The largest database of trusted experimental protocols

Raman microscope

Manufactured by Horiba
Sourced in France, Japan

The Raman microscope is an analytical instrument used for the characterization of materials. It provides a non-destructive method to identify and analyze the molecular composition of a sample by detecting the inelastic scattering of monochromatic light, typically from a laser source. The Raman microscope combines the capabilities of a Raman spectrometer with a high-magnification optical microscope, allowing for the examination of microscopic samples.

Automatically generated - may contain errors

11 protocols using raman microscope

1

Multimodal Characterization of Nanomaterials

Check if the same lab product or an alternative is used in the 5 most similar protocols
AFM images were obtained using a Bruker Dimension FastScan AFM (Billerica, MA, USA) in knockdown mode. Optical images were obtained using a Sunwoo RX50M microscope (Yuyao, China). Transmission electron microscopy was performed with a 200 kV Talos F200S transmission electron microscope. Room-temperature Raman and PL spectra were recorded with a Horiba Raman microscope (Irvine, CA, USA) with 532 nm laser excitation. The X-ray photoelectron spectroscopy (XPS) of the samples was carried out using a Thermo Scientific Kα XPS spectrometer (Waltham, MA, USA) equipped with a monochromatic Al-Kα X-ray source.
+ Open protocol
+ Expand
2

Raman Microscopy of Pretreated Wood

Check if the same lab product or an alternative is used in the 5 most similar protocols
The raw and pretreated wood sections were placed on a glass slide with a drop of D2O and then covered with a glass coverslip of 0.17 mm thickness for Raman detection. A LabRam Xplora exquisite full-automatic confocal Raman microscope (Horiba Jobin Yvon, Longjumeau, France) equipped with an MPlan 100× oil immersion microscope objective (Olympus, NA = 1.40) was utilized in this study. A linear polarized laser (diode-pumped green laser, λ = 532 nm), focused with a diffraction-limited spot size (0.61 λ/NA), was used to conduct measurements. The laser power on the sample was approximately 8 mW. The Raman light was detected by an air-cooled, front-illuminated, spectroscopic charge-coupled device (CCD) behind a grating (1200 grooves·mm−1) spectrometer. For mapping, an integration time of 2 s was chosen and every pixel corresponds to one scan with a spectrum acquired every 0.5 µm by averaging 2 s cycles. Labspec 5 software (Horiba Jobin Yvon, Longjumeau, France) was used for spectra analysis and image processing.
+ Open protocol
+ Expand
3

Raman Spectroscopy of Single Cells

Check if the same lab product or an alternative is used in the 5 most similar protocols
Fixed cell pellets were washed two times with pure water, and then resuspended into water to be cell solution to avoid influence from salt crystals after drying. The cell solution containing 5k cells was added dropwise on a glass slide. After air dry, glass slides with cells were then used to take Raman spectra. Spontaneous Raman spectra were acquired using an upright confocal Raman spectrometer (Horiba Raman microscope; Xplora plus). A 532 nm YAG laser is used to illuminate the sample with a power of 12 mW on sample through a 100 x, N.A. 0.9 objective (MPLAN N; Olympus) with slit 100 μm and hole 500 μm. Spectro/Raman shift center was set to be 2000.04 cm−1. With a 1200 grating (750 nm), Raman shift ranges from 690.81 cm−1 to 3141.49 cm−1 was acquired to cover whole cellular Raman peaks. Acquisition time for one spectrum was set to be 5 s, and then averaged for 5 times. The target cell was chosen randomly and spectra of five points (center, top, bottom, left, right) on the cell were acquired. The acquired spectra were processed by the LabSpec 6 software for baseline correction.
+ Open protocol
+ Expand
4

Raman Spectroscopy of HAP Crystallization

Check if the same lab product or an alternative is used in the 5 most similar protocols
Raman
spectroscopy was used to investigate the mineral phase transformation
during HAP crystallization in samples with and without the peptides.
The concentrations of the peptides were 0.2 mg/mL at pH 7.4. The concentrations
of calcium and phosphate were 1.5 and 9.5 mM, respectively. The Raman
spectra were collected continuously up to 3 h, from 100 to 4000 cm–1 under backscattering geometry using a Raman microscope
(HORIBA Scientific, Japan, equipped with LabRAM ARAMIS) operated at
a resolution of 1 cm–1 with an excitation wavelength
of 532 nm and laser power of 2.5 mW. A 60× objective with numerical
aperture of 0.75 was used to focus the sample and to collect the spectra
for 20 s.
+ Open protocol
+ Expand
5

Raman Spectroscopy of Analytes

Check if the same lab product or an alternative is used in the 5 most similar protocols
Sample (prepared in the following method session) was sandwiched between a polylysine-coated cover slide and a cover glass. A HORIBA Raman microscope was used to acquire the Raman spectra of analytes. Excitation laser was 532 nm with a power on the sample ~8 mW. Spectrum acquisition time for each Raman spectrum was 30 s.
+ Open protocol
+ Expand
6

