The largest database of trusted experimental protocols

Dxr smartraman

Manufactured by Thermo Fisher Scientific
Sourced in United States

The DXR SmartRaman is a compact, benchtop Raman spectrometer designed for routine analysis. It provides high-quality Raman data with minimal sample preparation. The instrument features automated alignment and optimization, enabling consistent and repeatable measurements.

Automatically generated - may contain errors

8 protocols using dxr smartraman

1

Comprehensive characterization of advanced carbon materials

Check if the same lab product or an alternative is used in the 5 most similar protocols
The morphological, elemental and chemical composition, and structure of the ACs were characterized using field emission scanning electron microscopy (FE-SEM, JEOL JSM-700F), X-ray photoelectron spectroscopy (XPS, PHI Quantera II) and Raman spectroscopy with a laser wavelength of 532 nm (Thermo Scientific DXR SmartRaman), respectively. Nitrogen adsorption–desorption isotherm measurements were performed by degassing at 350 °C for 12 h using a gas adsorption analyzer (Micromeritics ASAP 2020). The Brunauer–Emmett–Teller (BET) surface area was analyzed from the nitrogen adsorption isotherms, while the pore size distribution and pore volume were calculated via the density functional theory (DFT) method. The Hall effect measurement was utilized to characterize the carrier density of ACs using van der Pauw method, employing a four-point probe under a magnetic field of 0.2 T with a sample dimension of 10 mm diameter and 0.3 mm thickness (KEITHLEY 6221 DC and AC current source and 2182 NANOVOLTMETER).
+ Open protocol
+ Expand
2

Raman Spectroscopy of Powdered Mucilage

Check if the same lab product or an alternative is used in the 5 most similar protocols
The chemical composition of powdered mucilage was analyzed using a Raman spectrophotometer (DXR™ Smart Raman, Thermo Scientific, Waltham, MA, USA) equipped with a 785 nm excitation diode laser. Spectra were measured with an average scan time of 1.0 s using a laser power of 20.0 mW. A total of 20 scans per spectra were performed to improve the signal-to-noise ratio.
+ Open protocol
+ Expand
3

Raman Characterization of NMC Materials

Check if the same lab product or an alternative is used in the 5 most similar protocols
Characterization of the NMC was performed using a confocal Raman microscope (DXR SmartRaman, Thermo Fisher Scientific). A 532 nm laser with a power of 2 mW was used as the excitation source. The laser beam has been focused on the NMC surface with a 50 × objective, resulting in a focused spot with a diameter of ≈ 0.8 μm, and a spot distance of 10 µm. The detector was arranged in the backscattering configuration and equipped with a grating with 900 grooves mm−1. All Raman spectra were recorded in the wavenumber range between 60 and 1800 cm−1.
+ Open protocol
+ Expand
4

Comprehensive Characterization of BCOF-1 Material

Check if the same lab product or an alternative is used in the 5 most similar protocols
The nuclear
magnetic resonance data (1HNMR and 13CNMR) were
obtained by NMR-400 MHz (Bruker), and 13CSS-NMR data were
obtained by Bruker Avance HD 400 MHz w/4 mm HX and 1.6 mm HFX solid
probes. Attenuated total reflection infrared spectroscopy (ATR-IR)
was used to collect the infrared spectra. The surface topography images
were captured at high magnifications by scanning electron microscopy
(SEM, Hitachi SU-70 FE-SEM). A 3Flex surface analyzer (Micrometrics)
was utilized at 77 K to collect the nitrogen adsorption/desorption
isotherms and the Brunauer–Emmett–Teller (BET) surface
area measurements. BCOF-1 was activated by degassing at 80 °C
for 4 h and then was increased to 110 °C for 7 h at a rate of
5 min °C1 under a 10–6 bar vacuum. X-ray photoelectron microscopy (XPS) measurements were
carried out using a PHI VersaProbe III scanning XPS microprobe. For
Raman spectroscopy studies, Thermo Scientific DXR Smart Raman was
used at a 532 nm laser with a power of 5 mW to excite the samples.
+ Open protocol
+ Expand
5

