The largest database of trusted experimental protocols

Acquity uplc hss t3

Manufactured by Waters Corporation
Sourced in United States, United Kingdom, Japan

The ACQUITY UPLC HSS T3 is an ultra-high performance liquid chromatography (UPLC) column designed for the separation and analysis of a wide range of small molecule compounds. It features a High Strength Silica (HSS) particle technology and T3 bonding chemistry, which provide enhanced peak shape, resolution, and sensitivity for improved chromatographic performance.

Automatically generated - may contain errors

206 protocols using acquity uplc hss t3

1

Quantification of Bacterial Lipopeptides

Check if the same lab product or an alternative is used in the 5 most similar protocols
The samples were homogenized with A11 basic ultra-turrax (IKA, Staufen, Germany) and weighed on an XSE204 balance (Mettler-Toledo, Greinfesee, Switzerland), mixed using an MS3 vortex mixer (IKA, Staufen, Germany), and shaken in an automatic horizontal shaker (Hannuo Instruments, Shanghai, China). We performed centrifugation with a Sorvall Biofuge Stratos system (Thermo Fisher Scientific, Waltham, MA, USA). The PCR reaction was performed using an Agilent Sure Cycler 8800 (Agilent, Santa Clara, CA, USA) and results were visualized using Bioshine GelX1520 (Bioshine, Shanghai, China)
TeA was quantified with a Waters Acquity UPLC-tandem quadrupole (TQD) mass spectrometer (Waters Corporation, Milford, MA, USA), which contained an Acquity UPLC HSS T3 (1.8 µm, 2.1 mm × 100 mm) column for separation.
Purification of crude lipopeptides was used with the UltiMate 3000 (Thermo Fisher Scientific, Waltham, MA, USA), which contained a Hypersil PREP HS C18 HPLC (5 µm, 10 mm × 250 mm) column for separation. In addition, the 34 lipopeptides were screened using a Waters Acquity UPLC ion mobility quadrupole time-of-flight mass spectrometer (Waters Corporation, Milford, MA, USA), which contained an Acquity UPLC HSS T3 (1.8 µm, 2.1 mm × 100 mm) column for separation.
+ Open protocol
+ Expand
2

UPLC-MS Separation and Identification of AB23A Metabolites

Check if the same lab product or an alternative is used in the 5 most similar protocols
AB23A and its hydrolytic metabolite were separated by ACQUITY UPLC I-Class plus system (Waters Corporation, Milford, United States) coupled with ACQUITY UPLC HSS T3 (100 mm × 2.1 mm, 1.8 μm, Waters, United States), and identified by using Q-Exactive quadru-pole-Orbitrap mass spectrometer. The mobile phases consisted of 0.1% formic acid in water (A) and acetonitrile (B) with the elution gradients were as follows: 0–2 min, 5% B, 2–4 min, 5%–30% B, 4–8 min, 30%–50% B, 8–10 min, 50%–80% B, 10–14 min, 80%–100%, 14–15 min, 100% B, 15–15.1 min, 100%–5% B, 15.1–16 min, 5% B. The flow rate was set at 0.35 mL/min, while the temperature of chromatographic column was maintained at 45°C. The mass range was set from m/z 80 to 1,200. Other MS conditions were as follows: Resolution and AGC target were set at 30000 and 1e6, the maximum IT was 50 ms and loop count was set 10. (N) CE/stepped (N)CE was set 10, 20, 40 ev. All data were analyzed by PeakView 2.2 software and Thermo Xcalibur.
+ Open protocol
+ Expand
3

Comprehensive Analytical Instrumentation Suite

Check if the same lab product or an alternative is used in the 5 most similar protocols
Following instruments were included in this study: an ultra-high performance liquid chromatograph (Exion; Sciex Co., Ltd., Framingham, MA, USA), a high-sensitivity mass spectrometer (QTRAP 6500+; Sciex, Framingham, MA, USA), an ultra-performance liquid chromatography (UPLC) column (Acquity UPLC HSS T3, 100 mm × 2.1 mm × 1.8 μm; Waters Co., Ltd., Milford, MA, USA), a desktop colorimeter (CM-5; Minolta Co., Ltd., Osaka, Japan), a spectrophotometer (UV-1800PC UV/Vis; Shanghai Meipuda Instrument Co., Ltd., Shanghai, China), a desktop electric fan-drying oven (HGZF-9053; Shanghai Yuejin Medical Equipment Co., Ltd., Shanghai, China), a 1/10,000 balance (AL204-IC; Mettler Toledo Instruments (Shanghai) Co., Ltd., Shanghai, China), and a refrigerated centrifuge (HR/T20MM; Hunan Hercy Instrument Equipment Co., Ltd., Changsha, China).
+ Open protocol
+ Expand
4

