The largest database of trusted experimental protocols

27 protocols using 670 ir spectrometer

1

Measuring ATR-FTIR Spectra with Ge Crystal

Check if the same lab product or an alternative is used in the 5 most similar protocols
Measurements of ATR-FTIR background-corrected spectra (25 scans for each sample) were carried out with the use of a HATR Ge trough (45° cut, yielding 10 internal reflections) crystal plate at 20 °C, and were recorded with a 670-IR spectrometer (Agilent, USA). The Ge crystal was cleaned with ultra-pure organic solvents (Sigma-Aldrich). The instrument was continuously purged with argon for 40 min. before and during measurements. Absorption spectra at a resolution of one data point per 1/cm (to the highest measurement accuracy) were obtained in the region between 4000 and 400/cm. Scans were Fourier-transformed and averaged with Grams/AI 8.0 software (Thermo Fisher Scientific, USA).
+ Open protocol
+ Expand
2

FTIR Imaging of Tissue Samples

Check if the same lab product or an alternative is used in the 5 most similar protocols
Cary 620 Imaging FTIR equipped with an Agilent 670‐IR spectrometer and 128 × 128‐pixels FPA mercury cadmium telluride (MCT) detector was used in the transmission mode. Background spectra were collected on CaF2 substrates using 128–256 scans at 8 cm−1 spectral resolution. Sample spectra were recorded using 16–128 scans for tissues at 8 cm−1 spectral resolution. A 15 × IR objective (pixel size of 5.5 µm, 0.62 NA) and a 25 × IR objective (pixel size of 3.3 µm, 0.81 numerical aperture (NA)) were used for the imaging.
+ Open protocol
+ Expand
3

Infrared Spectroscopy of Powdered Samples

Check if the same lab product or an alternative is used in the 5 most similar protocols
Measurements of infrared spectra were conducted with the use of a 670-IR spectrometer (Agilent, Santa Clara, CA, USA)24 at 23 °C. An ATR (Attenuated Total Reflection) attachment was used in the form of a ZnSe crystal with adequate geometry (truncated at 45°) to ensure 20-fold internal reflection of the absorbed beam. The powders were pressed onto the crystal surface. During the measurement, 16 scans were registered and subsequently the programme averaged the results for all spectra. Prior to the measurement, the ZnSe crystal was cleaned using pure organic solvents by Sigma-Aldrich (Darmstadt, Germany). Spectral measurements were recorded in the region from 400 to 4000 cm−1 at the resolution of 1 cm−1.
+ Open protocol
+ Expand
4

Infrared Spectrometry of Biological Samples

Check if the same lab product or an alternative is used in the 5 most similar protocols
The infrared spectra of the analyzed samples were measured using 670-IR spectrometer (Agilent, USA). To ensure 20-fold internal reflection of the absorbed beam, Attenuated Total Reflectance (ATR) attachment was used in the form of a ZnSe crystal with adequate geometry (truncated at 45°). Sixteen scans were registered during the measurement, and subsequently, the program averaged the results for all spectra. Prior to the measurement, the ZnSe crystal was cleaned using ultraclear solvents (Sigma-Aldrich). Before (1 h) and during the experiment, the measurement chamber was kept in an inert N2 atmosphere. Spectral measurements were recorded in the region from 700 to 3800 cm−1 at a resolution of 1 cm−1. The measurements were taken at the Central Apparatus Laboratory of the University of Life Sciences in Lublin. The spectra were analyzed and processed using Grams/AI software developed by ThermoGalactic Industries (USA). All the spectra were measured at 23 °C.
+ Open protocol
+ Expand
5

Infrared Spectroscopy Analysis Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
Measurements of infrared spectra for the analyzed samples were conducted with the use of a 670-IR spectrometer (Agilent, Santa Clara, CA, USA). An ATR (attenuated total reflection) attachment was used in the form of a ZnSe crystal with adequate geometry (truncated at 45°) to ensure 20-fold internal reflection of the absorbed beam. During the measurement, 24 scans were registered, and, subsequently, the programme averaged the results for all spectra. Prior to the measurement, the ZnSe crystal was cleaned using ultra-clear solvents by Sigma-Aldrich (Sigma-Aldrich, St. Louis, MO, USA). Prior to (1 h) and during the experiment, the measurement chamber was kept in an inert N2 atmosphere. Spectral measurements were recorded in the region from 550 to 3800 cm−1 at the resolution of 0.5 cm−1. The measurements were conducted at the Department of Biophysics Laboratory of the University of Life Sciences in Lublin. The spectra were analyzed and processed with the use of the Grams/AI software (version Suite 9.3) by ThermoGalactic Industries (Waltham, MA, USA). The spectra were normalized at the wave number of 1025 cm−1 (for easier analysis). All the spectra were measured at room temperature.
+ Open protocol
+ Expand
6

