The largest database of trusted experimental protocols

Hp 5ms capillary column

Manufactured by Agilent Technologies
Sourced in United States, Germany, United Kingdom, Italy, Canada

The HP-5MS capillary column is a gas chromatography column designed for a wide range of applications. It features a 5% phenyl-methylpolysiloxane stationary phase and is suitable for the separation and analysis of a variety of organic compounds.

Automatically generated - may contain errors

465 protocols using hp 5ms capillary column

1

GC-MS Analysis of Essential Oils

Check if the same lab product or an alternative is used in the 5 most similar protocols
Essential oilswere analyzed by GC-MS using an Agilent 6890 N (Agilent Technologies SpA, Cernusco s/N, Milan, Italy) equipped with a 5973 N mass spectrometer. Separation was achieved on an HP-5 MS capillary column (5% phenylmethylpolysiloxane, 30 m, 0.25 mm i.d., 0.1 μm film thickness; J & W Scientific, Folsom, CA, USA) using helium as the carrier gas (1 mL min−1). The temperature of the oven was set to at 60 °C for 5 min, then raised at 4 °C min−1 up to 220 °C, finally 11 °C min−1 up to 280 °C. The TICs were acquired at 70 eV scanning in the 29–400 m/z range. The oil samples were diluted 1:100 in n-hexane, and the volume injected was 2 μL (three replicates). Data were analysed using MSD ChemStation software (Agilent, Version G1701DA D.01.00) and the NIST Mass Spectral Search Program for the NIST/EPA/NIH Mass Spectral Library v. 2.0. The identification of essential oil components was performed by a comparison of retention indices, calculated using a C7–C30 series of n-alkanes (Merck, Milan, Italy) and mass spectra of unknown peaks with those contained in the commercial libraries WILEY275, NIST 17, ADAMS and FFNSC3, as well as those in a homemade library. Percentage values of essential oil components were obtained from the peak areas in the chromatogram without the use of correction factors.
+ Open protocol
+ Expand
2

Multi-Modal Analytical Techniques

Check if the same lab product or an alternative is used in the 5 most similar protocols
NMR spectra were recorded on a Bruker 400/500/750 MHz. ESI-MS spectra were recorded on a Bruker QTOF ESI-MS instrument. UV-vis kinetic studies were performed using an Agilent 8453 diode array based UV-vis spectrophotometer. Single crystals were diffracted using a Rigaku X-ray single crystal diffractometer. GC-MS was performed on a Thermo Scientific ISQ QD Mass Spectrometer attached with a Thermo Scientific TRACE 1300 gas chromatograph using an HP-5ms capillary column (30 m × 0.25 mm × 0.25 μm, J&W Scientific) with helium as the carrier gas and an Agilent 7890A GC system connected with a 5975C inert XL EI/CI MSD (with a triple axis detector). EPR was recorded with a Bruker EMX plus series instrument using an X-band at 4 K. Mössbauer spectra were recorded at 5.5 K or 80 K on a lowfield Mössbauer spectrometer equipped with a Janis CCR cryostat and at 5.5 K on a strong-field Mössbauer spectrometer equipped with an Oxford Instruments Spectromag 4000 cryostat containing an 8 T split-pair superconducting magnet. Both spectrometers were operated in a constant acceleration mode in transmission geometry. The isomer shifts were referenced against that of a metallic iron foil at room temperature. Analysis of the data was performed with the program WMOSS (WMOSS4 Mössbauer Spectral Analysis Software, www.wmosss.org, 2009–2015) and a home-made program.27
+ Open protocol
+ Expand
3

Gas Chromatography-Mass Spectrometry Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
Analyses were conducted using a 6890N gas chromatograph arrayed with an HP-5MS capillary column (30 m length, 0.25 mm id and 0.25 µm film thickness; J&W Scientific, Folsom, CA, USA) and paired with a time-of-flight mass spectrometer (Leco Corp., St. Joseph, MI, USA). The operating conditions were as follows: GC: split 5:1; oven temperature: initially 40 °C increasing by 5 °C per min to 150 °C and then 40 °C per min to 250 °C; TOF-MS: 40–280 u, centroid mode, 0.1 s scan time, 0.02 s inter-scan time, 2750 V and −70 eV. The GC chromatograms were compared using retention times, mass spectra of reference compounds, and mass spectra of library entries.
+ Open protocol
+ Expand
4

