power supply. A hole has been drilled through the chip carrier socket and the PCB board. A DC power supply was used for measurements that did not require a high switching speed, and an arbitrary function generator (Tektronix AFG3022B) was used for temporal measurements. A DAQ (USB-3114) was used for experiments that required multiple switchable power sources. In order to make sure the DAQ pins support adequate power into the system, multiple pins have been connected to the contact and programmed to supply the required voltage to the contacts simultaneously.
+ Open protocol
+ Expand
2

Acoustic Field Characterization for Ultrasound

Check if the same lab product or an alternative is used in the 5 most similar protocols
Ultrasound fields were generated as described previously [28 ] and the acoustic exposure set-up is illustrated in Figure A1 (Appendix A). Briefly, a 1-cm diameter unfocused piezoceramic transducer was mounted at the bottom of a plastic exposure tank filled with degassed, deionized water. The transducer was driven by a continuous sinusoidal signal at its fundamental frequency (8.8 MHz) using a function generator (AFG3022B; Tektronix, Beaverton, OR, USA,), attenuator (837; KayPENTAX, Montvale, NJ, USA), and RF power amplifier (2100L; ENI, Rochester, NY, USA). Acoustic fields were characterized using both needle (HNC-0400; Onda, Sunnyvale, CA, USA) and capsule (HGL-0085; Onda) hydrophones. A location in the far field was selected (10.5 cm from transducer) such that the transaxial beam width was 3 mm. Acoustic fields were calibrated in the free field before and after each experiment for both amplitude (peak positive and peak negative pressure, MPa) and spatial peak pulse average intensity (ISPPA, W/cm2). Values from each calibration were averaged across all experiments and are reported as mean ± SEM in Table 1.
+ Open protocol
+ Expand
3

Acoustic Actuation for Microfluidic Separation

Check if the same lab product or an alternative is used in the 5 most similar protocols
The chip was acoustically actuated by a piezoelectric transducer glued underneath the separation channel and connected to a waveform generator (Tektronix AFG3022B, UK Ltd., Bracknell, UK) through an in-house built amplifier. The driving voltage was monitored with an oscilloscope (Tektronix TDS2002C, UK Ltd., Bracknell, UK) and held at 15 V peak-to-peak through an in-house written software controlling the waveform generator. The actuation voltage was chosen to be the highest possible where temperature feed-back loop could still retain the chip at 25 °C. The actuation frequency was tuned by visual optimization of the acoustic focusing and was set to 1.99 MHz unless otherwise stated.
+ Open protocol
+ Expand
4

Acoustic Field Characterization for Ultrasound

Check if the same lab product or an alternative is used in the 5 most similar protocols
Ultrasound fields were generated as described previously [28 ] and the acoustic exposure set-up is illustrated in Figure A1 (Appendix A). Briefly, a 1-cm diameter unfocused piezoceramic transducer was mounted at the bottom of a plastic exposure tank filled with degassed, deionized water. The transducer was driven by a continuous sinusoidal signal at its fundamental frequency (8.8 MHz) using a function generator (AFG3022B; Tektronix, Beaverton, OR, USA,), attenuator (837; KayPENTAX, Montvale, NJ, USA), and RF power amplifier (2100L; ENI, Rochester, NY, USA). Acoustic fields were characterized using both needle (HNC-0400; Onda, Sunnyvale, CA, USA) and capsule (HGL-0085; Onda) hydrophones. A location in the far field was selected (10.5 cm from transducer) such that the transaxial beam width was 3 mm. Acoustic fields were calibrated in the free field before and after each experiment for both amplitude (peak positive and peak negative pressure, MPa) and spatial peak pulse average intensity (ISPPA, W/cm2). Values from each calibration were averaged across all experiments and are reported as mean ± SEM in Table 1.
+ Open protocol
+ Expand
5

Acoustophoresis Imaging of Blood Cells

Check if the same lab product or an alternative is used in the 5 most similar protocols
To visualize the acoustophoretic motion of cells, the chip was mounted upside down in an inverted microscope (Eclipse Ti2, Nikon, Tokyo, Japan) equipped with a CMOS camera (Prime 95B, Teledyne Photometrics, Tucson, Arizona). Stained PBMCs and neutrophils were spiked into the whole blood and imaged at the end of the separation channel. To monitor the density-adjusted medium at the channel outlet, we added a fluorescent tracer molecule (Dextran, Cascade Blue, 3000 MW, Thermo Fisher Scientific) under the assumption that iodixanol and dextran molecules diffuse at the same rate (Diodixanol ≈ 2.5 × 10–10 m2 s−1 and Ddextran ≈ 2.2 × 10–10 m2 s−1). A laser illumination unit (Celesta light engine, Lumencor, OR, USA) was used with a multiband filter set (CELESTA-DA/FI/TR/Cy5/Cy7-A-000, Semrock optical filters, IDEX Health & Science) in three excitation channels with peak wavelengths at 365 nm, 488 nm, and 561 nm. The excitation channels and the camera were activated through external triggering.
When operating the chip outside of the AcouWash system the PZT transducer was driven by a function generator (AFG3022B, Tektronix, Inc., Beaverton, Oregon, USA) to deliver a resonant frequency (1.99 MHz) and different applied voltages as measured over the piezo with an oscilloscope (TDS1002, Tektronix, Inc., Beaverton, Oregon, USA).
+ Open protocol
+ Expand
6

