The largest database of trusted experimental protocols

Truebeam accelerator

Manufactured by Agilent Technologies
Sourced in United States

The TrueBeam accelerator is a medical device designed for the delivery of radiation therapy. It is used to generate and deliver high-energy ionizing radiation to targeted areas of the body, typically for the treatment of cancer. The TrueBeam accelerator is capable of producing multiple types of radiation beams, including photon and electron beams, to optimize treatment plans for individual patients.

Automatically generated - may contain errors

11 protocols using truebeam accelerator

1

VMAT for Liver SBRT: Dosimetric Objectives

Check if the same lab product or an alternative is used in the 5 most similar protocols
In the original VMAT plans, a total dose of 45 Gy in 3 fractions was prescribed for all patients with the following planning objectives: prescription dose to cover at least 95% of PTV, conformity index (CI) < 1.2, and appropriate dose sparing of the OARs. The normal tissue dose constraints used at our institution for liver SBRT in 3 fractions are presented in Table 2 [26 (link)]. The Eclipse analytical anisotropic algorithm was used for dose calculation with a grid size of 1.25 mm in the VMAT plans. Using the Eclipse planning system, inverse planning for each patient was performed with a 10 MV flattening-filter-free beam of a Varian TrueBeam accelerator equipped with a high-definition 120 MLC, and 2 full arcs (clockwise rotation from 181–179° with the collimator at 30°, and counterclockwise rotation from 179–181° with the collimator at 330°). The mean isodose line prescribed to 45 Gy was 90.6% (range, 87–92%) of the maximum dose.
+ Open protocol
+ Expand
2

Validating TrueBeam SRS Isocentricity

Check if the same lab product or an alternative is used in the 5 most similar protocols
We recently installed a Varian TrueBeam accelerator (Varian Medical Systems, Palo Alto, CA, USA) with SRS capability. This system uses cone beam computed tomography (CBCT) as the method of image guidance combined with Varian's PerfectPitch, six degree of freedom couch. The isocentricity of the combined couch, gantry, and collimator systems was determined to have a 0.59‐mm radius at the time of acceptance. The coincidence of the kV imaging system and MV treatment beam was within 0.3 mm per the Varian IsoLock procedure during this period. Monitoring of patient positioning can be accomplished via an optical surface‐monitoring system, but that system was not employed for this study. The phantom was initially aligned using CBCT and no adjustment was made after each couch rotation.
+ Open protocol
+ Expand
3

Precise Monte Carlo Simulation of Radiation Therapy

Check if the same lab product or an alternative is used in the 5 most similar protocols
The MC simulation code used in this study is the EGSnrc21 code and its user codes BEAMnrc22, 23 and DOSXYZnrc.24 The modulated realistic beams from the Varian TrueBeam accelerator with a Millennium 120 multileaf collimator (MLC) have been simulated by using BEAMnrc/DOSXYZnrc codes and calculated dose distributions have been validated.25, 26, 27Varian TrueBeam phase‐space files28 (version 2.0) are used as the radiation sources at the plane before entering the secondary collimators, or jaws, and MLC. The jaw openings and MLC modulations are modeled in detail in the simulations as described in the study by Lobo and Popescu.27 The typical source phase‐space file for each beam energy used for simulation is about 20 GB in size containing about 900 million particles. The large number of particles used is necessary to achieve a statistical uncertainty of about 1% for MC calculations with a calculation grid size of 2.5 mm, which is the same as in Eclipse calculations. The EXACT boundary crossing algorithm is used and the electron cutoff energy (ECUT) is set at 0.7 MeV.
+ Open protocol
+ Expand
4

sCT-DRR Guided Prostate Radiotherapy

Check if the same lab product or an alternative is used in the 5 most similar protocols
When all QA-steps were completed, the MRI-only treatment plan was approved and passed on to daily treatment on a TrueBeam accelerator (Varian Medical systems, Palo Alto, California, USA). Patients were positioned with corresponding fixation as during MRI and aligned using the patient tattoos. Set-up verification was performed with daily kilo voltage (kV)-image registration. The synthetic GFMs, represented in the sCT-DRR, were manually registered towards the physical GFMs seen in the orthogonal kV-image pairs. Operators were instructed to match the center of the GFM in the orthogonal kV-images to the center of the corresponding synthetic GFM in the sCT-DRR. From the eleventh patient and forward, the cylindrical GFM shape was added to the sCT as a RT structure around the synthetic GFMs. This was an attempt to facilitate easier detection of prostate and GFM rotations.
+ Open protocol
+ Expand
5

Beam data analysis for AXB configuration

Check if the same lab product or an alternative is used in the 5 most similar protocols
The same set of beam data (including percentage depth dose curve, profiles and output factors) used by AAA and measured in a three-dimensional water scanning system (PTW, Germany) for field sizes from 3 × 3 to 40 × 40 cm2 were imported in Eclipse treatment planning system (Version 10.0, Varian Medical Systems, Palo Alto, CA) for the configuration of AXB. All data presented in this study were collected from a commissioned Varian Truebeam™ accelerator equipped with a Millennium 120 multileaf collimator (MLC, with spatial resolution of 5 and 10 mm for the central and outer 20 cm, respectively.
+ Open protocol
+ Expand
6