Serum and Drug Metabolites Analysis by UPLC-MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
Metabolites from serum and drug samples were analyzed using ultra-performance liquid chromatography-mass spectrometry (UPLC-MS), under identical chromatographic and mass spectrometric conditions. Chromatographic separation was carried out using ACQUITY HSS T3 (Waters, USA) and ACQUITY UPLC HSS T3 columns (2.1 × 100 mm, 1.8 μm) (Waters, USA). The column temperature was maintained at 35 °C, and metabolites were detected using a mass spectrometer detector Q-Exactive (Thermo Fisher Scientific, USA). For metabolite detection, 5 μl of sample solution was used. The mobile phase consisted of a mixture of solvent A (H2O, containing 0.1% formic acid) and solvent B (CH3CN) with linear gradient elution. The mass spectrometry conditions were set as follows: spray voltage of 3.5/-3.2 kV, 40/45 Sheath gas, 11/10 AUX gas flow, and a capillary temperature of 350 °C. The mass charge ratio (M/Z) scan range was collected from 70 DA to 1000 DA in both positive and negative modes. To ensure system stability and reproducibility, combined QC samples were analyzed every 10 runs when analyzing serum samples.
+ Open protocol
+ Expand
5

Metabolomic Analysis of Samples via LC-MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
Metabolomic analysis was performed on the samples described above by LC-MS according to previously described methods (Xue et al., 2017 (link)) with modifications. The freeze-dried samples were extracted with methanol and an internal standard at a ratio of 500:8:0.1. All samples were ground to a fine powder using a grinding mill at 65 Hz for 90 s. Then, the samples were ultrasonicated for 30 min at 40 KHz and allowed to stand for 1 h at −20°C. The samples were centrifuged at 12000 rpm at 4°C for 15 min, and 200 μL of supernatant was transferred to vials for LC-MS analysis.
The analysis platform was LC-Q/TOF-MS (Agilent, 1290 Infinity LC, 6530 UHD and Accurate-Mass Q-TOF/MS, Agilent Technologies, Santa Clara, CA, United States) with a Waters column (ACQUITY UPLC@HSS T3, 2.5 μm 1002.1 mm). Water with 0.1% formic acid and acetonitrile with 0.1% formic acid were used for mobile phase A and B, respectively; the flow rate was 0.4 mL/min, the injection volume was 4 μL, the automatic injector temperature was 4°C, and the column temperature was 40°C. The capillary voltage was 4 and 3.5 kV in positive and negative mode, respectively; the drying gas flow was 11 L/min, and the gas temperature was 350°C. Centroid data were collected from 100 to 1000 m/z.
+ Open protocol
+ Expand
6

UPLC-MS/MS Metabolomics Profiling

Check if the same lab product or an alternative is used in the 5 most similar protocols
Chromatographic separation was accomplished in an Thermo Ultimate 3000 system equipped with an ACQUITY UPLC® HSS T3 (150×2.1 mm, 1.8 μm, Waters) column maintained at 40°C. The temperature of the autosampler was 8°C. Gradient elution was carried out with 0.1% formic acid in water (C) and 0.1% formic acid in acetonitrile (D) or 5 mM ammonium formate in water (A) and acetonitrile (B) at a flow rate of 0.25 mL/min. Injection of 2 μL of each sample was done after equilibration. An increasing linear gradient of solvent B (v/v) was used as follows: 0~1 min, 2% B/D; 1~9 min, 2%~50% B/D; 9~12 min, 50%~98% B/D; 12~13.5 min, 98% B/D; 13.5~14 min, 98%~2% B/D; 14~20 min, 2% D-positive model (14~17 min, 2% B-negative model).
The ESI-MSn experiments were executed on the Thermo Q Exactive Focus mass spectrometer with a spray voltage of 3.5 kV and -2.5 kV in positive and negative modes, respectively. Sheath gas and auxiliary gas were set at 30 and 10 arbitrary units, respectively. The capillary temperature was 325°C. The orbitrap analyzer scanned over a mass range of m/z 81-1 000 for a full scan at a mass resolution of 70,000. Data-dependent acquisition (DDA) MS/MS experiments were performed with a HCD scan. The normalized collision energy was 30 eV. Dynamic exclusion was implemented to remove some unnecessary information in MS/MS spectra.
+ Open protocol
+ Expand
7

