The largest database of trusted experimental protocols

12 protocols using 1200 series autosampler

1

HPLC Analysis of High-Purity Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
The HPLC instrument consisted of an Agilent Technologies series 1200 autosampler, series 1200 UV/Vis detector, series 1100 pump, using ChemStation software (Agilent Technologies, Santa Clara, CA, USA). The analytical column was a reverse-phase Waters Nova-pak® C18 150 mm × 3.9 mm column. A gradient elution was used (flow rate 0.2 mL/min), starting with 80% water and progressing to 100% acetonitrile over a period of 1 h, with both solvents containing 0.1% formic acid. All compounds have purity ≥95% (254 nm) by HPLC.
+ Open protocol
+ Expand
2

HPLC Analysis of Tissue Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
HPLC apparatus involved Agilent Series 1200 quaternary gradient pump, Series 1200 autosampler, Series 1200 UV-Vis detector, and HPLC 2D-Chemstation software. The chromatographic column was a reversed-phase column (C18, 4.6 mm, 250 mm, 5 µm, Agilent Co.). Solid-phase extraction (SPE) cartridges (Bond Elut C18, 500 mg/3 mL) were used to clarify tissue matrixes.
+ Open protocol
+ Expand
3

HPLC-UV Analysis of ETO Quantification

Check if the same lab product or an alternative is used in the 5 most similar protocols
ETO concentrations in samples were measured by HPLC-UV (Hewlett Packard series 1100 HPLC Pump combined with Agilent Technologies 1200 Series Autosampler and a ChemStation software (Agilent Technologies, B.04.03 version, Santa Clara, CA, USA). Injection volume was 50 µL and UV detection wavelength was 254 nm (Agilent 1100 Series UV lamp). Linearity was demonstrated between 50 µM and 500 µM with calibrated curves prepared in mobile phase from a ETO methanol stock solution (7 mM). Limit of detection (LOD) of and limit of quantification (LOQ) were 2.2 µM and 6.6 µM, respectively. The mobile phase was a mixture of 20:80 (v/v) water:acetonitrile both containing 0.1% (v/v) trifluoroacetic acid. After 2.3 min, the peak signal of ETO was observed using a reversed-phase C-18 column (Bondapak 250 mm × 4.6 mm—5 micron) and a 1 mL/min flow rate.
+ Open protocol
+ Expand
4

Polymer Analysis via SEC

Check if the same lab product or an alternative is used in the 5 most similar protocols
Polymer samples were dissolved
in HPLC-grade tetrahydrofuran (THF)
(∼1 mg/mL), filtered through alumina to remove the Cu catalyst,
and then filtered through a 0.45 μm PTFE filter. Size exclusion
chromatography was then performed on a PL-Gel 300 × 7.5 mm (5
μm) mixed C column using an Agilent 1200 series autosampler,
an inline degasser, and a diode array detector set to monitor the
absorbance at 254 nm. The column and detector temperatures were 35
°C. HPLC-grade THF was used as the eluent (1 mL/min). Linear
polystyrene standards were used for calibration.
+ Open protocol
+ Expand
5

HPLC-UV Quantification of DKT

Check if the same lab product or an alternative is used in the 5 most similar protocols
DKT concentrations in the samples were measured by HPLC-UV (Hewlett Packard series 1100 HPLC Pump combined with Agilent Technologies 1200 Series Autosampler). A volume of 75 µL was injected into the HPLC system (Waters 515 HPLC Pump with Waters 717 Autosampler). DKT was detected with an UV lamp at 262 nm (Water 996 Photodiode Array Detector). The mobile phase consisted of 60:40 mixture of acetonitrile and purified water A (both containing 0.1% TFA). Stationary phase was a C-18 Agilent Eclipse XDB (4.6 × 150 mm; 3.5 µm). Elution flow was 1 mL/min and retention time for DKT was 3.95 min. Calibration curves were made in mobile phase based on a stock solution of DKT in methanol. Linearity was observed between 1.5 µg/mL and 300 µg/mL covering all the experimental sample values. The observed peaks were integrated using Millenium software (Agilent Technologies, County of Santa Clara, CA, USA). The developed analytical method met the standards for precision and accuracy.
+ Open protocol
+ Expand
6

HPLC-UV Analysis of Loratadine

Check if the same lab product or an alternative is used in the 5 most similar protocols
All samples derived from the GIS studies were analyzed for loratadine by HPLC-UV (Hewlett Packard series 1100 HPLC Pump combined with Agilent Technologies 1200 Series Autosampler (Santa Clara, CA, USA). A volume of 100 µL was injected into the HPLC system connected to a UV-lamp that was able to detect loratadine at a wavelength of 248 nm (Agilent 1100 Series UV lamp). An isocratic run with a mixture of 70% acetonitrile and 30% purified water was used to detect loratadine using a C-18 column (Kinetex C18 HPLC column, 250 × 4.60 mm‒5 micron, Phenomenex, Golden, CO, USA) and a 1 mL/min flow rate. Calibration curves were made in methanol, based on a stock solution of loratadine in methanol (0.1 mg/mL). Linearity was observed between 40 µg/mL and 0.156 µg/mL with a regression coefficient of at least 0.995 between the AUC of the obtained peaks versus the spiked concentrations. The peaks were integrated using ChemStation software (Agilent Technologies, B.04.03 version, Santa Clara, CA, USA).
+ Open protocol
+ Expand
7

