The largest database of trusted experimental protocols

13 protocols using orbitrap exactive

1

Orbitrap Mass Spectrometry for Compound Detection

Check if the same lab product or an alternative is used in the 5 most similar protocols
Detection of compounds was performed using a thermo Orbitrap Exactive mass spectrometer (Thermo Fisher Scientific, San Jose, USA). Electrospray ionization source (HESI-II) was used for ionization of the target compounds in positive and negative ion modes. Instrument calibration in positive mode and negative mode were done prior to sample injection (ion calibration solution 88323 and 88324, for positive and negative ions, respectively, Thermo Scientific Pierce). The tuning parameters were set to avoid fragmentation and keep ions intact as follows: ESI voltage, +3.3 kV/−3.0 kV; sheath gas pressure, 30.0 arbitrary units and capillary temperature, 373 K. The RF of the ion guides were optimized to obtained the maximum transmission efficiencies. For the compounds of interest, a scan range of m/z 500–4000 was chosen. The automatic gain control (AGC) target for a maximum capacity in C-trap was set at 2 × 106 ions for a maximum injection time of 250 ms.
+ Open protocol
+ Expand
2

Targeted Metabolomics Using Orbitrap LC-MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
Sample pellets were extracted using a 1:3:1 (v:v:v) chloroform/methanol/water extraction, subjected to liquid chromatography using a 2.1 mm ZIC-HILIC column (Sequant), and analyzed using the Orbitrap Exactive (Thermo Fisher Scientific) platform at the Glasgow Polyomics Center (University of Glasgow, Glasgow, Scotland) exactly as described elsewhere [28 (link),29 (link)]. Both negative and positive ion modes were run in parallel with rapid polarity switching. Samples were analyzed in randomized order and the same analytical batch. Additionally, to aid accurate metabolite identification, verify LC-MS stability, and identify contaminants, the following control samples were included as well: (1) solvents blanks, (2) serial dilutions of a pooled sample (Undiluted, 1/2, 1/4, 1/8, and 1/16), (3) Authentic standard mixes containing in total 217 metabolites (50–400 Da) and representing a wide array of metabolic classes and pathways, (4) An amino acid standard mix (Sigma Product No. A9906).
+ Open protocol
+ Expand
3

Quality Control for Metabolomic LC-MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
Samples were analyzed with LC-MS using an Orbitrap Exactive mass spectrometer (Thermo Fisher) coupled to a 2.1 mm ZIC-HILIC column at Glasgow Polyomics (University of Glasgow, Scotland) (Sequant), exactly as previously described (Berg et al., 2015 (link)). Both negative and positive ion modes were run in parallel with rapid polarity switching. Several quality control samples were run at the start, middle and end of the LC-MS run to aid accurate metabolite identification and verify LC-MS stability (Berg et al., 2013a ). These included: 1) Authentic standard mixes containing in total 217 metabolites (50–400 Da) representing a wide array of metabolic classes and pathways. 2) An amino acid standard mix (Sigma Product No. A9906). 3) Serial dilutions (Undiluted, 1/2, 1/4, 1/8 and 1/16) of a pooled sample of all extracts to filter out spurious signals that do not follow the dilution trend (Jankevics et al., 2012 (link)).
+ Open protocol
+ Expand
4

Purification and Characterization of Organic Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
All of the commercially available reagents, catalysts, bases, and solvents were used as purchased without further purification. Starting materials and reaction products were purified by flash chromatography using SiO2 as a stationary phase, eluting with n-hexane/ethyl acetate mixtures. 1H NMR (400.13 MHz), 13C NMR (100.6 MHz), and 19F spectra (376.5 MHz) were recorded with a Bruker Avance 400 spectrometer. Splitting patterns are designed as s (singlet), d (doublet), t (triplet), q (quartet), m (multiplet), or bs (broad singlet). HRMS of samples were recorded on an Orbitrap Exactive (Thermo Fisher, Waltham, MA, USA). Melting points were determined with a Büchi B-545 apparatus and are uncorrected.
+ Open protocol
+ Expand
5

HRMS Analysis of Metabolite Extracts

Check if the same lab product or an alternative is used in the 5 most similar protocols
HRMS analysis was performed by ionization (ESI) mass spectrometry (MS). The extracts were dissolved to a final concentration of 1-2 pmol/μL in methanol. Compounds were measured in negative and positive modes by full mass scanning (m/z 50 to 1000) using a Thermo Scientific Orbitrap Exactive mass spectrometer equipped with a heated electrospray ionization source (HESI-II). The instrument parameter settings are as follows: sheath gas 10 in positive mode and 20 in negative mode (arbitrary units), spray voltage 3.5 kV and 3 kV in positive and negative mode, respectively, with a 275°C capillary temperature. Mass spectra were collected at a resolution of 100,000. Data processing was performed using the associated software, Xcalibur 2.2 and Exactive 1.1.
+ Open protocol
+ Expand
6

Identification of γ-Butyrolactone-like Molecules

Check if the same lab product or an alternative is used in the 5 most similar protocols
For identification of R. jostii RHA1 γ-butyrolactone-like molecules, HPLC-MS analysis was performed using an Accella1250™ HPLC system coupled with the benchtop ESI-MS Orbitrap Exactive™ (Thermo Fisher Scientific, San Jose, CA). A Reversed Phase C18 (Shim Pack Shimadzu XR-ODS 3 × 75 mm) column was used and a gradient from 2% to 95% of acetonitrile:water (0.1% Formic Acid) as follows: 2 min 2% acetonitrile, 2–10 min gradient to 95% acetonitrile, 1 min 95% acetonitrile. To separate further the peaks from A-factor and 6-dehydro SCB2, a gradient from 2% to 80% acetonitrile was applied to the separation: 2 min 2% acetonitrile, 2–25 min in 2–80% acetonitrile, 1 min 80% acetonitrile. Data was analysed using Xcalibur software from Thermo Scientific. LC-MS analysis was performed with 2–4 biological replicates per strain.
+ Open protocol
+ Expand
7

