The largest database of trusted experimental protocols

Chromeleon 7

Manufactured by Thermo Fisher Scientific
Sourced in United States, Germany, Spain

Chromeleon 7 is a chromatography data system software from Thermo Fisher Scientific. It provides data acquisition, processing, and reporting capabilities for various chromatography techniques, including liquid chromatography (LC) and gas chromatography (GC).

Automatically generated - may contain errors

173 protocols using chromeleon 7

1

HPLC Quantification of LVX and BTZ043

Check if the same lab product or an alternative is used in the 5 most similar protocols
APIs were quantified using a Dionex Ultimate 3000 U-HPLC (Thermo Fischer, Dreieich) equipped with a Synchronis C18 50 × 2.1 mm, 1.7 µm column (Thermo Fischer, Germany) and a UV–VIS detector (Thermo Fischer, Germany). (i) In case of LVX, 10 mg of dry powder was dissolved in 1 mL PBS containing 0.1% Triton-X and eluted with a 18% of mobile phase A (acetonitrile (ACN)) and 72% of mobile phase B (0.5% trimethylamine buffer at pH 2.5) at a flow of 0.3 mL/min (see Supporting Information, Fig. 7). (ii) In case of BTZ043 (Selleckchem; Houston, TX, USA), powder was dissolved in a 60:40 mixture of ACN:PBS at a concentration of 10 mg/mL of dry powder. A binary solvent gradient was applied at a flow rate of 0.3 mL/min and programmed as follows: 80% H2O and 20% ACN at 0 min to 0.5 min, progressing linearly at 5% H2O and 95% ACN at 2 min, followed by the hold in H2O 5% ACN to 95% for 1 min, and finally returning to the initial gradient until 5 min (cf. Supporting Information, Figs. 8 and 9). Data analysis was performed with Chromeleon 7 software (Thermo Fischer, Germany).
+ Open protocol
+ Expand
2

Rumen Fluid Analysis: Ammonia and VFA

Check if the same lab product or an alternative is used in the 5 most similar protocols
The rumen fluid samples were thawed at 4 °C, and the concentration of ammonia nitrogen was determined by the phenol-sodium hypochlorite colorimetric method. VFA concentration was determined by gas chromatography (GC) using methyl valerate as the internal standard in an Agilent 6890 series GC equipped with a capillary column (HP-FFAP19095F-123, 30 m, 0.53 mm diameter, and 1 cm thickness). Samples were injected using an auto-sampler (AI 3000, Thermo Scientific, Waltham, MA, USA) into an AE-FFAP capillary column (30 m × 0.25 mm × 0.33 μm, ATEO, Lanzhou, China) on a Varian GC (TRACE 1300, Thermo Scientific, MA, USA). Samples were run at a split ratio of 20:1 with a column temperature of 45 °C to 150 °C with an increase of 10 °C/min followed by a 5 min hold. The injector and detector temperatures were 200 °C and 250 °C, respectively. Peak integration was performed using Chromeleon® 7 Software (Thermo Fisher Scientific Inc., Waltham, MA, USA). All ruminal fluid samples were assayed in duplicate.
+ Open protocol
+ Expand
3

HPLC-based Quantification of ATP and ADP

Check if the same lab product or an alternative is used in the 5 most similar protocols
Indicated ATPase reactions were set-up as described above in microcentrifuge tubes. After completion of reactions, each tube was centrifuged at 10,000 x g for 2 minutes, and then 15 μL of supernatant was loaded onto a 96 well plate (Thermo Scientific), diluted with 35 μL of 50 mM potassium phosphate (pH 6.0) as running buffer. HPLC analysis was conducted with an autosampler using a 2.6 μm, 4.6 x 150 mm Accucore aQ C18 Polar Endcapped column (ThermoFisher Scientific). Each sample (15 μL) was eluted with an isocratic gradient of running buffer at a flow rate of 0.700 mL/min for 10 min at a column temperature of 30 °C. The presence of ATP and ADP was detected at a wavelength of 260 nm and quantified using Chromeleon 7 software (ThermoFisher Scientific) using ATP/ADP standards that were analyzed under the same conditions.
+ Open protocol
+ Expand
4