Skin Dose Evaluation of Eclipse AAA

Check if the same lab product or an alternative is used in the 5 most similar protocols
The commercial TPS evaluated in this study is Varian Eclipse Version 11.0 (Varian Medical Systems, Palo Alto, CA, USA) with the model‐based AAA dose calculation algorithm. The 6, 10, and 15 MV photon beams from a Varian TrueBeam accelerator are used in the study. The heterogeneity correction is employed in all AAA dose calculations; and, unless otherwise specified, the calculation grid size is 2.5 mm, which is a typical size in clinical practice. Although the grid size may affect the skin dose calculation accuracy, switching from 2.5 mm grid size to 1 mm grid size, which is smallest for AAA in Eclipse, only slightly improves the accuracy.20 In addition, the calculation time is much longer with 1 mm grid size, making it clinically unattractive. Furthermore, the grid size for MC dose calculations in this study (see below) is also set at 2.5 mm, which makes the comparison meaningful and justified. To avoid confusion, the skin dose is defined in this study as the mean dose to the skin structure of 5 mm thickness for the CT based dose calculations. To quantify the skin dose, the skin was contoured to be an area of 2 × 2 cm2, corresponding to a volume of about 2 cm3, which is of clinical interest. The skin dose predicted by Eclipse is compared with that of MC calculations which are benchmarked by measurements in phantoms. The term “entrance dose” is used for the phantom measurements.
+ Open protocol
+ Expand
7

VMAT Radiotherapy for Head and Neck Cancer

Check if the same lab product or an alternative is used in the 5 most similar protocols
Patients were treated on a Varian TrueBeam™ accelerator using 6 MV photon beams. The treatment plans were generated using the Acuros dose calculation algorithm [31] (link), [32] (link) in ARIA 15.6 (Varian Medical Systems, Palo Alto, CA, USA). The treatment plans consisted of 2 or 4 VMAT (RapidArc) beams (see Table 1). The dose to the PTV was: D98% > 95 % of prescribed dose. The dose to the parotid glands, submandibular glands, oral cavity and swallow structures was minimised as much as possible. The CTV-PTV margin was 3 mm based on in-house analysis.
+ Open protocol
+ Expand
8

Dosimetry Validation for Radiation Therapy

Check if the same lab product or an alternative is used in the 5 most similar protocols
Dose measurements were performed using gafchromic MD-V3 film, which had been tested in a wide dose range, namely 1–100 Gy [53 (link),54 (link)]. Using the Truebeam accelerator (Varian Medical Systems, Inc., of Palo Alto, CA, USA), we constructed a calibration curve by irradiating films with known doses of 0, 1, 3, 5, 10, 15, 20, 30, 40, and 50 Gy. The linearity of dose and exposure with Monitor Units (MU) were also checked.
CT scans of every setup configuration were acquired with a LightSpeed™ Pro16 (General Electric Medical Systems, Waukesha, WI, USA) and fed to a TPS (Eclipse, Varian Medical Systems, Inc., of Palo Alto, CA, USA) to calculate the MU required for the irradiation of each sample. Gafchromic films were placed under the holder samples (flask or uvette) to check the accuracy of the dose delivery.
+ Open protocol
+ Expand
9

Photon Beam Quality Optimization

Check if the same lab product or an alternative is used in the 5 most similar protocols
On the Eclipse treatment planning system V13.5, the aforementioned RapidPlan model was applied to estimate the best achievable DVHs under five photon beam qualities from a Varian TrueBeam accelerator equipped with Millennium 120 multi-leaf collimator (MLC), including 6-MV flattened (6X), 6-MV flattening-filter-free mode (6F), 8X, 10X and 10F respectively, for 20 historical patients that were not included in the model library. Higher energies are not used at our center for the consideration of secondary neutron contamination [19 ], hence were not tested in this study. Without any human intervention, VMAT plans were optimized using the RapidPlan-generated patient-specific objectives [20 ], keeping the original beam geometries of the clinical plans unchanged. The prescription dose was 41.8Gy for PTV and 50.6Gy for PTVboost. The volume dose was calculated using analytical anisotropic algorithm (AAA). All plans were normalized to cover 95% target volume with 100% prescription dose before comparison.
+ Open protocol
+ Expand
10

Pre-Clinical and Clinical IMRT/VMAT Plan Validation

Check if the same lab product or an alternative is used in the 5 most similar protocols
For pre‐clinical validation, 17 anonymized intensity‐modulated treatment plans were selected for analysis. All plans were generated in Eclipse. Dose was calculated with the AcurosXB algorithm (v11). The treatment machine for all plans was a Varian TrueBeam accelerator with high‐definition (0.25 and 0.5 cm) MLCs. The accelerator is equipped with 6 and 10 MV flattened photon beams. Plans were chosen to encompass a clinically relevant variety. Nine plans used volumetric modulated arcs (VMAT) for delivery. Of the VMAT plans, six used a beam energy of 6 MV, and three used a beam energy of 10 MV. The eight remaining plans utilized sliding window intensity‐modulation (IMRT) at static gantry angles for delivery. Of the IMRT plans, two used 6 MV, three used 10 MV, and three were mixed‐beam (6 and 10 MV). Both IMRT and VMAT plans incorporated tracking of the primary jaws to minimize MLC leakage. Treatment plan characteristics for these 17 cases can be found in Table 1.
For the post‐implementation clinical plan comparison, the first 36 treatment plans evaluated using Mobius were chosen for inclusion in this study. The only requirement for inclusion was intensity modulation; this included IMRT as well as VMAT. Of these 36 clinical cases, 33 used a beam energy of 6 MV, one used a beam energy of 10 MV, and two used mixed energy (6 and 10 MV).
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!