UHPLC-HRAM-MS Metabolomics Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
The prepared samples were applied on an analytical system consisting of ultra-high performance liquid chromatography (UHPLC) Ultimate 3000 (Thermo-Fisher Scientific, Waltham, MA, USA) connected with a high-resolution accurate mass spectrometer (UHPLC/HRAM-MS, Orbitrap Exactive, Thermo-Fisher Scientific, Waltham, MA, USA). Methanol (100%) and water with 5 mM ammonium formate were used as mobile phases for gradient elution. The gradient began at 10% (0 min to 2 min) and then grew to 100% of A in 10 min, which was maintained until 15 min for washing of column, and finally was returned to initial conditions (10% A at 16 min) and kept for 4 min for system conditioning. The separation was performed using Kinetex Synergi Hydro-RP column (2.5 µm, 100 × 2.1 mm, Phenomenex, Torrance, CA, USA) tempered to 35°C and the flow rate set to 250 µL/min. The injection volume was 5 µL. The MS analysis was performed using positive ESI ionization and data was collected in FullMS/AIF scanning mode (resolution of 70,000 FWHM, m/z range from 170 to 900).
+ Open protocol
+ Expand
8

Targeted Metabolomic Analysis by LC-MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
LC–MS analysis was performed as described in [30 (link), 49 (link)]. Metabolites were extracted from supernatant media before and after treatment and by lysing cells in ice-cold methanol/acetonitrile/H2O (50:30:20). Samples were shaken at 4 °C for 10 min and then centrifuged for 15 min at 16,000 g, and the supernatant was collected and analyzed by LC–MS. Analytes were separated using hydrophilic interaction liquid chromatography with a SeQuant ZIC-pHILIC column (2.1 3 150 mm, 5 mm) (Merck) and detected with high-resolution, accurate-mass mass spectrometry using an Orbitrap Exactive in line with an Accela autosampler and an Accela 600 pump (Thermo Scientific). The elution buffers were acetonitrile for buffer A and 20 mM (NH4)2CO3 and 0.1% NH4OH in H2O for buffer B. A linear gradient was programmed starting from 80% buffer A and ending at 20% buffer A after 20 min, followed by wash (20% buffer A) and re-equilibration (80% buffer A) steps with a flow rate of 100 ml/min. The mass spectrometer was fitted with an electrospray-ionization probe and operated in full-scan and polar-switching mode with the positive voltage at 4.5 kV and negative voltage at 3.5 kV. Metabolite identification and data analysis were carried out using LCQUAN software (Thermo Scientific).
+ Open protocol
+ Expand
9

Metabolomic Analysis of Nanovibration-Stimulated MSCs

Check if the same lab product or an alternative is used in the 5 most similar protocols
MSCs were stimulated with nanovibration for 7 and 14 days in 2D and 3D (collagen gels; 2 mg ml−1) culture. Nonstimulated samples cultured in expansion media and OGM were used as controls. Metabolites were extracted using a 1:3:1 chloroform/methanol/water extraction buffer and vigorously shaken at 4°C for 1 hour. Following this, metabolite extraction solution was collected, transferred to 1.5-ml tubes, and centrifuged for 3 min at 13,000g at 4°C. Metabolomics was performed through hydrophilic interaction LC-MS analysis (UltiMate 3000 RSLC, Thermo Fisher Scientific) with a 150 mm by 4.6 mm ZIC-pHILIC column running at 300 μl min−1and Orbitrap Exactive (Thermo Fisher Scientific). A standard pipeline, consisting of XCMS (58 (link)) (peak picking), MzMatch (59 (link)) (filtering and grouping), and IDEOM (60 (link)) (further filtering, postprocessing, and identification), was used to process the raw mass spectrometry data. Identified core metabolites were validated against a panel of unambiguous standards by mass and retention time. Further putative identifications were allotted mass and predicted retention time (22 (link)). Means and SEs of the mean were generated for every group of picked peaks, and the resulting metabolomics data were uploaded to IPA software for pathway analysis.
+ Open protocol
+ Expand
10

UHPLC-MS Quantification of Sugars

Check if the same lab product or an alternative is used in the 5 most similar protocols
Ultra-high performance liquid chromatography Orbitrap Exactive – Thermo Scientific (UHPLC-MS) was performed using a NH4 column apHera 15 cm × 4.6 mm (Supelco Analytical– Sigma Aldrich). UHPLC parameters were: 40 °C temperature column, 10 μl injection volume, 0.3 ml/min flow injection, isocratic mode (ACN: H2O/70:30) and 19 min total retention time. Mass spectrometer parameters were: 2.5 KV spray voltage, 10 eV collision energy, negative mode ESI, 100 °C capillary temperature, 250 °C aux gas heater temperature, and a resolution of 70000.
The limit of quantification for the instrument was established with glucose, sucrose and fructose standards (Sigma Aldrich – Merck) and calibration curves for each of the main sugars. The range of calibration curve was from 25 to 200 μg/L for sucrose and from 50 to 400 μg/L for glucose and fructose, respectively. The lower limit of quantification was 9.82 μg/L glucose, 8.05 μg/L fructose and 10.69 μg/L sucrose, and the lower limit of detection was 2.95 μg/L glucose, 2.42 μg/L fructose and 3.2 μg/L sucrose.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!