The largest database of trusted experimental protocols

Zorbax extend c18

Manufactured by Agilent Technologies
Sourced in United States, Germany

The Zorbax Extend C18 is a high-performance liquid chromatography (HPLC) column designed for the separation and analysis of a wide range of analytes. It features a silica-based stationary phase with a C18 bonded ligand, providing good retention and selectivity for a variety of organic compounds.

Automatically generated - may contain errors

51 protocols using zorbax extend c18

1

Quantifying Hepatic Long-Chain Acyl-CoA Esters

Check if the same lab product or an alternative is used in the 5 most similar protocols
The level of hepatic LCACoAs was analyzed using UHPLC/MS/MS as described by Blachnio-Zabielska et al. [26 (link)]. LCACoA extraction was performed according to Minkler et al. [27 (link)] with the use of internal standards (C15:0-, 16:0(d4)-, C17-, C19:0-, C21:0-, C23:0- and 24:0(d4)-CoA) spiked into each sample. Samples were resolved on 2.1 mm × 150 mm Agilent ZORBAX Extend-C18, using a reverse-phase gradient of ammonium hydroxide (NH4OH) in water and NH4OH in ACN. The quantitative analysis of LCACOAs was performed in positive electrospray ionization mode (ESI+) on Sciex QTRAP 6500 + LC/MS/MS system (AB Sciex, Darmstadt, Germany).
+ Open protocol
+ Expand
2

Steroid Hormones Analysis Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
Analyses of estradiol, progesterone, androsterone, testosterone, DHEA, DHT, 17α-hydroxyprogesterone, pregnenolone and d3-testosterone were accomplished with the use of ZORBAX Extend C18 chromatographic column (2.1 mm × 100 mm, 3.5 μm; Agilent Technologies, Waldbronn, Germany). The mobile phase was composed of 0.1% aqueous solution of formic acid (phase A) and 0.1% formic acid solution in methanol (phase B). The gradient elution was utilized starting from 60% of phase B to 80% of B in 10 min, then was set at 80% of B for 4 min. The time for stationary phase equilibration was set at 6 min. The flow rate was 0.35 mL/min, the injection volume was 2 µL and the column temperature was maintained at 40°C.
+ Open protocol
+ Expand
3

HPLC Analysis of Pterostilbene Derivatives

Check if the same lab product or an alternative is used in the 5 most similar protocols
Samples (2 µL) were analyzed by HPLC/UV (1290 Infinity LC System, Agilent Technologies, Cernusco sul Naviglio (Milano), Italy) using a reverse phase column (Zorbax Extend-C18, 1.8 µm, 50 × 3.0 mm i.d.; Agilent Technologies, cat. n. 727975-302) and a UV diode array detector (190–500 nm). Solvents A and B were water containing 0.1% trifluoroacetic acid (TFA) and acetonitrile, respectively. The gradient for B was as follows: 10% for 0.5 min, then from 10% to 100% in 3.5 min, 100% for 1 min; the flow rate was 0.6 mL/min and the column compartment was maintained at 35 °C. Eluate was preferentially monitored at 286, 300, and 320 nm (corresponding to absorbance maxima of the internal standard, derivatives/metabolites, and pterostilbene, respectively).
+ Open protocol
+ Expand
4

HPLC-ELSD Alkaloid Quantification Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
The HPLC‐ELSD conditions were used for the determination of the alkaloids, and an LC‐20AT Shimadzu high‐performance liquid chromatography system (Kyoto, Japan) equipped with an Alltech 2000 evaporative light‐scattering detector (Grace,) was used. The chromatographic separations were performed over a ZORBAX Extend‐C18 (250 mm × 4.6 mm, 5 µm; Agilent) column at a column temperature of 25℃. The column was eluted with a mixture of acetonitrile (mobile phase A) and water‐ammonium bicarbonate (PH = 10.10, mobile phase B) at a flow rate of 1.0 ml/min. The elution conditions were as follows: 0–10 min, 75%A to 60% A; 10–20 min, 60% A; 20–30 min, 60% A to 30% A; 30–45 min, 30% A to 10% A; 45–55 min, 10% A; 55–60 min, 10% A to 75% A; 60–65 min, 75% A. The drift tube temperature of the ELSD was set at 115℃, and using nitrogen as the carrier gas at a flow rate set at 3.0 L/min, the gain value was 2, and the injection volume was 20 µl for analysis. The concentration of each compound was calculated based on external standard method, using the calibration curves of each compound.
+ Open protocol
+ Expand
5

Optimized UHPLC Method for Compound Separation

Check if the same lab product or an alternative is used in the 5 most similar protocols
Chromatographic separation was performed on a Zorbax Extend C18 UHPLC column (2.1 mm × 100 mm, 1.8 µm, Agilent, Palo Alto, CA, USA). The column temperature was set at 25 °C. Flow rate was 0.3 mL/min and the sample injection volume was 3 μL. The mobile phase consisted of solvent A (water) and solvent B (acetonitrile). The optimized UHPLC gradient elution program was as follows: 0–20 min, 5–70% B; 20–24 min, 70–100% B; and 24–26 min, 100% B.
+ Open protocol
+ Expand
6

