The largest database of trusted experimental protocols

Bio amp fe231

Manufactured by ADInstruments
Sourced in Australia, United Kingdom

The Bio Amp FE231 is a high-performance biological amplifier designed for recording and analyzing a variety of biological signals. It features low-noise, high-gain circuitry and supports multiple input channels to accommodate different types of electrodes and transducers.

Automatically generated - may contain errors

8 protocols using bio amp fe231

1

Perinatal ECG Monitoring in Neonates

Check if the same lab product or an alternative is used in the 5 most similar protocols
On the day of birth, neonates were briefly separated from the mother to record
the 30-min period at rest, and then they were euthanized. The animals were kept
unsedated and allowed to rest on a heated surface (V831, SonoBel, Brazil),
restricted only by the weight of ECG electrodes. Adhesive disposable ECG
electrodes (ML02 MedLevensohn, Shanghai INTCO Electrode Manufacturing Co., Ltd.,
China) were cut to reduce their sizes, removing excess sponge, and were adapted
to provide more comfort to the animal, while the part with the silver sensor
(Ag/AgCl) was gently placed on the skin of the newborn (Figure 1A). Three Micro-Hook electrodes (1.5 mm Socket,
ADInstruments Ltd., New Zealand) were connected to the sensors of the electrodes
placed on the right and left limbs of the newborn, as negative and positive
poles, respectively, as well as on the back of the animal, as a neutral wire. RR
intervals were recorded over a 30-min period using an ECG system (BioAmp FE231,
ADInstruments Ltd.) and LabChart software (version 8, ADInstruments Ltd.). The
sampling rate was 4000 Hz.
+ Open protocol
+ Expand
2

Graphene-Based EMG-Controlled Robotic Prosthetics

Check if the same lab product or an alternative is used in the 5 most similar protocols
The alginate-coated SHP–graphene composite was fixed to the skin using Tegaderm film (3M, Maplewood, MN, USA) for EMG measurement. The signals were recorded using a bio-signal amplifier (Bio Amp FE231, AD Instruments, Dunedin, New Zealand) and data acquisition device (PowerLab 8/35, AD Instruments). Action potential signals were filtered on the authority of the ISEK (International Society of Electrophysiology and Kinesiology) standard (1500-Hz low-pass filter). We used LabChart 8 Pro (AD Instruments, Bella Vista, New South Wales, Australia) software to obtain all data. The obtained data were transferred to a conventional microcontroller (Arduino Mega 2560) that moves the robot’s hand (DFRobut, Shanghai, China). The authors obtained Institutional Review Board (IRB) approval (NO. SKKU 2022-07-035) from Sungkyunkwan University for measuring EMG signals.
+ Open protocol
+ Expand
3

Multimodal Monitoring of Vital Signs

Check if the same lab product or an alternative is used in the 5 most similar protocols
Blood pressure, respiratory rate, and heart rate were monitored using an electrophysiological data acquisition system (PowerLab 16/35, AD Instruments, Castle Hill, Australia). Mean blood pressure was measured at the carotid artery with a 24G catheter using a reusable BP transducer (MLT0380/D, AD Instruments, Castle Hill, Australia). The carotid artery was also used to collect blood samples and to create a haemorrhage. The tail vein was cannulated for intravenous fluid infusion. Respiratory rate and heart rate were measured non-invasively and recorded (Figure 1). The respiration rate was calculated using PowerLab-connected pulse transducers (TN1012/ST, AD Instruments, Castle Hill, Australia). Heart rate was measured using PowerLab-connected biological amplifiers (Bio Amp FE231, AD Instruments, Castle Hill, Australia).
+ Open protocol
+ Expand
4

Evaluation of Conductive Hydrogel as IE Platform

Check if the same lab product or an alternative is used in the 5 most similar protocols
To verify the performance of Alg-BA/OEGCG/NaCl conductive hydrogel as an IE platform, an ex vivo muscle defect model was set up using rat hind limb muscles (12-week-old, male) labeled as ‘distal muscle’ and ‘proximal muscle’. Two muscle tissues were freshly dissected and spaced with a distance of 4 mm and the hydrogel was filled in the space. One proximal muscle was electrically stimulated by waveform generator (Arbitrary, 1 Channel, 20 MHz, 33511B, Keysight Technologies, Inc., Santa Rosa, CA, USA). The EMG signals were recorded in another distal muscle by biosignal amplifier (Bio Amp FE231, AD instruments, Sydney, Australia) and the data acquisition device (PowerLab 8/35, AD instruments). Action potential signals were filtered on the authority of the ISEK (International Society of Electrophysiology and Kinesiology) standard (1500 Hz low-pass). As a control, a direct contact muscle-to-muscle model was prepared without a space between two muscles.
+ Open protocol
+ Expand
5

