The largest database of trusted experimental protocols

C18 reversed phase column

Manufactured by Waters Corporation
Sourced in United States

The C18 reversed-phase column is a type of chromatographic column used for the separation and analysis of a wide range of organic compounds. It features a stationary phase composed of silica particles chemically bonded with octadecyl (C18) alkyl chains. This column is commonly used in high-performance liquid chromatography (HPLC) and is designed to provide efficient and reproducible separation of analytes based on their hydrophobic interactions with the stationary phase.

Automatically generated - may contain errors

44 protocols using c18 reversed phase column

1

Enzymatic Production of Acryloyl-CoA

Check if the same lab product or an alternative is used in the 5 most similar protocols
PrpE, a propionate-CoA ligase from D. shibae DFL 12, was used to produce acryloyl-CoA. Reaction system of PrpE contained 0.5 mM acrylate, 0.5 mM coenzyme A (CoA), 0.5 mM adenosine triphosphate (ATP), 5 mM MgCl2, 100 mM Tris-HCl (pH 8.0) and 10 μM recombinant PrpE. After a 2-h reaction at 37°C, RdAcuH was added to the mixture at a final concentration of 1 μM. Hydrochloric acid was added to a final concentration of 1 mM after 5 min to terminate the reaction. The enzyme activity of RdAcuH was measured by determining the production of 3-HP-CoA using liquid chromatography-mass spectrometry (LC-MS). Components of the reaction system were separated on a reversed-phase C18 column (4.6 × 250 mm, 5 μm particle size; Waters, United States) connected to a high performance liquid chromatography (HPLC) system (Dionex, United States). The detection wavelength was 260 nm. The samples were eluted with a linear gradient of 1-20% (v/v) acetonitrile in 20 mM ammonium acetate (pH 5.5) over 15 min at a flow rate of 1 ml/min. The injection volume was 5 μl and the column temperature was 25°C. The HPLC system was coupled to an impact HD mass spectrometer (Bruker, Germany) for molecular weight (MW) determination.
+ Open protocol
+ Expand
2

Phenolic Profile Analysis of Cassia fistula Flowers

Check if the same lab product or an alternative is used in the 5 most similar protocols
Various types of phenolic compounds in C. fistula flower extracts were analyzed using HPLC analysis and were then compared with standard gallic acid, catechins, protocatecheuic acid, vanillic acid, chlorogenic acid, ferulic acid and coumaric acid. The flower extracts were dissolved in 50% ethanol and were further assessed by HPLC (Agilent Tecnologies, CA, USA) using reversed-phase C18 column (WATER, MA, USA). The mobile phase consisted of methanol (A) and 0.1% trifluoroacetic acid (TFA) in water (B) with gradient condition. The flow rate was set at 1.0 mL/min and the detection wavelength was recorded at 280 with a UV detector. The concentration levels of the phenolic compounds were calculated and compared with the standard curve considering the standard concentrations and peak areas (mg/g extract).
+ Open protocol
+ Expand
3

HPLC Quantification of Curcumin and Demethoxycurcumin

Check if the same lab product or an alternative is used in the 5 most similar protocols
The concentrations of Cur or DZG in the serum or organ tissues were determined using the HPLC method. The tissues were homogenized in 200 μl of 50% acetonitrile and vortexed for 5 min. Samples were centrifuged at 96.5 g at 4°C for 20 min. Fifty microliters of serum or tissue homogenate was mixed with 1.5-mg/mL trichloroacetic acid in 100 µl of acetonitrile. The substance was then centrifuged at 96.5 g at 4°C for 20 min. Sixty microliters of the supernatant was collected and subjected to HPLC using a reversed-phase C18 column (Waters Corporation, Milford, MA, USA). The mobile phase was composed of 52% acetonitrile and 48% citric buffer (1% w/v citric acid, pH 3.0), and the flow rate was set at 1 mL/min. The detection wavelength for Cur and DZG were 425 and 340 nm, respectively [11 (link),34 (link)]. Concentrations of Cur and DZG in the serum were calculated and compared to the standard curve (0–3200 ng/mL). The limits of quantification for Cur and DZG were 100 and 50 ng/mL, respectively. The extraction recovery from the plasma was approximately 90% for both Cur and DZG.
+ Open protocol
+ Expand
4