Quantitative Detection of Urinary VOCs

Check if the same lab product or an alternative is used in the 5 most similar protocols
The quantitative detection of VOCs in human urine was performed using ultra performance liquid chromatography coupled with electrospray tandem mass spectrometry (UPLC-ESI/MSMS) as described in previous studies (20 (link)). The Acquity UPLC® HSS T3 (Part no. 186003540, 1.8 µm x 2.1 mm x 150 mm, Waters Inc.) column, with 15 mM ammonium acetate and acetonitrile as the mobile phases, was applied for chromatographic separation. In brief, the eluent from the column is firstly ionized using an electrospray interface to generate and transmit negative ions into the mass spectrometer. Then the comparison of relative response factors (ratio of native analyte to stable isotope labeled internal standard) with known standard concentrations yields individual analyte concentrations (https://wwwn.cdc.gov/nchs/data/nhanes/2015-2016/labmethods/UVOC_UVOCS_I_MET.pdf). More details of the laboratory method detecting urine VOCs could be found in the website (https://wwwn.cdc.gov/nchs/data/nhanes/2015-2016/labmethods/UVOC_UVOCS_I_MET.pdf).
+ Open protocol
+ Expand
8

UPLC-MS Gradient Separation Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
Chromatographic separation was accomplished in a Thermo Vanquish system equipped with an ACQUITY UPLC® HSS T3 (150×2.1 mm, 1.8 µm, Waters) column maintained at 40 °C. The temperature of the autosampler was 8 °C. Gradient elution of analysis was carried out with 0.1% formic acid in water (A2) and 0.1% formic acid in acetonitrile (B2) or 5 mM ammonium formate in water (A3) and acetonitrile (B3) at a flow rate of 0.25 mL/min. Injection of 2 μL of each sample was done after equilibration. An increasing linear gradient of solvent B2/B3 (v/v) was used as follows: 0~1 min, 2% B2/B3; 1~9 min, 2%~50% B2/B3; 9~12 min, 50%~98% B2/B3; 12~13.5 min, 98% B2/B3; 13.5~14 min, 98%~2% B2/B3; 14~20 min, 2% B2-positive model (14~17 min, 2% B3-negative model).
+ Open protocol
+ Expand
9

UPLC-MS/MS Quantification of Oxidized Phospholipids

Check if the same lab product or an alternative is used in the 5 most similar protocols
A wide-targeted analysis was performed using an ACQUITY UPLC system (Waters) coupled with a triple quadrupole MS (QTRAP 6500, Sciex). The LC separation was per- formed using a reverse-phase column [ACQUITY UPLC HSS T3 (50 mm × 2.1 mm inner diameter, 1.8 μm particle size; Waters)] with a gradient elution consisting of mobile phase A (methanol/acetonitrile/water = 1:1:3 v/v/v containing 50 mM ammonium acetate and 10 nM EDTA) and mobile phase B (100% isopropanol containing 50 mM ammonium acetate and 10 nM EDTA). The LC gradient consisted of holding solvent (A/B:100/0) for 1 min, which was linearly converted into solvent A/B:50/50 for 4 min, linearly converted to solvent A/B:36/64 for 7 min, linearly converted to solvent A/B:5/95 for 1 min, and then held for 1 min. It was then returned to solvent A/B:100/0 and held for 5 min for re-equilibration. The injection volume was 3.5 μl, the flow rate was 0.350 mL/min, and the column temperature was 50°C. The multiple reaction monitoring (MRM) mode was used to quantify oxidized phospholipids in biological samples.
+ Open protocol
+ Expand
10

Quantitative Analysis of Purified RPEFs

Check if the same lab product or an alternative is used in the 5 most similar protocols
The purified RPEFs were quantitatively analyzed by using ACQUITY UPLC HSS T3 (50 mm × 3.0 mm, 1.8 μm, Waters Technologies Co., Massachusetts, USA) coupled with Triple-TOF 5600+ Mass Spectrometry (AB SCIEX CO., Framingham, USA). 0.1 % formic acid solution (A) and acetonitrile (B) were adopted as the mobile phase. The extract was filtered through 0.45 μm membrane prior to 4 μL of injection into the system. The flow rate was maintained at 0.5 mL/min and detected at 280 nm UV length. The elution procedure was presented as following: 0–10 min, 5 % to 40 % B; 10–17 min, 40 %-95 % B; 17–25 min, 95 %-5% B. The MS conditions was set as positive/negative ion sweeping mode: scan range 100–1500 m/z, ion source gas 1 and 2 (air) 55 psi, curtain gas (N2) 35 psi, ion source temperature 600℃ (positive) and 550℃ (negative), source voltage −5.5 kv (positive) and −4.5 kv (negative), collision energy spread 40 ± 20 eV. Data was processed using Analyst software version 4.1 (MassLynx).
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!