GC-MS Analysis of Cameroonian Essential Oils

Check if the same lab product or an alternative is used in the 5 most similar protocols
The chemical constituents of the Cameroonian essential oils were analyzed on an Agilent 6890 N gas chromatograph coupled to a 5973 N mass spectrometer (Santa Clara, CA, USA) and equipped with a HP-5 MS capillary column (5% phenylmethylpolysiloxane, 30 m, 0.25 mm i.d., 0.1 μm film thickness; J & W Scientific, Folsom, CA, USA). For the separation of the volatile constituents, the following temperature program was used: 5 min at 60 °C then 4 °C/min up to 220 °C, then 11 °C/min up to 280 °C held for 15 min. The injector and detector temperatures were: 280 °C; carrier gas: Helium; flow rate: 1 mL/min; split ratio: 1:50; acquisition mass range: 29–400 m/z; and mode: electron-impact (EI, 70 eV). The essential oil was diluted 1:100 in n-hexane, and then 2 µL of the solution was injected into the GC-MS system. For the identification of essential oil components, co-injection with the authentic standards available in our laboratory (purchased from Sigma-Aldrich) was performed, together with a comparison of the retention indices and the mass spectra of those occurring in the ADAMS, NIST 08, and FFNSC2 libraries [41 ,42 ,43 ]. The percentage values of the volatile components were the means of three chromatographic analyses and were determined from the peak areas without the use of correction factors.
+ Open protocol
+ Expand
5

GC/MS Analysis of Essential Oils

Check if the same lab product or an alternative is used in the 5 most similar protocols
A solution 1:100 v/v in ethyl acetate of each EO was prepared and filtered, then 1 μL of each solution was injected into the GC/MS. Analyses of EOs were carried out on an Agilent 6890 N gas chromatograph hyphenated with a 5973 N mass spectrometer provided with HP-5 MS capillary column (J & W Scientific, Folsom), using the conditions reported in our previous work [28 (link)].
+ Open protocol
+ Expand
6

GC-MS Analysis of Organic Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
For the GC-MS analysis, the Agilent 5860 gas chromatograph was used under the control of AgilentChemStation software E.02.02.1431. Sampling was performed using the autosampler Agilent 7893 in the “splitless” mode, 1 µL volume was injected. Separation was performed on a J&W HP-5MS capillary column (30 m in length, 0.25 mm in diameter, fixed phase film thickness (5% biphenyl, 95% dimethyl polyoxane) 0.1 μm). The carrier gas was helium, with a constant flow of 1.3 mL/min, and the evaporator temperature was 250 °C. Column thermostat temperature: base temperature was 70 °C increased at a speed of 4 °/min to 320 °C and then sustained for 10 min. The chromatogram was recorded by a Agilent 5975C mass-selective detector with m/z range 50–850. The source temperature was 230 °C.
+ Open protocol
+ Expand
7

Pesticide Residue Analysis by GC-MS/MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
A GC-MS/MS system of Agilent (model 7890B-7000C, Santa Clara, CA, USA) was operated in multiple reaction monitoring mode (MRM) with two ion transitions to obtain the maximum sensitivity for the detection of the target molecules, the mass transitions used are presented in Table 1. A HP-5MS capillary column (30 m × 250 µm × 0.25 µm) from J&W Scientific (Forson, CA, USA) was used for pesticide residue analysis, using the multi-mode inlet at 280 °C in splitless mode. The oven was programmed at 150 °C for two min, ramped at 3 °C/min−1 to 200 °C, then ramped to 280 °C at 8 °C/min−1 and then held for 10 min., the carrier gas was Helium (He) at a flow rate of 1 mL/min−1, with a retention time (Rt) as shown in Table 1. MS was operated in electron impact ionization mode (EI). The transfer line, MS source, quad1, and quad2 temperatures were 280, 300, 180, and 180 °C, respectively. Helium (He) quenching gas and N2 collision gas were used at 2.25 and 1.5 mL/min−1, respectively. The system was back flushed at 300 °C at 50 psi for five min., and the method retention time was locked to chlorpyrifos-methyl at 13.093 min. The Rt of the TTP used as IS was 24.162 min.
+ Open protocol
+ Expand
8

