The largest database of trusted experimental protocols

Van guard beh c18 pre column

Manufactured by Waters Corporation

The Van Guard BEH C18 pre-column is designed to protect the analytical column from contaminants and particulates, thereby extending the column's lifetime and ensuring reliable chromatographic results. It features a Waters Ethylene Bridged Hybrid (BEH) C18 stationary phase packed in a robust PEEK housing.

Automatically generated - may contain errors

3 protocols using van guard beh c18 pre column

1

Quantitative Analysis of Bisphenol A Analogues

Check if the same lab product or an alternative is used in the 5 most similar protocols
The separation and quantitation of bisphenol A analogues were performed on a Waters Acquity Ultra Performance Liquid Chromatography (UPLC) system (Milford, MA, USA) coupled to a Waters Xevo TQD triple quadrupole mass spectrometer (Milford, MA, USA). Ten μL of sample extract was injected in full loop and the separation was performed at 30 °C on an Acquity UPLC BEH C18 column (1.7 μm, 2.1 mm × 50 mm, Waters) with a Van Guard BEH C18 pre-column (1.7 μm, 2.1 × 5 mm, Waters). The mobile phase consisted of (A): 5% methanol in water and (B): 2 mM NH4OH in methanol. The gradient program was as follows: initial gradient 100% (A) hold for 1min, to 55% (B) in 1 min, to 100% (B) in 10 min, back to 100% (A) in 0.5 min, and hold for 3.5 min at 100% (A). Flow rate was set at 0.2 mL/min. The MS/MS was operated in multiple reaction monitoring (MRM) using electrospray ionization in negative ion mode (ESI-). The MRM transitions and MS parameters were obtained and optimized using Waters IntelliStart (Waters, MA, USA). MRM transitions and optimized MS parameters are summarized in Table 1. Source temperature, desolvation temperature, and gas flow were set at 150 °C, 350 °C and 600 L/h, respectively. Capillary voltage was set at 2.5 kV and extractor voltage at 10 V.
+ Open protocol
+ Expand
2

Lipidomic Analysis of Serum Samples

Check if the same lab product or an alternative is used in the 5 most similar protocols
Lipidomic analysis was performed according to Simcox et al43 (link),44 (link); briefly, 40μl aliquots of sera were combined with 250μls PBS and 225μls ice-cold MeOH containing internal standards (Avanti Splash cat#3307–07), and homogenized. Samples were then mixed with 750μls of ice-cold MTBE (methyl tert-butyl ether), rehomogenized, and separated by centrifugation (17,000g for 5 mins/4°C). The upper phase was transferred to a new tube, lyophilized and resuspended in 150 μls of isopropanol. Lipids were analyzed by UHPLC/MS/MS in positive and negative ion modes, at a dilution suitable to eliminate saturation issues. Extracts were separated on an Agilent 1260 Infinity II UHPLC system using an Acquity BEH C18 column (Waters 186009453; 1.7 μm 2.1 × 100 mm) maintained at 50 C with VanGuard BEH C18 precolumn (Waters 18003975), using the chromatography gradients described by Jain et al. The UHPLC system was connected to an Agilent 6546 Q-TOF MS dual AJS ESI mass spectrometer and run in both positive and negative modes as described. Samples were injected in a random order and scanned between 100 and 1,500 m/z. Tandem MS was performed at a fixed collision energy of 25 V. The injection volume was 2 μl for positive mode and 5 μl for negative mode.
+ Open protocol
+ Expand
3

UHPLC-HRMS Metabolite Profiling Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
UHPLC-HRMS measurements for determination of exact mass, sum formula and retention time were carried out with a Dionex Ultimate 3000 SL system consisting of SWPS 3000 SL autosampler, P680 pump module, TCC100 column oven, and PDA100 UV-detector coupled to a maXis 4 G UHR-TOF platform. The ionization mode was ESI with the following MS settings: capillary voltage 4000 V, end plate off-set -500 V, nebulizer gas pressure 1 bar, dry gas flow rate 5 L/min, dry gas temperature 200 °C, mass scan range 150–2500 m/z. The mass spectrometer was externally calibrated to sodium formate cluster using subsequent lock mass calibration. Metabolite profiling was performed using Waters Acquity BEH C18 columns (50 × 2.1 mm, 1.7 µm and 100 × 2.1 mm, 1.7 µm) connected to a Waters VanGuard BEH C18 pre-column. The standard methods for both columns are linear gradients with 5–95% ACN + 0.1 % FA over 6, 9, or 18 min. The LC-MS results were analyzed by Bruker Compass DataAnalysis V4.4.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!