The largest database of trusted experimental protocols

3t mr system

Manufactured by Philips
Sourced in Netherlands

The 3T MR system is a magnetic resonance imaging (MRI) scanner that operates at a magnetic field strength of 3 Tesla. It is designed to capture high-quality images of the human body for diagnostic and research purposes. The system uses strong magnetic fields and radio waves to generate detailed images of internal structures, allowing healthcare professionals to assess and monitor various medical conditions.

Automatically generated - may contain errors

24 protocols using 3t mr system

1

T1-weighted MRI Acquisition Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
Imaging data were acquired using Philips 3T MR systems (Philips, Best, The Netherlands) located at Guangzhou Psychiatric Hospital, Affiliated Hospital of Guangzhou Medical University. For each subject, an anatomical image was obtained using a sagittal 3-D gradient-echo T1-weighted sequence (TR = 7.6 ms, TED = 3.7 ms, TI = 795 ms, flip angle = 8°, 180 slices, slice thickness = 1 mm, gap = 0 mm, matrix = 256 × 256, and inversion time = 0).
+ Open protocol
+ Expand
2

Gradient Impulse Response Measurement for MRI

Check if the same lab product or an alternative is used in the 5 most similar protocols
The methods described in this paper have been tested using 2 separate Philips 3T MR systems (both Achieva). Scanner 1 has an experimental 8‐channel body coil30 with scan software capable of GIRF measurement using an image‐based procedure similar to.31 Scanner 2 is a standard clinical system with 2‐port birdcage transmitter with software implementation of SMS imaging techniques. Since neither system had both capabilities, SMS imaging was performed by scanner 2 using pulses corrected for the GIRF measured from scanner 1. Though not ideal, prior experience suggests that the gradient performance of both systems is similar17; errors that may arise from assuming this equivalence will be discussed later.
Figure 1 depicts the idealized (Hi) and the measured transfer functions along the y (Hy) and z (Hz) direction with frequency resolution δf=76.3 Hz. The 2 measured GIRFs both have low‐pass characteristics but differ in cutoff frequency (4200 Hz for Hy and 3750 Hz for Hz) where the magnitude response is reduced by a factor of 1/2. Moreover, the phase deviation potentially leads to a varying phase shift of different frequency components. The GIRF data and details about its application are publicly available (https://github.com/mriphysics/reVERSE-GIRF).
+ Open protocol
+ Expand
3

Resting-State fMRI Acquisition Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
Philips 3T MR systems (Philips, Best, The Netherlands) were used to acquire imaging data. Subjects were asked to wear earplugs and positioned in the coil for support. Participants were requested to lie still with their eyes closed but to remain awake, and all participants reported that they did not fall asleep during the whole experiment. A total of 180 volumes of echo planar images were obtained axially (repetition time, 3,000 ms; echo time, 30 ms; slices, 36; thickness, 4 mm, no slice gap; field of view, 240 × 240 mm2; resolution, 64 × 64; flip angle, 90°).
+ Open protocol
+ Expand
4

3T MRI Structural Brain Imaging

Check if the same lab product or an alternative is used in the 5 most similar protocols
Imaging data were acquired using Philips 3T MR systems (Philips, Best, The Netherlands) located at Guangzhou Brain Hospital. For each subject, an anatomical image was obtained using a sagittal 3-dimensional gradient-echo T1-weighted sequence (TR = 7.6 ms, TED = 3.7 ms, TI = 795 ms, flip angle = 8°, 180 slices, slice thickness = 1 mm, Gap = 0 mm, matrix = 256 × 256, inversion time = 0).
+ Open protocol
+ Expand
5

Resting-state fMRI of Healthy Individuals

Check if the same lab product or an alternative is used in the 5 most similar protocols
Imaging data were acquired using Philips 3 T MR- systems (Best, the Netherlands) located at the LUMC, AMC, and UMCG, equipped with a SENSE-8 (LUMC and UMCG) and a SENSE-6 (AMC) channel head coil respectively. Resting-state fMRI (RS-fMRI) data were acquired using a T2-weighted gradient echo echo-planar imaging with the following scan parameters in Amsterdam and Leiden: 200 whole-brain volumes; repetition time (TR) 2300 ms; echo time 30 ms; flip angle 801; 35 transverse slices; no slice gap; field of view 220 × 220 mm; in-plane voxel size 2.3 × 2.3 mm; slice thickness 3 mm; duration 7.51 min. Parameters in Groningen were identical, apart from: echo time 28 ms; 39 transverse slices; in-plane voxel size 3.45 × 3.45 mm. A sagittal 3-dimensional gradient-echo T1-weighted image was acquired for registration purposes and gray matter analysis with the following scan parameters: repetition time 9 ms; echo time 3.5 ms; flip angle 801; 170 sagittal slices; no slice gap; field of view 256 × 256 mm; 1 mm isotropic voxels; duration 4.5 min. In the darkened MRI room participants were instructed to lie still with their eyes closed and not to fall asleep. Participants confirmed wakefulness after the scanning session. No abnormalities were found upon inspection of the subjects’ structural images by a neuroradiologist.
+ Open protocol
+ Expand
6

Multiparametric MRI Protocol for Comprehensive Tissue Evaluation

Check if the same lab product or an alternative is used in the 5 most similar protocols
The pulse sequences were based on slightly modified standard clinical sequences provided by Philips for the 3-T MR system and included a T1-weighted spin echo (SE) sequence, a T1-weighted turbo field-echo (TFE) sequence, a three-dimensional (3D) multiecho fast field-echo (mFFE) sequence, a multiecho gradient- and spin-echo (mGraSE) sequence, a short tau inversion-recovery (STIR) sequence, a 3D T1-weighted high-resolution isotropic volume excitation (eTHRIVE) sequence, a two-dimensional (2D) MIXED sequence, a T2-weighted turbo SE (TSE) sequence, and a fluid attenuated inversion-recovery (FLAIR) sequence. The sequence parameters are reported in Table 1.