Characterization of O. compactum Essential Oil

Check if the same lab product or an alternative is used in the 5 most similar protocols
The essential oil of O. compactum (obtained from flowering aerial parts) was kindly supplied by Pranarôm International (http://www.pranarom.com). Its analysis was achieved on an Agilent 6890 N gas chromatograph equipped with a 5973-N mass spectrometer (MS). The settings for the MS were as follows: EI mode, 70 eV, and mass to charge ratio (m/z) scan between 35 and 400. A HP-5 MS capillary column (30 m × ID 0.25 mm × 0.25 μm film thickness, J & W Scientific, Folsom, CA, USA) using helium gas flow (1.0 ml/min) was used for separation. The GC temperature program was as follows: initial 50 °C for 5 min, then increasing with 20 °C/min to 300 °C. The injector temperature was 150 °C. Quali-quantitative analysis of the essential oil was performed according to the work of Benelli et al. (2017 (link)). Thymol and carvacrol were purchased from Sigma-Aldrich (Madrid, Spain).
+ Open protocol
+ Expand
9

Multimodal Spectroscopic Characterization

Check if the same lab product or an alternative is used in the 5 most similar protocols
X-band EPR
spectra were recorded
on a Bruker 300 spectrometer equipped with an Oxford ESR-910 liquid
helium cryostat. For both instruments, the microwave frequency was
calibrated with a frequency counter and the magnetic field with an
NMR gaussmeter. Mössbauer spectra were recorded with two spectrometers
using a Janis Research Super Vari-temp Dewar. UV–vis spectral
studies were carried out using an Agilent diode array 8453 spectrophotometer
with an attached electrically controlled thermostat. Gas chromatography
(GC) was performed on a PerkinElmer Arnel Clarus 500 instrument equipped
with a hydrogen flame ionization detector; BP20 columns (polar) (12
m × 0.53 mm × 1 μm) were used with helium as the carrier
gas at a flow rate of 1 mL min–1. GC-MS was performed
on an Agilent 5977A mass selective detector interfaced with an Agilent
7890B gas chromatograph using an HP-5ms capillary column (30 m ×
0.32 mm × 0.25 μm, J&W Scientific). HR-MS was performed
in a Thermo Scientific Q-Exactive Orbitrap analyzer using an electrospray
ionization source connected with a C18 column (150 m × 4.6 mm
× 8 μm).
+ Open protocol
+ Expand
10

Soil Chemical Analysis Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
The major chemical molecules and ions in soil mentioned above were measured following previously described methods (Xu et al., 2015 (link)). Briefly, Fe(II) concentration was measured based on 1,10-phenanthroline spectrophotometer colorimetric method after 0.5 mol L-1 HCl extraction for 24 h. Concentrations of anions (SO42- and NO3-) and DOC/DON concentrations were determined through Milli-Q water extraction at ratio of 1:10 (w/v) before analysis by ion chromatography and TOC analyzer, respectively. The concentrations of PCP and its intermediate products (2,3,4,5-TeCP, 3,4,5-TCP, and 3,5-DCP) in soils were extracted by ultrasonic extraction and subsequent derivatization by mixing with K2CO3 (10 mL, 0.2 M) and acetic anhydride (0.5 mL) (Lin et al., 2014 (link)). A gas chromatograph (Agilent 6890N, Agilent, Santa Clara, CA, United States) equipped with a 63Ni electric capture detector (Hewlett-Packard 6890, Hewlett-Packard, Palo Alto, CA, United States) and a HP-5 MS capillary column (30 m by 0.32 mm diameter by 0.25 μm) (J&W Scientific, Inc., Folsom, CA, United States) was used to determine the quantity and the species of chlorophenols. The concentration of released CO2 and CH4 were monitored by gas chromatography (GC) equipped with a flame ionization detector (FID) combined with a methane converter (TECHCOMP, China).
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!