Quantification of Chlorogenic Acid by HPLC

Check if the same lab product or an alternative is used in the 5 most similar protocols
CGA was quantified by HPLC. A Merck Hitachi LaChrome equipment with a L7000 interface module, a L7200 autosampler, a L7490 RI detector, a UV detector, a L7350 column oven and a L7100 pump, was used associated with the D-7000 HSM software. A reversed-phase C18 column (5 μm, 250 × 4.6 mm, Waters) at RT was used with a flow rate of 0,9 mL/min. The mobile phase consisted of A: water/acetic acid (19:1) and B: methanol (0 min: 95%, 3 min: 85%, 13 min: 75%, 25 min: 70%, 27 min: 40%, 28 min: 0%, 30 min: 95% and 38 min: 95% of A). CGA was detected at 320 nm and eluted at 15 min retention time. CGA was quantified in Cc leaf phenolic-derived extracts from the linear equation (y = 122,049 × − 1,236,603 where x and y represent CGA concentration and peak area, respectively; r2 = 0.999), prepared with the injection of CGA pure standard (Sigma, USA) in methanol/water (50:50) (10–800 μg/mL). Cc leaf phenolic-derived extracts were diluted in methanol/water (50:50) (10 mg/mL) in duplicate and filtered through cellulose acetate filter, 0.22 μm pore size, prior to injection. The data acquisition was carried out using the D-7000 HSM software (Hitachi, Japan).
+ Open protocol
+ Expand
5

HPLC Analysis of MH Concentrations in Rat Spinal Cord

Check if the same lab product or an alternative is used in the 5 most similar protocols
1, 3, 7 and 21 days after MH administration, the rats were euthanized with Euthasol. MH concentrations in C5 segments were measured by HPLC as reported by Milane et al. [46 ]. Each segment was homogenized in 100 μL ice-cold phosphate buffer (0.01 M, pH 7.4). 2.5 μl of phosphoric acid (H3PO4) was added to the homogenate, and the mixture was vortexed and centrifuged at 2,000 rpm for 30 min at 4°C. MH was extracted to the supernatant. The precipitate was re-dissolved in 100 μL phosphate buffer and centrifuged. The second supernatant was combined with the first supernatant for HPLC analysis to measure MH concentration. HPLC analysis was performed on a Waters 1525 Binary HPLC system (Milford, MA) equipped with a UV/visible detector and a reversed phase C18 column (Waters, 75 × 4.6 mm, 3.5 μm). The mobile phase consists of phosphate buffer (25 mM, pH 3.0)/methanol/acetonitrile at a volume ratio of 91:6:3 and ran at a flow rate of 1.53mL/min. The sample injection volume was 203μL and the UV detection wavelength was 2453nm.
+ Open protocol
+ Expand
6

Silibinin Identification in Mouse Plasma

Check if the same lab product or an alternative is used in the 5 most similar protocols
Silibinin was identified in mice blood plasma using Liquid chromatography coupled with electrospray ionization mass spectrometry (LC-ESI-MS). The High Pressure Liquid Chromatography (HPLC) separation was performed at 25°C using a Waters® reversed phase C18 column (100 mm × 2.1 mm dimension; 5 μm particle size). ESI-MS was carried out using Waters® micromass® Q-TOF Mass Spectrometer at Regional Sophisticated Instrumentation Centre (RSIC), Panjab University, Chandigarh, India. For HPLC, the mobile phase consisted of acetonitrile and 10 mM ammonium acetate (pH 5.45) in the ratio of 50:50 (v/v). The detection of Silibinin was carried out at 288 nm and the flow rate was 100 μl/min with a total run time of 10 minutes. Mass spectrometric (MS) analysis was performed in negative ion mode under the following conditions: capillary voltage 2960 V, sample cone voltage 30 V, extraction cone voltage 1.0 V and desolvation temperature 350°C. The mass spectra were recorded in the m/z range of 100–700.
+ Open protocol
+ Expand
7

Characterization of Organic Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
Optical rotations were measured on a JASCO P-2000 instrument. The 1D and 2D NMR spectra were recorded on either a Bruker DPX 400 instrument with tetramethylsilane as an internal standard and MeOH-d4 as a solvent. HR-ESI-MS experiments were conducted using a Thermo Fisher QE Focus spectrometer. The semi-preparative HPLC procedure was conducted on a Shimadzu LC-16D instrument with an RID-20A (reflective index detector) and a reversed-phase C18 column (250 × 10 mm, 5 μm, Waters SunFire). Column chromatography was performed using silica gel (200–300 mesh, China), octadecyl silica (ODS) (50 μm, Merck, Germany).
+ Open protocol
+ Expand
8

