The largest database of trusted experimental protocols

9 protocols using sq detector

1

Enzymatic Synthesis of Ginsenoside C-Mc

Check if the same lab product or an alternative is used in the 5 most similar protocols
The minor ginsenoside C-Mc was prepared from American ginseng PPD ginsenosides, using 6% PPD ginsenoside substrate in acetate buffer (0.02 M and pH 5.0). Ginsenoside C-Mc was reacted with a volume of crude enzyme (A. niger g.848 strain) in the bioreactor.
In brief, chromatographic analysis was completed by a Waters ACQUITY UPLC system using XBridge C-18 chromatographic column (5 μm, φ4.6 × 250 mm) was used to analyze the samples. The mobile phase was acetonitrile (A) and water (B), column oven was 35°C, and flow rate is 0.6 ml/min. Detection wavelength of the DAD was 203 nm.
The mass examination was completed by a Waters SQ detector through a positive electrospray ionization pattern. Briefly, cone gas flow was set at 50 L/h, and desolvation gas flow was 550 L/h. Capillary voltage and cone voltage were 3.5 kV and 30 V, respectively.
The sample of ginsenoside C-Mc was identified as 20-O-[α-L-arabinofuranosyl-(1→6)-β-D-glucopyranosyl]-20(S)-protopanaxdiol using a BrukeAvance 600 (1H: 600 MHz and 13C: 150 MHz) NMR spectrometer (Switzerland) in our previous study [42 (link)].
+ Open protocol
+ Expand
2

UPLC-MS Quantification of Artemisinin

Check if the same lab product or an alternative is used in the 5 most similar protocols
Analysis of artemisinin was achieved by UPLC using a Waters Acquity UPLC™ system fitted with a single quadrupole (SQ) detector (Waters Corporation, Milford, MA). The mobile phase was an isocratic system consisting of 60% v/v aqueous solution of 0.1% formic acid and 40% v/v acetonitrile. The column was a Waters BEH300 C4 1.7 μm 100 mm × 2.1 mm column. The flow rate was 0.4 mL/min. For the four lowest concentration standards, a full loop (10.0 μL) of each sample was injected, and for the remaining standards 2.0 μL of the standard was injected using the partial loop with needle overfill injection mode.
Detection was achieved with a Waters SQ detector using selective ion monitoring of the [M-H2O+H]+ with m/z of 265.3 in positive ion mode. The operation parameters were a capillary voltage of 3.5 kV, a cone voltage of 41 V, extractor of 3 V, and RF lens of 0.5 V. The source and desolvation temperatures were 135°C and 250° C, respectively. Pure 99.99% nitrogen was used at a flow rate of 500 L/hr for the desolvation and 2 L/hr for the cone. The operational parameters including the target mass were determined by running a mass scan and tuning the instrument to the desired mass.
The retention time was 2.2 minutes and quantification was based on the UPLC-MS peak area of artemisinin.
+ Open protocol
+ Expand
3

Synthesis and Analytical Characterization of PACAP Peptides

Check if the same lab product or an alternative is used in the 5 most similar protocols
PACAP1-27 and PACAP1-38 were synthesized in our laboratory on a CEM Liberty microwave peptide synthesizer (Matthews, NC, USA) and were dissolved in the following sterile vehicles: (i) 0.9% saline solution, (ii) benzalkonium chloride solution for ophthalmic use (SOCB), (iii) thimerosal solution for ophthalmic use and (iv) water for injection, obtained from the Faculty Central Pharmacy, Faculty of Medicine, University of Szeged. A commercially available artificial tear solution (Systane Ultra®, Alcon, Switzerland) was also used in the experiment.
Analytical reversed-phase high-performance liquid chromatography (RP-HPLC) was performed on an Agilent 1200 Series separation system with diode array and multiple wavelength detector (Waldbronn, Germany), with a Luna C18(2) 100 Å column (10 μm, 250 × 4.6 mm; Phenomenex, Torrance, CA, USA). The chromatography was carried out at room temperature (RT), with a flow rate maintained at 1.2 mL min−1 at a wavelength of 220 nm [mobile phase solvent A: 0.1% TFA in Milli-Q water; solvent B: 0.1% TFA in acetonitrile (AcN)] using gradient elution. Mass spectrometry (MS) data were collected on a Waters SQ Detector (Milford, MA, USA) with an API mass spectrometer in positive ion mode.
+ Open protocol
+ Expand
4

UPLC-MS Analysis of Photoreactive Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
A2E (18 mM in DMSO) was diluted in DPBS to a final A2E concentration of 100 μM. Quercetin and cyanidin-3-glucoside were then added to obtain concentrations of 100 and 200 μM, and were irradiated (430 ± 30 nm, 1.5 milliwatts/cm2) for 3 min. The samples were examined by UPLC with a Waters SQ Detector (ion source; electrospray ionization mode (ESI), mass analyzer; single quadrupole). To acquire full scans, 5 μL of each sample was directly injected (without column) with a solvent of 70% acetonitrile/methanol (1:1) in 30% water containing 0.1% formic acid at a flow rate of 0.5 mL/min. UPLC-MS analysis was performed using positive ionization mode and voltages as follows: capillary voltage, 3.0 V; cone voltage, 30 V; extractor voltage, 3 V; and radio frequence lens voltage, 0.1 V. The desolvation gas flow was 800 liters/hr and the cone gas flow was 50 liters/hr; the source temperature was 150 °C, and desolvation temperature was 400 °C.
+ Open protocol
+ Expand
5