Acoustophoretic System Setup and Characterization

Check if the same lab product or an alternative is used in the 5 most similar protocols
The standing wave field was created using piezoceramic transducers glued underneath the pre-focusing channel as well as underneath the main separation channel. Frequencies were set to 4.831 MHz for the pre-focusing channel with 5 Vpp amplitude and 1.956 MHz for the main separation channel. A dual channel function generator (AFG3022B, Tektronix, Beaverton, OR, USA), connected to signal amplifiers (in-house build) was used to drive both transducers while the voltage over each transducer was measured via a two-channel digital oscilloscope (TDS 1002, Tektronix). Temperature regulation was achieved using a Peltier element and a PT100 resistance temperature detector attached to the acoustophoretic system.
+ Open protocol
+ Expand
7

Optogenetic Stimulation of Neuronal Synapses

Check if the same lab product or an alternative is used in the 5 most similar protocols
L4 transgenic worms were transferred to standard agar plates with food and 450 µM all-trans-Retinal overnight (Liewald et al., 2008 (link)). A custom array of 457 nm blue LEDs controlled by a Tektronix AFG 3022B pattern generator, delivered 10 ms light pulses at 20 Hz for 30 seconds every minute with an intensity of 3 mW/ mm2. For image analysis, postsynaptic AMPARs in AVA (marked by SEP or GFP) showing clear overlap with VAMP∷mCherry in PLM were identified as PLM synapses (Figure 9C, 9D and 9G, red arrowheads). We typically identified two synapses in the region with VAMP∷mCherry fluorescence that met this criterion, which is consistent with the number of synaptic inputs from PLM to AVA predicted by electron microscopic reconstruction of the C. elegans nervous system. PLM also provides synaptic inputs to other neurons in the ventral cord, consistent with the large region of VAMP∷mcherry fluorescence (http://wormweb.org/neuralnet#c=PLM&m=1). SEP or GFP puncta underlying VAMP∷mCherry, but without overlap, were termed neighboring synapses (Figure 9D and 9G, yellow arrowheads), and SEP or GFP puncta outside of the VAMP∷mCherry regions were termed distal synapses (Figure 9D and 9G, light blue arrowheads). Signals were corrected for fluorescence background and quantified in ImageJ.
+ Open protocol
+ Expand
8

Targeted Gene Delivery via Ultrasound

Check if the same lab product or an alternative is used in the 5 most similar protocols
Either 300 μg Cre-recombinase-IRES-GFP or 300 μg GFP plasmid in 50 μl volume, mixed with 200 μl OPTISON (GE Healthcare) microbubbles (MBs) (~1.2 × 108 MBs), were injected into the tail vein 18 days after the first STZ injection using previously established protocols [42 (link)]. Immediately following MB injection, mice were anesthetized with 3% isoflurane and an ultrasound (US) Applicator Head probe was placed on the shaved skin over the dorsal kidney area, using the following acoustic parameters: 1.0 MHz ultrasound frequency, 0.7 MPa peak negative pressure, 30 s pulse repetition period, and 2 min duration of exposure. The custom experimental ultrasound (US) (supported by UAB Small Animal Imaging Shared Facility) setup included a single element (0.75 in.) immersion transducer (Olympus, Waltham, MA) connected to a signal generator (AFG3022B, Tektronix, Beaverton, OR) and power amplifier (A075, Electronics and Innovation, Rochester, NY). The plasmid/MB mixture was injected via tail vein (i.v.) 18 days following STZ injection and this was repeated 2 weeks later.
+ Open protocol
+ Expand
9

Ex vivo Liver Vibration Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
Two small identical vibration sources (Model DAEX320-8, 8 Ohm, 20 W, Dayton Audio, Ohio, USA) were put in contact with the ex vivo liver tissue as shown in Figure 3 (c); these generated the RSW within the tissue for the ultrasound scan. To create an effective contact surface between each vibration source and the tissue, a conical knob was attached and fixed on each source as shown in the figure. These vibrator sources were connected to a power amplifier (BKA1000-4A, ButtKicker, Westerville, OH, USA) driven by a dual-channel function generator (AFG3022B, Tektronix, Beaverton, OR, USA) which provided input signals to the vibration sources. Vibration frequency ranges between 200 – 400 Hz were used for the ex vivo liver experiment.
+ Open protocol
+ Expand
10

Memristor Hysteresis Loop Characterization

Check if the same lab product or an alternative is used in the 5 most similar protocols
We used an AC voltage to draw a memristor hysteresis loop and a function generator (AFG3022B, Tektronix, Beaverton, OR, USA.) to apply the AC voltage and a picoammweter (6487, Keithley) to measure the ion current. AC voltages were used ranging from −5 V to 5 V; frequencies of 10, 1, and 0.1 Hz; and concentrations of 1, 10, and 100 mM. The order of injecting the solutions and connecting platinum (PT) electrodes was the same as in Section 2.2. The hysteresis loop is plotted for five periods with the gray line from −5 V to 5 V, and the bold black line is the average of the five period ion currents.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!