The largest database of trusted experimental protocols

10 protocols using beh glycan column

1

Fluorescent Glycan Labeling and HILIC Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
Released glycans were fluorescently labeled via reductive amination with 2-aminobenzoic acid (2-AB) and sodium cyanoborohydride in 30% v/v acetic acid in DMSO at 65° C for 3 hours. Excess labeling reagents and reducing agent were removed from the samples using GlycoClean S solid-phase extraction cartridges according to the manufacturer’s instructions. Hydrophilic interaction chromatography (HILIC) separation of 2-AB labeled glycans was carried out using an Agilent 1100 HPLC system coupled to an Agilent HPLC fluorescence (FLD) detector. Separations were performed using Waters BEH Glycan column, 100 mm × 2.1 mm i.d., 2.5 μm amide sorbent, with the column heated to 60° C. The injection volume was 20 μl. All separations were performed using 100 mM ammonium formate, pH 4.5, as solvent A and 100% acetonitrile as solvent B. The gradient conditions were as follows: 0–5 min, 85–75% B, 0.3 ml min-1; 5–35 min, 75–64% B, 0.3 ml min-1; 35–40 min, 64–50% B, 0.3 ml min-1; 40–42 min, 50–50% B, 0.3–0.1 ml min-1; 42–43 min, 50–10% B, 0.1 ml min-1; 43–48 min, 10–10% B, 0.1 ml min-1; 48–50 min, 10–85% B, 0.1 ml min-1; 50–60 min, 85–85% B, 0.1–0.3 ml min-1. The fluorescence detector excitation and emission wavelengths were set at 260 and 430 nm, respectively.
+ Open protocol
+ Expand
2

Glycan Characterization by Online FLD-MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
Online coupled fluorescence (FLD)-mass spectrometry detection was performed using a Waters Xevo G2 QTof with Acquity UPLC and BEH Glycan column (1.0×150 mm, 1.7 µm particle size). MS data was acquired in negative mode with the following conditions: 2500 V capillary voltage, 50 V cone voltage, 280°C desolvation temperature, 600 L/hour desolvation gas and 100°C source temperature. The analyzer was set to sensitivity mode. The fluorescence data rate was 1 pts/second and a PMT gain = 10 with altering excitation and emission wavelengths that varied based on the experiment and glycan label used. Sample injection volumes were 10 µl and the flow rate was 0.150 µl/min. Solvent A was 50 mM ammonium formate pH 4.4 and solvent B was acetonitrile. A 40 minute linear gradient was used and was as follows: 28–43% A for 32 minutes, 70% A for 4 minutes and 28% solvent A for 4 minutes. Samples were diluted in 65% acetonitrile prior to analysis.
+ Open protocol
+ Expand
3

Online Coupled FLR-MS Analysis of Glycans

Check if the same lab product or an alternative is used in the 5 most similar protocols
Lyophilised cell line secretome samples, were reconstituted in 3 μl of water and 9 μl acetonitrile. Online coupled fluorescence (FLR)-mass spectrometry detection was performed using a Waters Xevo G2 QTof with Acquity® UPLC (Waters Corporation, Milford, MA, USA) and BEH Glycan column (1.0 × 150 mm, 1.7 μm particle size). For MS acquisition data, the instrument was operated in negative-sensitivity mode with a capillary voltage of 1.80 kV. The ion source block and nitrogen desolvation gas temperatures were set at 120 and 400°C, respectively. The desolvation gas was set to a flow rate of 600 L/h. The cone voltage was maintained at 50V. Fullscan data for glycans were acquired over m/z range of 450–2,500. Data collection and processing were controlled by MassLynx 4.1 software (Waters Corporation, Milford, MA, USA). The fluorescence detector settings were as follows ex = 330 nm, em = 420 nm. Sample injection volume was 8 μL. The flow rate was 0.150 mL/min and column temperature was maintained at 60°C; solvent A was 50 mM ammonium formate in water (pH 4.4) and solvent B was acetonitrile. A 40 min linear gradient was used, and was as follows: 28% (v/v) A for 1 min, 28–43% (v/v) A for 30 min, 43–70% (v/v) A for 1 min, 70% (v/v) A for 3 min, 70–28% (v/v) solvent A for 1 min and finally 28% (v/v) A for 4 min.
+ Open protocol
+ Expand
4