Fab Molecular Weight Determination by ESI-MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
Molecular weights of Fab fractions were determined by electrospray ionization mass spectrometry (ESI-MS) conducted at the Center for Metabolomics and Mass Spectrometry. Samples were run through an Agilent PLRP-S 100 Å 5 µm column, number 0006140735-10. A column guard was used to protect the column from overloading and from particulates. The autosampler used was an Agilent Technologies 1200 series autosampler, and the instrument an Agilent Technologies 6230 TOF LC/MS with a Dual AJS ESI ion source. The LC/MS gradient makes use of two solvents: Solvent A being 0.1% Formic Acid in H2O, and Solvent B being 0.1% Formic Acid in ACN. The gradient consisted of three steps: the first flowing 95% Solvent A and 5% B for 5 min, the second ramping up the concentration of B to 90% for 10 more min, and finally returning the concentration of B to 5%, flowing for another 1 min. The gradient was run at a flow of 300 µL/min at a pressure of 400 bar (40 MPa).
+ Open protocol
+ Expand
8

HPLC-UV Analysis of Ibuprofen Solubility

Check if the same lab product or an alternative is used in the 5 most similar protocols
Solubility samples were analyzed by HPLC–UV (Hewlett Packard series 1100 HPLC Pump, Santa Clara, CA, USA), combined with Agilent Technologies 1200 Series Autosampler (Santa Clara, CA, USA). A volume of 5 µL was injected into the HPLC system connected to a UV lamp that was able to detect ibuprofen at a wavelength of 220 nm (Agilent 1100 Series UV lamp, Santa Clara, CA, USA). An isocratic run containing 70% acetonitrile (VWR International, West Chester, PA, USA) and 30% purified water (both containing 0.1% TFA) was used to detect ibuprofen at a retention time of 2.9 min using a reversed-phase C-18 column (Eclipse Plus C18, 4.6 × 150 mm, 5.5 µm, Agilent Technologies) and a 1 mL/min flow rate. The calibration curve was made in methanol based on a stock solution of ibuprofen in methanol (1 mM). Linearity was observed between 10.32 µg/mL and 0.32 µg/mL. The observed peaks were integrated using ChemStation software (Agilent Technologies, B.04.03 version). The developed analytical method met the FDA requirements for bioanalytical method validation [21 ].
+ Open protocol
+ Expand
9

Posaconazole Quantification in GIS Samples

Check if the same lab product or an alternative is used in the 5 most similar protocols
All samples derived from the GIS studies were analyzed for posaconazole by HPLC-UV (Hewlett Packard series 1100 HPLC Pump combined with Agilent Technologies 1200 Series Autosampler). A volume of 100 μL was injected into the HPLC system connected to a UV-lamp that was able to detect posaconazole at a wavelength of 254 nm (Agilent 1100 Series UV lamp). A gradient run with a mixture of acetonitrile and purified water (both containing 0.1% TFA) was used to detect posaconazole at a retention time of 8.9 min using a C-18 column (Kinetex C18 HPLC column, 250 × 4.60 mm -5 μm, Phenomenex, Golden, CO) and a 1 mL/min flow rate. Calibration curves were made in mobile phase based on a stock solution of posaconazole in methanol (7 mM). Linearity was observed between 50 μM and 39 nM. The observed peaks were integrated using ChemStation software (Agilent Technologies). The developed analytical method met the FDA requirements for bioanalytical method validation (Food and Drug Administration, 2001).
+ Open protocol
+ Expand
10

HPLC-ESI-MS for Compound Identification

Check if the same lab product or an alternative is used in the 5 most similar protocols
For qualification and peak identification purposes, an HPLC-ESI-MS system consisting of a binary HPLC pump (1100 series) and autosampler (1200 series) from Agilent Technologies (Waldbronn, Germany) equipped with an LC-ESI-MS/MS ion-trap system (HCT Ultra ETD II, Bruker Daltonics, Bremen, Germany) was used. Mass spectra were recorded in the negative ionization mode with the capillary voltage set at 3500 V, end plate −500 V, and capillary exit −115.0 V. Drying gas was nitrogen at 330 °C, and 10.0 L/min flow rate with nebulizer pressure of 50 psi, target mass setting m/z 350, scan range from m/z 100–2000 in Ultra Scan mode, fragmentation amplitude 1 V. Compass Hystar Software (version V. 3.2, Bruker Daltonics) was used for analysis and data collection. HPLC separation was adapted from Kiene and co-workers and performed on a C18-column (Aqua 3u, 100 Å, 3 μm, 150 mm × 2.0 mm i.d.) from Phenomenex (Aschaffenburg, Germany) with a guard column of the same material at a flow rate of 0.20 mL/min [35 (link)]. The mobile phase consisted of 2% aqueous acetic acid (v/v) (A) and acetonitrile (B). HPLC conditions for ESI-MS/MS analysis were 0 min (20% B), 20 min (30% B), 40 min (50% B), 50 min (80% B), 55 min (80% B), 60 min (20% B), 70 min (20% B).
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!