HPLC Analysis of Phenolic Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
The HPLC analysis was performed on an Agilent Series 1200 chromatographic system (Agilent Technologies, Santa Clara, CA, USA) coupled with a photodiode array detector (PAD) (Agilent Technologies, Santa Clara, CA, USA) and using a Zorbax Extend C18 (4.6 × 250 mm, 5 μm i.d.) chromatographic column (Agilent Technologies, USA). The elution was performed in a gradient with a mobile phase consisting of (A) 1% (v/v) formic acid solution in water and (B) 1% (v/v) formic acid solution in acetonitrile. The regimen was as follows: 0 min—7% B; 5 min—7% B; 45 min—40% B; 47 min—70%; 52 min—70% B. The flow was 1 mL min−1, the injection volume 5 μL and the column temperature 25 °C. The chromatograms were recorded at 278 and 320 nm. The analysis for all samples was performed in duplicate. All samples were filtered through a 0.45 μm membrane filter (Nylon Membranes, Supelco, Bellefonte, PA, USA) prior to the analysis. Identification of individual phenolic compounds and caffeine was performed by comparing the retention times and the characteristic absorption spectrum (190–400 nm) with commercially available standards, while quantification was enabled by establishing calibration curves (20–100 µg mL−1).
+ Open protocol
+ Expand
7

Quantitative LC-TOFMS Analysis Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
An LC instruments set (Agilent Technologies 1260 Infinity II, Santa Clara, CA, USA) was connected to a time-of-flight mass spectrometer (TOFMS; Agilent Technologies G6230B) with an electrospray ionization source (Dual Agilent Jet Stream ESI). TOFMS detection was operated in negative ionization mode using the following scan source parameters: capillary voltage 3.5 kV, ion source gas temperature 300 °C, drying gas flow 10 L/min, and sheath gas temperature 350 °C with sheath gas flow 11 L/min. The column used was ZORBAX Extend-C18 (2.1 × 50 mm, particle size 1.8 μm; Agilent Technologies, P/N: 727700-902), and the column temperature was set at 45 °C. Column equilibration was carried out 1 day before actual measurement by passing mobile phases A and B at the starting ratio according to each measurement method.
+ Open protocol
+ Expand
8

Determination of Rosmarinic Acid by HPLC

Check if the same lab product or an alternative is used in the 5 most similar protocols
The content of rosmarinic acid was determined on an Agilent Series 1200 chromatographic system (Agilent Technologies, Santa Clara, CA, USA) using a Zorbax Extend C18 (4.6 × 250 mm, 5 μm i.d.) chromatographic column (Agilent Technologies, Santa Clara, CA, USA) and coupled with a Photodiode Array Detector (PAD) (Agilent Technologies, Santa Clara, CA, USA) (Table S1, Figure S1). The gradient elution was performed with a two-component mobile phase consisting of 1% (v/v) formic acid solution in water and 1% (v/v) formic acid solution in acetonitrile, as described by Šeremet et al. [26 (link)].
+ Open protocol
+ Expand
9

Optimizing ZEA Degradation Enzymes

Check if the same lab product or an alternative is used in the 5 most similar protocols
For ZEA degradation, ZHDCP, ZHD101 and carboxypeptidase enzymes had chosen. The ZEA degradation test (in a total volume of 1 mL) was started by adding concentrated purified protein, where ZEA 5 ppm, 100-μg enzyme, 900-μL reaction buffer (150 mM NaCl, 25 mM Tris-HCl, pH: 9.5). The reaction was completed by heating the reaction mixture at 98 °C for 2 min. For identification of ZEA and its metabolite, the reaction mixture was filtered and analyzed by UV-HPLC (274 nm). The optimum pH was investigated at 30 °C in different buffers in the pH range 4-11. The optimum temperature was studied at pH 7.5 using temperatures ranging from 20 to 45 °C. The UV-HPLC running parameter deployed are as follows: Agilent Zorbax Extend-C18, 3 × 150 mm, 3.5 μm particle size); mobile phase: methanol/water/formic acid (10/90/0.1, v/v/v), flow rate: 0.2 mL/min; nitrogen drying gas: 10 L/min, nebulizer pressure: 25 psi, drying gas temperature: 350 °C, capillary voltage: 4 kV.
+ Open protocol
+ Expand
10

Determining Lipophilicity of Synthesized Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
The lipophilicity of synthesized compound was evaluated using Agilent 1200 Series HPLC instrument (Agilent technologies, Waldbron, Germany) controlled through the Chemstation software (version B.04.03). The separation was performed on ZORBAX Extend-C18 (3 × 150 mm, 3.5 µm) column (Agilent Technologies, Waldbron, Germany). Methanol (constituent A) and purified water (constituent B) were used as a mobile phase. Total flow rate was set up to 0.4 mL/min; the injection volume was 1 µL; and the column temperature was maintained at 25 °C. The detection wavelength of 210 nm was used. Retention factors of the compounds were measured under five different isocratic conditions in the 65:35–80:20 range (A:B; v/v), in duplicate. The log kw values (indexes of lipophilicity) were estimated by linear regression analysis based on the following equation,

where logk represents a logarithm of an individual isocratic retention factor, Φ is the organic phase concentration and S is a constant derived from linear regression analysis.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!