Automated Glycan Analysis by UHPLC-FLR-MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
Following derivatization of released glycans with 2-aminobenzamide, glycans were chromatographically resolved by UHPLC using a Waters BEH Glycan column (1.7 µm, 2.1 mM × 150 mM) on a Waters ACQUITY UPLC H-Class Bio (Waters, Milford, MA). Analytes were separated using a 38.5 min gradient of 25–52% 50 mM ammonium formate, pH 4.5, against acetonitrile over a 38.5 min interval with a flow rate of 0.4 mL/min. Analytes were then optically detected prior to MS using an ACQUITY FLR detector with excitation and emission wavelengths of 250 and 428 nm, respectively. Corresponding m/z values for each glycan were determined using a Waters Synapt G2-S QToF instrument (Waters). Mass spectrometric settings included a capillary voltage of 3 kV, cone voltage of 40 V, desolvation temperature of 350°C, desolvation gas of 600 L/h and a source temperature of 150°C. Data were acquired using MassLynx 4.1 and processed using UNIFI 1.7.1 (Waters).
+ Open protocol
+ Expand
5

N-Glycan Analysis by UPLC-QTOF MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
The prepared N-glycans were analyzed by a Waters Acquity ultra-performance liquid chromatography instrument equipped with a fluorescence detector and a Xevo-G2S qTOF mass spectrometer. The system was controlled by MassLynx 4.2 (Waters, Milford, MA, USA). Separations were performed by a Waters BEH Glycan column, 100 × 2.1 mm i.d., 1.7 μm particles, using a linear gradient of 75–55% acetonitrile (Buffer B) at 0.4 mL/min in 42 min, using 50 mM ammonium formate pH 4.4 as Buffer A. An amount of 5 μL of sample was injected using partial loop mode in all runs. The sample manager temperature was 15 °C, and the column temperature was 60 °C during each analysis. The fluorescence detection excitation and emission wavelengths were λex = 308 nm and λem = 359 nm. During the MS analysis, 2.2 kV electrospray voltage applied to the capillary. The desolvation temperature was set to 120 °C, while the desolvation gas flow was 800 L/hr. Mass spectra were acquired using positive ionization mode over the range of 500–2000 m/z. MS/MS fragments were obtained using 45 kV collision energy during the analysis.
+ Open protocol
+ Expand
6

N-Glycan Analysis by HILIC-UHPLC-MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
N-linked glycans (N-glycans) were evaluated by glycan map analysis using hydrophilic interaction liquid chromatography (HILIC) with fluorescence detection. N-glycans were released by treatment with PNGase F. The reducing termini of the released glycans were then labeled through reductive amination with a fluorescent tag, 2 aminobenzoic acid (2-AA, Sigma Aldrich). Following incubation at 80°C, precipitated proteins were pelleted and the supernatant, containing released and labeled glycans, was transferred to a new vial. The labeled glycans were separated by HILIC-UHPLC on a BEH Glycan column (2.1 × 150 mm, 1.7 μm; Waters Corporation), using a 100 mM ammonium formate, pH 3.0 and acetonitrile gradient. The gradient was 22% 100 mM ammonium formate to 40% over 111 min. Peak identification was performed using MS by coupling the HILIC UHPLC with an ion-trap MS for verification against the expected glycan mass.
+ Open protocol
+ Expand
7

N-Glycan Analysis by UPLC-MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
The prepared N-glycans were analyzed using a Waters Acquity ultra-performance liquid chromatography system equipped with a fluorescence detector and a single quad mass detector. The system was controlled by the Empower 3 chromatography software (Waters, Milford, MA, USA). Separations were performed with a Waters BEH Glycan column, 100 × 2.1 mm i.d., 1.7 μm particles, using a linear gradient of 72–55% acetonitrile (Buffer B) at 0.4 mL/min in 42 min, with 50 mM ammonium formate pH 4.4 as Buffer A. Samples were dissolved in 75%/25% acetonitrile/water and 5 μL was injected in all runs. The sample manager temperature was set at 15 °C, while the column temperature was 60 °C during the runs. The fluorescence detection excitation and emission wavelengths were λex = 309 nm and λem = 359 nm. During the MS analysis, a 2.2 kV electrospray voltage was applied to the capillary. The desolvation temperature was set to 120 °C, while the desolvation gas flow was 500 L/h. Mass spectra were acquired over the range of 500–2000 m/z in positive ionization mode and data-independent acquisition.
+ Open protocol
+ Expand
8