Pulse sequence parameters employed in this study

SequenceRepetition time (ms)Echo time (ms)Flip angle (degrees)
T1-weighted spin-echo600–7001070
T1-weighted turbo field-echo“shortest”“shortest”8
3D multiecho fast field-echo“shortest”4.620
Multiecho gradient- and spin-echo“shortest”“shortest”90
Short tau inversion-recovery5,0003090
T1-weighted high-resolution isotropic volume excitation“shortest”“shortest”10
2D MIXED920 (spin-echo) 2,300 (inversion-recovery)“shortest”90
T2-weighted turbo spin-echo3,0008090
Fluid-attenuated inversion-recovery11,00012590

2D Two-dimensional, 3D Three-dimensional

+ Open protocol
+ Expand
7

3D T1-Weighted MRI Brain Imaging Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
All experiments were conducted on a 3T MR system (Philips Healthcare, Best, The Netherlands). The body coil was used for radiofrequency transmission, and an eight-channel sensitivity encoding (SENSE) head coil was used for receiving (Dai et al., 2008 (link); Aslan and Lu, 2010 (link)). A 3D T1-weighted magnetization-prepared-rapid-acquisition-of-gradient-echo (MPRAGE) scan was performed for anatomical reference and the estimation of brain volume. The MPRAGE sequence used the following imaging parameters: repetition time (TR) of 8.1 ms, echo time (TE) of 3.7 ms, flip angle (FA) of 12°, shot interval of 2100 ms, inversion time (TI) of 1100 ms, voxel size of 1 × 1 × 1 mm3, 160 slices with a sagittal slice orientation, and total scan duration of 3 min 57 s.
+ Open protocol
+ Expand
8

3T MRI Acquisition for Perfusion Imaging

Check if the same lab product or an alternative is used in the 5 most similar protocols
Imaging was performed on a 3T MR system (Philips Healthcare, Best, The Netherlands) with an eight-channel head coil. High-spatial-resolution 3D T1-weighted images were acquired with a turbo field echo sequence in the sagittal plane with the following parameters: repetition time/echo time, 8.21/3.78 msec; inversion time, 1026 msec; flip angle, 8°; effective section thickness, 1.0 mm; slab thickness, 190 mm; matrix, 240 × 240; field of view, 240 × 240 mm; number of signals acquired, 1; scan duration, 5 min 23 sec. All scans resulted in 190 contiguous slices through the brain.
The imaging parameters for the pCASL experiments were as follows: single-shot gradient-echo echo planar imaging in combination with parallel imaging (sensitivity encoding factor 2.0); repetition time/echo time, 4200/8.56 msec; matrix, 64 × 64; field of view, 240 × 240 mm; in-plane resolution, 3.75 × 3.75 mm; 20 slices acquired in ascending order; slice thickness, 6 mm; slice gap, 1 mm; labeling duration, 1650 msec; post-labeling delay, 1525 msec. Forty pairs of control/label images were acquired and then averaged. The total scan duration was 5 min 44 sec. The echo planar imaging M0 images were separately obtained with the same geometry and imaging parameters as the pCASL without labeling.
+ Open protocol
+ Expand
9

Automated Quantification of White Matter Hyperintensities

Check if the same lab product or an alternative is used in the 5 most similar protocols
WMH volumes were derived from axial fluid-attenuated inversion recovery (FLAIR) images obtained on a 3T MR system (Philips Healthcare, Best, The Netherlands) as part of a multi-modal MRI protocol. A previously described, an automated method [32 , 33 (link)] was used to quantify global WMH volumes. Because the distribution of baseline WMH volumes was skewed, all analyses were run using a continuous log(x + 1)-transformed WMH variable and a dichotomized WMH variable (1 = highest quartile considering WMH data across all time points, 0 = lower three quartiles).
+ Open protocol
+ Expand
10

Resting-state fMRI and Anatomical MRI Acquisition

Check if the same lab product or an alternative is used in the 5 most similar protocols
MRIs were acquired using a Philips 3T MR system (Philips, Achieva, the Netherlands) located at Guangzhou Brain Hospital. The participants were instructed to keep their eyes closed, to relax but not fall asleep, and to move as little as possible. For each subject, the rs-fMRI data were collected using an echo-planar imaging (EPI) sequence (64×64×33 matrix with 3.44×3.44×4.6 mm3 spatial resolution, echo time=30 ms, time to repeat=9,000 ms, acquisition time=2,000 ms, flip angle=90°, field of view=211×211 mm2). The structural MRI data were obtained using a sagittal three-dimensional gradient-echo T1-weighted sequence (256×256×188 matrix with the spatial resolution of 1×1×1 mm3, repetition time=8.2 ms, echo time=3.8 ms, flip angle=7°).
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!