Cardiovascular Monitoring in Anesthetized Rats

Check if the same lab product or an alternative is used in the 5 most similar protocols
After induction of anesthesia, rats were placed in a supine position, and cardiovascular monitoring consisting of pulse oximetry for SpO2 measurement, ECG and noninvasive measurement of NIBP was established.
Pulse oximetry was placed at the rodent’s paw by means of a commercially available infrared system (Radical-7, Masimo, Puchheim, Germany). A 3-channel ECG according to Eindhoven/Goldberger was derived using needle electrodes and a differential amplifier (BioAmp FE231, ADInstruments Ltd., Oxford, UK). In addition, NIBP measurement was performed using a commercially available device (IN125/R, ADInstruments Ltd., Oxford, UK) consisting of an inflatable tail cuff system in combination with a pulse sensor.
All parameters were measured and recorded during the entire time of surgery and stored for further analysis using Labchart (ADInstruments Ltd., Oxford, UK).
+ Open protocol
+ Expand
6

Multiparameter Biometrics in Rodents

Check if the same lab product or an alternative is used in the 5 most similar protocols
Hemoglobin oxygen saturation, respiratory rate per minute, heart rate, and core temperature were measured noninvasively and recorded. Blood pressure, respiratory rate, and heart rate were monitored using an electrophysiological data acquisition system (PowerLab 16/35, AD Instruments, Castle Hill, Australia). Blood pressure was measured at the carotid artery with a 24-catheter employing a reusable BP transducer (MLT0380/D, AD Instruments, Castle Hill, Australia). The carotid artery was also used to collect blood samples and to generate hemorrhage. The tail vein of rats was cannulated for IV fluid infusion. Respiratory rate and heart rate were measured non-invasively and recorded. The respiration rate was measured using PowerLab-connected pulse transducers (TN1012/ST, AD Instruments, Castle Hill, Australia). Heart rate was measured using PowerLab-connected biological amplifiers (Bio Amp FE231, AD Instruments, Castle Hill, Australia).
+ Open protocol
+ Expand
7

Concurrent Validity and Reliability of Cyton Biosensing Board

Check if the same lab product or an alternative is used in the 5 most similar protocols
The current investigation consists of two different data collections. The first examined the concurrent validity between the Cyton Biosensing Board with three-lead ECG (OpenBCI, Brooklyn, New York, USA) (herein referred to as cyton board) and an industry-standard ECG system (ADInstruments, Bio Amp FE231, Colorado Spring, Colorado, USA). The second quantified the between-day reliability of the cyton board. Prior to the commencement of both studies, written informed consent was collected from all participants. Ethical approval for the investigation was given by the University of Calgary Conjoint Health Research Ethics Board (REB15-1376).
+ Open protocol
+ Expand
8

Comprehensive Vital Sign Monitoring in Rodents

Check if the same lab product or an alternative is used in the 5 most similar protocols
Following induction of anesthesia, rats were positioned supine and monitoring was established, consisting of pulse oximetry for blood oxygen saturation (SpO2) measurement, electrocardiography (ECG), and noninvasive measurement of blood pressure (NIBP). Pulse oximetry was applied to the paw of the rodents using a commercially available infrared system (Radical-7, Masimo, Puchheim, Germany). A 3-lead Eindhoven/Goldberger ECG was derived using needle electrodes and a differential amplifier (BioAmp FE231, ADInstruments Ltd., Oxford, UK). In addition, NIBP was measured during first and second surgery with a commercially available device (IN125/R, ADInstruments Ltd., Oxford, UK) using an inflatable tail cuff in combination with a pulse sensor, to detect hypotension, as a consequence of hypovolemia or very deep anesthesia. Maintenance of adequate body temperature was ensured by a feedback-controlled heating pad (TCAT-2LV controller, Physitemp, Clifton, NJ, USA) with a rectal probe.
All parameters were measured and recorded throughout every procedure and stored for further analysis using Labchart (ADInstruments Ltd., Oxford, UK).
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!