Resonance Raman Spectroscopy of Organometallic Complexes

Check if the same lab product or an alternative is used in the 5 most similar protocols
Resonance Raman spectra were collected in a 180° (780 nm) or 90° (all other lines) geometry. Coherent Innova 70C (5W) Ar+ and 300C (1W) Kr+ ion lasers were used as the photon sources. The scattered radiation was passed through a longpass filter (Semrock RazorEdge) to remove Rayleigh scattered laser light and then dispersed onto a liquid N2 cooled 1” Infrared Associates CCD detector using a Princeton Acton spectrograph. The laser power at the sample was kept between 40 and 100 mW in order to prevent possible photo- and thermal degradation of the sample. 780 nm Raman spectra were collected with a Thermo-Scientific DXR SmartRaman, using a low 2 mW laser power due to the observed sensitivity of Cp2V(bdt) to photodegradation. Solid samples were prepared as finely ground powders and dispersed in a NaCl(s) matrix with Na2SO4 added as an internal standard. These samples were subsequently either sealed in a glass capillary tube and spun with a custom-made sample holder or thinly spread on carbon tape and held in a standard brass holder (780 nm). The construction of resonance Raman profiles was accomplished by comparing the integrated intensity of a Raman band at a given excitation wavelength relative to that of the 992.4 cm−1 band of Na2SO4 or solvent bands. All data were scan-averaged, and any individual data set with vibrational bands compromised by cosmic events was discarded.
+ Open protocol
+ Expand
6

Comprehensive Characterization of Lithium Anode SEI

Check if the same lab product or an alternative is used in the 5 most similar protocols
The surface composition was confirmed by X‐ray photoelectron spectroscopy (ESCALAB 250Xi, Thermo Fisher Scientific Inc., USA). The morphologies of lithium anode were analyzed by SEM SU‐8010 (Hitachi) at 5 kV. High‐Resolution Transmission Electron Microscope was performed on JEM‐2100 at 200 kV. The Cryoelectron microscopy was carried out on Talos F200C 200 kV. Attenuated total reflection Flourier transformed infrared spectroscopy (ATR‐FTIR) was carried out on Nicolet5700. Raman spectra were carried out on ThermoFisher DXR SmartRaman. 1H‐NMR spectra were performed on a 600 MHz DirectDrive2 spectrometer. Before test, the SEI residue from four disks was washed with DME solution and peeled off into glass bottle and dried for 12 h under 45 °C and 3 h in vacuum. Thereafter, DMSO‐d6 solution was added and heated under 40 °C for 1 h to dissolve SEI.
+ Open protocol
+ Expand
7

Detailed Physicochemical Characterization of Materials

Check if the same lab product or an alternative is used in the 5 most similar protocols
The purity of RHS was confirmed by a Raman spectrometer (DXR SmartRaman, Thermo Scientific). The crystallography of materials was analyzed by X-ray diffraction (XRD) technique (SmartLab, Rigaku). The surface area and porosity of materials were studied by N2 adsorption-desorption analysis technique (autosorb iQ, Quantachrome instruments) where the samples were degassed at 300 °C for 6 h with a heating rate of 10 °C min−1 before measurement. The structure and functional groups of materials were carried out by Fourier-transform infrared (FT-IR) spectroscopy (Spectrum Two, PerkinElmer). The thermal stability of materials was evaluated by Thermogravimetric analyzer (TG 209 F3 Tarsus, NETZSCH) under the N2 atmosphere. The morphology and composition of materials were observed by scanning electron microscopy (SEM) with the energy dispersive X-ray (EDX) spectroscopy (EVO MA10, Zeiss). The chemical identity of the samples was investigated by solid-state Nuclear magnetic resonance (NMR) spectroscopy (JNM-ECZ-400 R/S1, JEOL) at 400 MHz. The acidity was studied by NH3-TPD analysis (Belcat B, BEL JAPAN, Inc.) where the process for ammonium desorption analysis was carried out according to the literature [27 ].
+ Open protocol
+ Expand
8

Raman Analysis of Mucilage Compositions

Check if the same lab product or an alternative is used in the 5 most similar protocols
The surface composition of the Opuntia ficus-indica and aloe vera powdered mucilages were analyzed using a Raman spectrophotometer (DXR™ Smart Raman, Thermo Scientific, Waltham, MA, USA) equipped with a 785 nm excitation diode laser. Spectra were measured with an average scan time of 1.0 s using a laser power of 20.0 mW. A total of 20 scans per spectra was performed in order to improve the signal-to-noise ratio.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!