Infrared Spectroscopy of Analytical Samples

Check if the same lab product or an alternative is used in the 5 most similar protocols
Infrared spectra measurements for the analyzed samples were conducted with the 670-IR spectrometer (Agilent, Santa Clara, CA, USA). An ATR (attenuated total reflection) attachment was used in the form of a ZnSe crystal with adequate geometry (truncated at 45°) to ensure 20-fold internal reflection of the absorbed beam. During the measurement, 16 scans were registered and, subsequently, the program averaged the results for all spectra. Prior to the measurement, the ZnSe crystal was cleaned using ultra-clear solvents by Sigma-Aldrich. Prior to (1 h) and during the experiment, the measurement chamber was kept in an inert N2 atmosphere. Spectral measurements were recorded in the region from 500 to 4000 cm−1 at the resolution of 1 cm−1. The measurements were conducted at the Central Apparatus Laboratory of the University of Life Sciences in Lublin. The spectra were analyzed and processed with the Grams/AI software by ThermoGalactic Industries (Waltham, MA, USA). All the spectra were measured at 23 °C.
+ Open protocol
+ Expand
7

Imaging FTIR Analysis of Biological Cells

Check if the same lab product or an alternative is used in the 5 most similar protocols
Agilent Cary 620 Imaging FTIR equipped with an Agilent 670-IR spectrometer and 128 × 128-pixels FPA mercury cadmium telluride (MCT) detector was used in the transmission mode. A background spectrum was collected on a clean CaF 2 substrate using 128 scans at 8 cm -1 spectral resolution, suggesting that the IR absorbance were measured every 4 cm -1 . Cell spectra were recorded using 64-128 scans at 8 cm -1 spectral resolution. A ×25 IR objective (pixel size, 3.3 μm, 0.81 numerical aperture (NA)) was used for cell imaging.
+ Open protocol
+ Expand
8

Infrared Spectroscopy of Honey Powders

Check if the same lab product or an alternative is used in the 5 most similar protocols
Measurements of infrared spectra for 14 analysed samples (11 honey powders and 3 basic materials: honey, maltodextrin, and milk powder) were conducted with the use of a 670-IR spectrometer (Agilent, Santa Clara, CA, USA) at 23 °C. An ATR (Attenuated Total Reflection) attachment was used in the form of a ZnSe crystal with adequate geometry (truncated at 45°) to ensure 20-fold internal reflection of the absorbed beam. The powders were pressed onto the crystal surface. During the measurement, 16 scans were registered and subsequently, the programme averaged the results for all spectra. Prior to the measurement, the ZnSe crystal was cleaned using ultra-pure organic solvents by Sigma-Aldrich (Darmstadt, Germany). Prior to (1h) and during the experiment, the measurement chamber was kept in an inert N2 atmosphere. Spectral measurements were recorded in the region from 500 to 4000 cm−1 at the resolution of 1 cm−1.
+ Open protocol
+ Expand
9

ATR-FTIR Spectroscopy of Solvents

Check if the same lab product or an alternative is used in the 5 most similar protocols
The IR(ATR) measurements were done at of the facilities of the University of Life Sciences in Lublin using a 670-IR spectrometer (Agilent, Santa Clara, CA, USA) equipped with ZnSe crystal-containing ATR adaptor. The measurement chamber was kept in a N 2 atmosphere. All spectra were recorded with 16 scans with the subsequent software averaging. All solvents used for cleaning the crystals before each measurement were ultrapure grade (Sigma-Aldrich, St. Louis, MO, USA) The resolution for spectra recording, which was performed at 23 • C, was set to 1 cm -1 , while the utilized range was equal to 4000-500 cm -1 . Processing of the recorded spectra was carried out by Grams/AI software (ThermoGalactic Industries, Waltham, MA, USA).
+ Open protocol
+ Expand
10

Aggregation Kinetics of TTR Variants

Check if the same lab product or an alternative is used in the 5 most similar protocols
The aggregation kinetics of TTR-WT and TTR-L55P were studied by FTIR measurements in attenuated total reflection (ATR) as previously reported11 (link),64 (link). In particular, 3 µL of sample aliquots were deposited on the single reflection diamond crystal of the ATR device and the FTIR spectra were collected after solvent evaporation. A 670-IR spectrometer (Varian Australia, Mulgrave, Australia) equipped with a nitrogen-cooled mercury cadmium telluride detector was used under the following conditions: 2.0 cm−1 spectral resolution, 25 kHz scan speed, 1000 scan coadditions, and triangular apodization. The second derivatives of the measured spectra were obtained after Savitsky-Golay smoothing using the Resolutions-Pro software (Varian Australia).
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!