Measuring Serum Folate and Homocysteine Levels

Check if the same lab product or an alternative is used in the 5 most similar protocols
Serum folate levels were determined five days after STZ treatment and then every four weeks, using an automated chemiluminescence system (Immulite 1000; Siemens, Berlin, Germany).
Tail vein blood glucose level was monitored by using a glucometer (Accu-Chek; Roche, Penzberg, Germany). Blood glucose test strips were used to measure the blood glucose concentration in mice.
Plasma levels of Hcy were analyzed using high-performance liquid chromatography (HPLC) with a Waters 700 HPLC Pump and a reversed-phase C18 column (5 µm bead size; 4.6 mm × 250 mm) (Waters, Milford, CT, USA). Briefly, the mobile phase consisted of 0.08 M acetate buffer and 5% (v/v) methanol adjusted to pH 4.0 by addition of concentrated acetic acid and then filtered through a 0.45 µm membrane filter. The isocratic elution was performed using a flow rate of 1.0 mL/min at 30 °C and a pressure of 100–110 kgf/cm2 (1500–1800 psi). A fluorescence detector with excitation at 390 nm and emission at 470 nm was used to detect Hcy. Before analysis of Hcy, the system was calibrated with authentic DL-homocysteine standards in the range of 50 to 4000 ng. Plasma Hcy was quantified relative to the standard obtained from Sigma Chemical Co. (St. Louis, MO, USA).
+ Open protocol
+ Expand
9

HPLC Quantification of Ascorbic Acid

Check if the same lab product or an alternative is used in the 5 most similar protocols
HPLC analysis of ascorbic acid was used following a method modified from Asami et al. (41 (link)). Small pieces of frozen samples (1–5 g) were homogenized with 20 mL of 4.5% metaphosphoric acid (HO3P) in ultra turrax (IKA Laroratechnik, Malaysia) during 4 min. The extract was centrifuged at 16,096 × g for 15 min at 4°C, and the supernatant was filtered using filter paper (Whatman no. 1). ascorbic acid content was then quantified by Shimadzu’s HPLC system on reversed-phase C18 column (150 mm × 4.6 mm diameters, Waters Corporation, USA) under the following conditions: injected volume 20 µL; oven temperature 40°C; solvent potassium phosphate (10 mM); flow rate 1.5 mL/min, and the detection was performed with an ultraviolet wavelength of 242 nm. Total ascorbic content was calculated from calibration curves by comparison with external standard (ascorbic acid, Sigma-Aldrich, MO, USA) (42 , 43 (link)).
+ Open protocol
+ Expand
10

Quinone Capture and Quantification

Check if the same lab product or an alternative is used in the 5 most similar protocols
Owing to the instability and high reactivity of quinones, BSA was used to capture quinones to form the stable products. Hence, quinones were determined as quinone-BSA adducts. UHPLC-MS/MS was used to analyze polyphenols and quinone-BSA products. Samples were filtered with 0.22 μm Nylon membrane filter, of which 3 μL were injected into a Waters ACQUITY UPLC System coupled with a Waters Xevo TQ-S and a reversed-phase C18 column (2.1 × 100 mm, 1.7 μm particles, Waters). The mobile phases were composed of (A) 0.1 % formic acid in water and (B) 0.1 % formic acid in methanol. The flow rate was 0.3 mL/min, and the column temperature was 30 °C. The gradient was as follows: 0 min, 10 % B; 4 min, 40 %; 6 min, 60 % B; 7 min, 100 % B; 8 min, 10 % B; 9 min, 10 % B. Samples in the automatic sampler tray were held at 10 °C.
The analysis was carried out in negative ion mode with an ESI interface. The source parameters were as follows: capillary voltage, 1.57 KV; cone gas flow, 700 L•Hr-1; desolvation temperature, 550 °C. Multiple reaction monitor (MRM) was used, where the parent ion and daughter ion were m/z = 289 and 203 for CAT, m/z = 353 and 191 for CQA, m/z = 428.9 and 125.06 for CATQ-BSA, m/z = 492.9 and 301.05 for CQAQ-BSA.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!