Glycan Analysis of Monoclonal Antibodies

Check if the same lab product or an alternative is used in the 5 most similar protocols
The 2‐AA‐derivatized mAb1 samples and controls described earlier were profiled with a Waters Acquity Ultra‐Performance Liquid Chromatography (UPLC) H class System (Milford, MA, USA) equipped with a Waters BEH Glycan column (1.7 µm, 2.1 × 150 mm) (Milford, MA, USA). Solvent A consisted of 100 mm ammonium formate prepared with formic acid (1116701000; Fisher Scientific) and brought to pH 3.0 with ammonium hydroxide (338818; Sigma). Solvent B consisted of 100% acetonitrile (MX0486‐6; Fisher Scientific). The column temperature was maintained at 35°C, and the fluorescence detection was set at λex = 360 nm and λem = 425 nm. The flow rate was 0.25 mL·min−1 with an initial 22% solvent A ramped up to 40% solvent A in 111 min, followed by a 6.5‐min wash at 90% solvent A, then a 5‐min stepdown to 22% solvent A and a 25‐min final reequilibration. The flow rate was decreased by 20% for the wash step to avoid approaching the pressure limit of the system.
+ Open protocol
+ Expand
9

UPLC-MS Analysis of Fluorescent N-Glycans

Check if the same lab product or an alternative is used in the 5 most similar protocols
The fluorescently labeled and purified N-glycans were analyzed using a Waters Acquity ultra-performance liquid chromatography instrument with fluorescence and single quad mass spectrometric detection. Empower 3 chromatography software (Waters, Milford, MA, USA) was used to control the system. Chromatographic separations were performed by a Waters BEH Glycan column, 100 × 2.1 mm i.d., 1.7 μm particles, with a gradient of 72–55% acetonitrile in 42 min at a 0.4 mL/min flowrate using 50 mM ammonium formate pH 4.4 as the mobile phase. Samples were made up of 75%/25% acetonitrile/ water, and 5 μL was injected using partial loop injection in all runs. The sample manager temperature was 15 °C, and the column temperature was 60 °C during each analysis. The fluorescence detection excitation and emission wavelengths were λex = 309 nm and λem = 359 nm. In the MS setup, the electrospray voltage was 2.2 kV, with a desolvation temperature of 120 °C using a 500 L/hr desolvation gas flow. Mass spectra were acquired using a positive ionization mode over the range of 600–2000 m/z. Centroid data was acquired with a scan rate of 500 Da/s, and the MS target resolution was 0.5 Da.
+ Open protocol
+ Expand
10

HILIC-UPLC for Glycan Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
The N-glycan solution was either stored (at 4 C or À40 C) or dried in a rotary vacuum at 37 C. The glycans were re-dissolved in 6 lL of water and 14 lL of MeCN (30/70% solution) giving a 20 lL solution. Of this, 19 lL was injected onto the column for a 20 min runtime. The HILIC-UPLC was carried out on a BEH Glycan column on an Acquity UPLC (Waters, Massachusetts, USA) equipped with a fluorescence detector. Solvent A used was 50 mM ammonium formate solution, pH 4.4. Solvent B was acetonitrile (HPLC grade). The column temperature was set to 40 C. The following conditions were used: 20 min method using a linear increasing gradient of 30-70% solvent A and a linear decreasing gradient of solvent B at 0.561 mL/min. The samples were injected in 70% acetonitrile and 30% water. The fluorescence was measured at 420 nm with excitation at 330 nm and all analytical peaks resolved within 12 min. The system was calibrated using an external standard of hydrolysed, 2-ABlabelled glucose oligomers (Waters, Massachusetts, USA) to create a dextran ladder with retention times of all identified peaks expressed as GU. 30, 34
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!