The largest database of trusted experimental protocols

Precolumn

Manufactured by Knauer
Sourced in Germany

The Precolumn is a component used in liquid chromatography systems. Its primary function is to protect the analytical column from contaminants and particulates that may be present in the sample or mobile phase. The Precolumn acts as a guard column, filtering out these unwanted substances before they reach the main analytical column, helping to extend the column's lifetime and maintain the integrity of the separation.

Automatically generated - may contain errors

4 protocols using precolumn

1

HPLC Analysis of 5-KF and D-Fructose

Check if the same lab product or an alternative is used in the 5 most similar protocols
For qualitative and quantitative analysis of the prepared 5-KF and D-fructose solution, a Knauer Smartline HPLC-system (Knauer GmbH, Berlin, Germany) was used. The system was composed of a degasser (Knauer Smartline manager 5000), a pump (Knauer Smartline pump 1000), an autosampler (Knauer Smartline autosampler 3800), a column oven (Knauer Column-Thermostat Jetstream 2 Plus), an RI detector (AZURA RID 2.1L), and an ultraviolet (UV) detector (Knauer Smartline UV detector 2600), which measured the absorbance at a wavelength of 210 nm. Sample separation was achieved using the Eurokat H column (300 × 8 mm; Knauer GmbH) and a precolumn (30 × 8 mm; Knauer GmbH) heated to 65°C. The mobile phase, 5 mM H2SO4, was applied at a flow rate of 0.6 mL min–1. The injection volume was set to 20 μL. Data evaluation and control of the HPLC system were accomplished using ClarityChrom® 8.2.3 (Knauer GmbH). Peak assignment and quantification were performed by applying external standards of D-fructose and 5-KF in concentrations of 0.5, 1, and 2 mM.
+ Open protocol
+ Expand
2

LC-MS Analysis of Seagrass Extracts

Check if the same lab product or an alternative is used in the 5 most similar protocols
LC-MS analysis was performed on a Shimadzu HPLC system (controller CBM-20A, two pumps LC-20 AD, a column oven CTO-20 AC and a photo diode array detector SPD-M20A; Shimadzu, Darmstadt, Germany) coupled to a Triple Tof 4600 mass spectrometer (AB Sciex, Canby, USA). The separation of extracted compounds was realised on a Knauer Vertex Plus column (250 × 4 mm, 5 μm particle size, packing material ProntoSIL 120–5 C18-H) with precolumn (Knauer, Berlin, Germany). The column oven temperature was set to 30 °C and 25 μl of undiluted methanolic seagrass extract prepared as described above was injected. The solvent flow rate was 0.8 ml/min. In this time, a gradient was run from 10 to 90% B from minute 0 to 35, 2 min of 90% B, switch to 10% B in 1 min and subsequent equilibration at 10% B for 2 min. Solvent A (water) and B (methanol) were both supplemented with 2 mM ammonium acetate and 0.01% acetic acid. Mass spectra were monitored between 100 and 800 Da in negative ionisation mode. In addition, MS/MS spectra were generated with a collision energy of − 30 eV and measured between 50 and 800 Da. Spectra for the most prominent peaks were compared to database entries in MassBank [35 (link)] and ReSpect [36 (link)] for identification.
+ Open protocol
+ Expand
3

Comprehensive Metabolite Identification by LC-MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
For LC-MS analysis, an HPLC (Shimadzu, Darmstadt, Germany) coupled to a triple TOF mass spectrometer (AB Sciex, Canby, USA) was used. Water (A) and methanol (B) supplemented with 2 mM of ammonium acetate and 0.01% acetic acid were used as solvents. The flow rate was 0.8 mL/min, and a gradient was applied with 10% to 90% B over 35 min, two min of 90% B, and switching to 10% B in one min, followed by two min of 10% B. The column oven temperature was 30 °C, and 10 µL of the sample was injected into a Knauer Vertex Plus column (250 × 4 mm, 5 µm particle size, packing material ProntoSIL 120-5 C18-H) with precolumn (Knauer, Berlin, Germany). The ion source was set to negative electrospray ionization mode (ESI–) and mass spectra were recorded between 100–1000 Da. MS/MS spectra were measured at a collision energy of −30 eV between 50–800 Da. Precursor masses and MS/MS peaks were compared to database entries in MassBank [60 (link)] and ReSpect [61 (link)] for identification.
+ Open protocol
+ Expand
4

HPLC Analysis of Pharmaceuticals

Check if the same lab product or an alternative is used in the 5 most similar protocols
HPLC analysis was carried out using a Shimadzu LC 2010HT HPLC system (Shimadzu, Kyoto, Japan) equipped with a serial dual plunger pump, an autosampler, and an SPD-20A/20AV series UV-Vis detectors. LC-solution software was used for peak integration. Chromatographic separation was achieved on a C18 column (250 × 4.6 mm, Eurospher 100 with 5 μm) equipped with a pre-column (Knauer, Berlin, Germany). The mobile phase was filtered through a 0.22 μm filter and degassed by ultrasonication before use. The injection volume was 20 μL. The column temperature was kept at 33 °C during chromatographic operation.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!