Comprehensive Material Characterization Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
Morphological analysis was carried out on a SUPRA 55vp ZEISS field emission scanning electron microscopy (Oberkochen, Germany), while element mapping were obtained using a corresponding energy dispersive X-ray (EDX) spectra. Elemental analysis was carried out on Elementar vario MICRO cube elemental analyzer (Frankfurt, Germany). Carbon, hydrogen, nitrogen, and sulfur content (% CHNS) was determined by combustible analysis using a thermal conductivity detector. Differential scanning calorimetry (DSC), thermogravimetry analysis (TGA), and derivative thermogravimetry (DTG) were heated under nitrogen to 800 °C at a heating rate of 10 °C min−1 using a simultaneous thermal analyzer Setaram Labsys Evolution (Lyons, France). Fourier-transform infrared spectroscopy (FT-IR) was performed using a Bruker VERTEX-70 FT-IR spectroscopy (Ettlingen, Germany), between 650 cm−1 to 4000 cm−1. Raman spectra was acquired using a Horiba Scientific Raman microscope (Paris, France) at an excitation laser wavelength of 532 nm with a 50X objective.
+ Open protocol
+ Expand
7

Raman Analysis of Femur Cortical Bone

Check if the same lab product or an alternative is used in the 5 most similar protocols
After the three-point bending test, the broken left femurs were collected. A segment of cortical bone of the femur along the horizontal plane was excised. After alcohol dehydration, the embedding process in epoxy resin and grinding, the samples for Raman spectroscopy (RS) were obtained. A Raman microscope (Horiba LabRAM, France) with a 785-nm wavelength laser source was adopted to obtain the spectra of femur cortical tissue. The laser was focused on a cross section of the femur cortex with a 50× magnification objective. The average of spectra obtained from three different sites of the femur cortex were employed to reflect the composition of the bone matrix. Curve fitting of Raman spectra was performed with Origin analysis software (OriginLab, USA), and the band area was calculated. The definition and calculation of the main Raman spectra-based compositional parameters in this study are described below: mineral-to-matrix ratio (area of v1PO43− [960 cm−1] band: area of Amide I [1660 cm−1] band), carbonate to phosphate ratio (area of v1CO32− [1060 cm−1] band: area of v1PO43− [960 cm−1] band), mineral crystallinity (inverse of v1PO43− [960 cm−1] band width at half-maximum intensity).
+ Open protocol
+ Expand
8

Quantifying Sulfated Fucan Composition

Check if the same lab product or an alternative is used in the 5 most similar protocols
The sulfate content of the ultrapurified sulfated fucan Fuc1 was externally analyzed using an Agilent 7500 Series quadrupole ICP-MS. The sulfate content of the hydrolyzed sulfated fucans was analyzed by Raman spectroscopy as previously described [12 (link)]. Raman spectra were recorded at room temperature using a Horiba Jobin–Yvon LabRAM HR system equipped with a Raman microscope (HORIBA Ltd., Kyoto, Japan).
+ Open protocol
+ Expand
9

High Pressure Raman Spectroscopy of H2O-CH3OH Mixtures

Check if the same lab product or an alternative is used in the 5 most similar protocols
Distilled H2O mixed with specific wt% of CH3OH was loaded, together with a ruby ball, into a symmetric diamond-anvil cell (DAC) with a culet size of 500 μm. The mixture itself served as a pressure transmitting medium. The pressure was determined by ruby fluorescence43 and the uncertainties of the pressure measurements were typically ~0.1 GPa. The compression and decompression rates of the samples were typically ~0.1 GPa s−1. After the pressure within the DAC equilibrated and reached a stable value, the high pressure Raman spectra of H2O-CH3OH mixtures were measured at room temperature using a Raman microscope (Horiba Jobin Yvon) that employs a CW 514.5 nm Ar-ion laser to excite the mixture within the DAC and probes the Raman scattering and vibrational frequency shifts with a spectral resolution of ~2 cm−1.
+ Open protocol
+ Expand
10

Characterization of Multilayer Composite Film

Check if the same lab product or an alternative is used in the 5 most similar protocols
The cross-section of the film was observed using a Zeiss high-resolution field emission scanning electron microscope to explore the interface bonding ability of the three layers. Raman spectra of CNT-PDMS and pure PDMS films were obtained using a HORIBA Raman microscope with an excitation wavelength of 532 nm. The near-infrared light source used was a semiconductor laser integrated light source with a wavelength of 808 nm provided by Hite Optoelectronics Co., Ltd. (Beijing, China), and its power density was measured by a TP100 optical power meter. The surface temperature distribution and maximum temperature of the composite were measured and recorded by a T650sc infrared imager and paperless recorder. The bending deformation behavior of the film was recorded using an industrial camera, and the bending performance of the film was analyzed using the curvature calculation formula. The thermal expansion coefficient (CTE), thermal conductivity coefficient, and elastic modulus of each single layer film were measured by a thermomechanical analyzer (TMA), thermal conductivity meter, and intelligent electronic tension machine, respectively.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!