Purification and Characterization of Cf Peptides

Check if the same lab product or an alternative is used in the 5 most similar protocols
The crude Cf peptides
were purified using a C18 RP-HPLC on a PerfectSil 100 ODS-3 5 μm
column (250 × 10 mm). All compounds are >95% pure by HPLC
analysis.
In most cases, the two regioisomers (5)-Cf and (6)-Cf were well separable
on the RP-HPLC column (100% purity of the given isomer confirmed by
NMR); only for ERD-B, an isomeric ratio of 80:20 could be achieved
during purification.
The mass accuracy of the products was determined
by ESI-MS using a Waters SQ detector (Milford, MA, USA). The mass
spectra were recorded in positive ion mode in the 200.0–3000.0 m/z range.
+ Open protocol
+ Expand
6

Synthesis and Characterization of Diamidocalix[4]arene

Check if the same lab product or an alternative is used in the 5 most similar protocols
Solvents and reagents were obtained from commercial sources and used without further purification. Analytical TLC was performed using prepared plates of silica gel (Merck 60 Merck KGaA, Darmstadt, Germany F-254 on aluminum). 1H and 13C NMR spectra were recorded on a Bruker Billerica, MA, USA AV400 spectrometer. All chemical shifts are reported in parts per million (ppm) using the residual peak of the deuterated solvent, whose values are referred to tetramethylsilane (TMS, δTMS = 0), as internal standard. 13C NMR spectra were performed with proton decoupling. Mass spectra were recorded in ESI mode on a single quadrupole instrument, SQ Detector, Waters (capillary voltage 3.7 kV, cone voltage 30–160 eV, extractor voltage 3 eV, source block temperature 80 °C, desolvation temperature 150 °C, and cone and desolvation gas (N2) flow rates 1.6 and 8 L/min, respectively). UV-vis spectra were recorded on a Thermo Scientific Waltham, MA, USA Evolution 260 Bio spectrophotometer. Melting points were determined with a Gallenkamp apparatus.
Diamidocalix[4]arene 1 was synthesized according to a literature procedure [30 (link)].
+ Open protocol
+ Expand
7

HPLC Separation and Analysis Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
All HPLC separations involved a mobile phase of 0.05% (v/v) TFA in water (solvent A), and 0.04% (v/v) TFA in acetonitrile (solvent B).
Analytical HPLC–MS chromatographic separations were performed using a Waters Alliance e2695 Separations Module, an SQ Detector, and a Waters 2489 UV/Visible (UV/Vis) Detector equipped with an Agilent C18 column (5.0 μm, 4.6 × 150 mm) at a flow rate of 0.4 ml/min, or with a Welch-XB C4 column (3.0 μm, 3.0 × 150 mm) at a flow rate of 0.3 ml/min. The wavelengths of the UV detector were set to 210 and 220 nm.
Preparative HPLC separations were performed using a Hanbon Science & Technology NP7005C solvent delivery system and a Hanbon Science & Technology NU3010C UV detector equipped with an Exsil Pure 300 C18 column (10.0 μm, 20 × 250 mm) at a flow rate of 15 ml/min, or a Welch-XB C4 column (5 μm, 10 × 250 mm) at a flow rate of 4 ml/min. The wavelengths of UV-detector were set to 210 and 220 nm.
+ Open protocol
+ Expand
8

UPLC-MS analysis of lipid families

Check if the same lab product or an alternative is used in the 5 most similar protocols
Samples were analysed by reverse phase UPLC–MS (UPLC, Waters, Acquity class H) coupled with a single quadrupole MS (SQDetector, Waters, Acquity) on an Acquity UPLC BEH C18 (Waters) 2.1 × 50 mm, 1.7 μm column. We used a method, based on acetonitrile gradients, that allowed the simultaneous detection of all three LP families. Elution was started at 30% acetonitrile (flow rate of 0.60 ml min−1). After 2.43 min, the percentage of acetonitrile was brought up to 95% and held until 5.2 min. Then, the column was stabilized at an acetonitrile percentage of 30% for 1.7 min. Compounds were identified on the basis of their retention times compared with authentic standards (98% purity, Lipofabrik society, Villeneuve d'Asc, France) and the masses detected in the SQDetector. Ionization and source conditions were set as follows: source temperature, 130°C; desolvation temperature, 400°C; nitrogen flow, 1000 l h−1; cone voltage, 120 V.
+ Open protocol
+ Expand
9

Quantitative LC-UV/MS Analysis of Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
LC-UV/MS analysis was carried out on a Waters Acquity ultra-high performance liquid chromatographic system (Milford, MA, USA) equipped with a photodiode-array detector (PDA) and a single quadrupole mass spectrometer (Acquity SQ Detector) with an electrospray ionization source (ESI) operating in negative mode. An Agilent Poroshell EC-C18 column (150 mm × 4.6 mm, 2.7 µm) held at 50° C was used for the separation with 1 mM ammonium fluoride (NH4F) in water (solvent A) and acetonitrile (solvent B) as mobile phases. The flow rate was 0.6 mL/min with an elution gradient as follows: 0–2 min 75% A, 2–5 min 75–68% A, 5–14 min 68% A, 14–16 min 68–50% A, 16–18 min 50% A, 18–18.5 min 50–10% A, 18.5–22 min 10% A, 22–22.5 min 10–75% A, and 22.5–28 min 75% A. The injection volume was 5 µL. The mass acquisition window was from 50 to 2000 Da.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!