The largest database of trusted experimental protocols

42 protocols using nanobrook 90plus pals

1

Characterizing Nanoparticle Properties by DLS

Check if the same lab product or an alternative is used in the 5 most similar protocols
Mean hydrodynamic size and polydispersity index (PDI) were determined by dynamic light scattering (DLS) at an angle of 90° in previously diluted samples with ultrapure water (1:500, v/v) (NanoBrook 90Plus PALS, Brookhaven Instruments, Holtsville, NY, USA) [29 (link),45 (link)]. The same equipment was used to measure zeta potential of previously diluted samples (1:1000, 1 mM KCl) by electrophoretic mobility [29 (link),45 (link)].
The pH value was measured by potentiometry (NanoBrook 90Plus PALS, Brookhaven Instruments, USA). A previously calibrated (pH 4.0 and pH 7.0) electrode probe was placed directly in the samples.
All measurements were carried out in triplicate.
+ Open protocol
+ Expand
2

Nanoemulsion Droplet Size and PDI Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
The droplet size and PDI of a nanoemulsion were measured by DLS (Brookhaven NanoBrook 90Plus PALS) operated at a fixed scattering angle of 90° and a temperature of 25 °C. The sample was prepared by diluting 5 µL of the nanoemulsion solution with 3 mL deionized water in a cuvette. The dilution was performed to eliminate multiple scattering effects and ensure a consistent baseline. For each sample, five sets of 1 min measurements were done to determine the droplet size distribution.
+ Open protocol
+ Expand
3

Droplet Size and Zeta Potential Measurement

Check if the same lab product or an alternative is used in the 5 most similar protocols
The droplet size and ζ potential were both measured by a ζ
potential and nanoparticle size analyzer (NanoBrook 90Plus PALS, Brookhaven
Instruments, USA). The droplet size was tested based on the principle
of the dynamic light scattering (DLS) technique at a scattering angle
of 90°. The ζ potential was determined by using the phase
analysis light scattering (PALS) technique.
+ Open protocol
+ Expand
4

Nanoparticle Characterization and Stability Evaluation

Check if the same lab product or an alternative is used in the 5 most similar protocols
To determine the size and surface charge, nanoparticle suspension was added to a transparent cuvette and was then inserted into the ZetaPALS dynamic light scattering (DLS) detector (NanoBrook 90Plus PALS, Brookhaven Instruments, Holtsville, NY) as previously described (36 (link)). Scanning electron microscopy (SEM, Hitachi S-3000N, Hitachi, Pleasanton, CA) was used to visualize the morphology of nanoparticles. Briefly, 50 μl of the nanoparticle suspension air-dried on a coverslip was silver sputter-coated and inserted into the SEM instrument. To determine the in vitro stability, nanoparticles were suspended in saline (0.9% Sodium Chloride, NaCl, Crystalline, Fisher Scientific, Hampton, NH, USA) or Vasculife VEGF basal cell media with 10% Fetal Bovine Serum (LL-0003, Lifeline Cell Technologies) and incubated at 37°C for 48 h. Particle size was measured on predetermined time points using DLS as described earlier. The stability of the nanoparticles was represented as the percentage change of nanoparticle size measured at each time point with respect to initial particle size according to the following equation:
+ Open protocol
+ Expand
5

Nanoemulsion Droplet Size Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
A NanoBrook 90plus PALS (Brookhaven Instruments Corporation, USA) particle size analyzer was used to measure nanoemulsion droplet sizes and polydispersity index (PDI). Hydrodynamic droplet size was analyzed by dynamic light scattering. Samples were diluted to approximately 0.001 volume fraction. The dilution was performed to eliminate multiple scattering effects arising from components other than the nanodroplets and their inherent effects on the continuous phase viscosity.
+ Open protocol
+ Expand
6

Particle Size and Zeta-Potential Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
Particle size distribution and zeta-potential of the PG solutions (1 mg/mL) were measured using a NanoBrook 90Plus PALS instrument (Brookhaven Instruments, Holtsville, NY, USA) and analyzed by Particle Solutions v.3.6.07122 (Holtsville, NY, USA).
+ Open protocol
+ Expand
7

Particle Characterization by PALS

Check if the same lab product or an alternative is used in the 5 most similar protocols
The surface charge and average particle size was measured at 25 °C using a NanoBrook 90 Plus PALS (Brookhaven Instruments, USA).
+ Open protocol
+ Expand
8

Particle Characterization by DLS and Zeta Potential

Check if the same lab product or an alternative is used in the 5 most similar protocols
The preliminary assessment of the particle size and size distribution was carried out by dynamic light scattering (DLS) measurements on a NanoBrook 90 Plus PALS instrument (Brookhaven Instruments Corporation), equipped with a 35 mW red diode laser (λ = 660 nm) at a scattering angle of 90 °. The measurements were taken at 25 °C and 37 °C applying a dust cut-off of 30 micellar dispersions at concentrations of 1 mg·mL−1 and fixed volumes of 1.7 mL.
The apparent hydrodynamic radii (Rh90) were determined according to the Stokes–Einstein equation:
where k is the Boltzmann constant, η is the solvent viscosity at temperature T in Kelvin and D90 is the diffusion coefficient measured at an angle of 90°. Each measurement was performed in triplicate.
The electrophoretic light scattering measurements were carried out on the same instrument at a scattering angle of 15° and 25 °C and 37 °C. The principle of phase analysis light scattering (PALS) was applied for the measurements of electrophoretic mobility. The ζ potentials were calculated using the Smoluchowski equation:
where η is the solvent viscosity, υ is the electrophoretic mobility, and ε is the dielectric constant of the solvent.
+ Open protocol
+ Expand
9

Characterizing Freeze-Dried Ng Liposomes

Check if the same lab product or an alternative is used in the 5 most similar protocols
The freeze-dried Ng liposomes were dissolved in PBS for analysis. Before use, the samples were obtained by extruding the solution through a .22 μm microporous membrane three times at 25°C. The particle size distribution of the sample was determined using dynamic light scattering (DLS) via a particle size analyzer (NanoBrook 90plus PALS; Brookhaven Instruments Corporation, Holtsville, NY, United States) at 25°C, and the intensity was detected at 90°. The PDI and zeta potential of the freeze-dried Ng liposomes were determined using the same instrument. All measurements were averaged using at least three replicates, and Brookhaven software was used for the analysis.
+ Open protocol
+ Expand
10

Comprehensive Characterization of BP and BCL

Check if the same lab product or an alternative is used in the 5 most similar protocols
The morphologies of BP and BCL were assessed by transmission electron microscopy (TEM) (JEM-2100 F, Japan Electron Optics Laboratory Co., Ltd., Japan). The morphologies of BCL were observed by a PHILIPS XL30E scanning electron microscope (SEM). Fourier transform infrared (FT-IR) spectra (4000–500 cm single bond1) were recorded using a Nicolet 6700 spectrometer (Nicolet Instrument Co., USA). X-ray diffraction (XRD) patterns were obtained using an X-ray diffractometer (Rigaku Co., Japan) with Cu Kα radiation (λ = 0.15418 nm) at a scanning rate of 4°·min single bond1. Raman spectra were recorded using a laser micro-Raman spectrometer (Jobin Yvon Co., Ltd., France). The zeta potentials of the samples were determined by the Dynamic Light Scattering Instrument (NanoBrook 90 Plus PALS, Brookhaven). Fluorescence spectra were recorded using a fluorometer (F-7100 FL Spectrophotometer, Hitachi).
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!