Paracetamol Quantification in SODF

Check if the same lab product or an alternative is used in the 5 most similar protocols
For each formulation, three individual SODFs were dissolved in 100 mL of distilled water. Samples of the solutions were then diluted and the drug concentration was determined by ultra-high performance liquid chromatography (UHPLC, Thermofisher Scientific, Waltham, MA, USA) using a UHPLC-DAD system. It consisted of a Thermo Scientific™ Dionex™ UltiMate™ 3000 BioRS equipped with a WPS-3000TBRS autosampler and a TCC-3000RS column compartment set at 35 °C. The system was operated using Chromeleon 7 software (Thermofisher Scientific, Waltham, MA, USA). An Accucore C18 column (2.6 µm, 100 × 2.1 mm2) combined with a security guard ultra-cartridge (Phenomenex Inc., Torrance, CA, USA) was used. An isocratic binary solvent system was utilized, consisting of water/formic acid (1%, v/v) as solvent A and acetonitrile/formic acid (1%, v/v) as solvent B (90%A, 10%B). The flow rate of the mobile phase was 1.5 mL/minute, and the injection volume was 50 μL. Quantitative analysis of paracetamol in the SODFs was carried out using an external standard method. The calibration curve was constructed using 5 different standard levels in the concentration range 1–20 mg/L. The peak of paracetamol was monitored at 244 nm.
+ Open protocol
+ Expand
5

Enantioselective Separation of 2-Pentanol

Check if the same lab product or an alternative is used in the 5 most similar protocols
Eight different chiral columns were selected, and the chiral column types were shown in Table 2. 2-Pentanol racemate was determined using a Thermo Fisher gas chromatograph (Trace 1300, Thermo Fisher, Waltham, Massachusetts, USA), and the best column was selected according to the separation resolution. Take 5 µL 2-pentanol racemate with anhydrous ethanol constant volume to 10 mL, filtered through a 0.22 µm nylon syringe filter membrane into the injection vial, and stored in a refrigerator at 4 °C until GC analysis. The 2-pentanol racemate was injected into the injection port in split mode (injector temperature 250 °C, split ratio: 20:1). The separation was carried out on eight different chiral columns under the same chromatographic conditions. GC conditions: column oven of 40 °C, held for 1 min, increased to 120 °C at a rate of 2 °C/min, and finally increased to 210 °C at 3 °C/min and held for 1 min. The carrier gas was nitrogen (99.999%) at a constant flow rate of 1 mL/min. Chromatograms were analyzed using Chromeleon 7 software (Thermo Fisher, Waltham, Massachusetts, USA).
+ Open protocol
+ Expand
6

HPLC Analysis of Antibody Monomer and HMWI

Check if the same lab product or an alternative is used in the 5 most similar protocols
SEC was used to assess the monomer and HMWI content and product losses. Experiments were carried out on a Dionex UltiMate 3000 HPLC system (Thermo Fisher Scientific, Waltham). Isocratic elution was carried out with a 50 mM sodium phosphate buffer with 150 mM NaCl at pH 7.0. 10 μl of 0.22 μm filtered sample were applied to a TSKgel® G3000SWXL HPLC column (5 μm; 7.8 mm × 300 mm) with a TSKgel® SWXL guard column (7 μm; 6.0 × 40 mm) (both Tosoh, Japan). Chromeleon™ 7 software (Thermo Scientific) was used to monitor the signals at 215 nm (for HMWI) and 280 nm (for monomer content and product loss). The starting material is referred to as T0 material. Monomer content is based on the ratio of product peak area (monomer, detected at 280 nm) to the sum of all peak areas. HMWI are calculated by taking the sum of all peaks eluting before the main antibody peak (at 215 nm) and dividing it by the sum of main peak and earlier eluting species. Concentration in relation to T0 is the ratio between the monomer peak at 280 nm of the respective sample with the T0 sample of the respective capture method (protein A or precipitation). Average values and standard deviations are based on triplicate measurements. Injection volume and dilutions were kept constant for all samples.
+ Open protocol
+ Expand
7

HPLC-SEC-MALS Analysis of Sample Composition

Check if the same lab product or an alternative is used in the 5 most similar protocols
Relevant samples were analyzed by HPLC‐SEC‐MALS in order to determine particle composition and estimate purity. All experiments were performed using the HPLC system mentioned in 2.11 with the Chromeleon 7 software (Thermo Fisher Scientific) for method programming, control and data acquisition. A TSKgel G5000PWXL column (300.0 × 7.8 mm i.d.) combined with a TSKgel PWXL guard column (40.0  × 6.0 mm i.d.) (Tosoh Bioscience, Stuttgart, Germany) was used. A volume of 50 µL of each sample was injected. The flow rate was 0.3 mL/min. Isocratic elution was performed with 50 mM HEPES, 100 mM NaCl, pH 7.2. UV signals at 280 and 260 nm were recorded by the Chromeleon software and LS signal was acquired with a DAWN HELEOS 18‐angle detector (Wyatt) with the Astra Software 5.3.4 (Wyatt). Data evaluation was performed in Astra 6.1.2.
+ Open protocol
+ Expand
8

MALS Analysis of Biomolecular Samples

Check if the same lab product or an alternative is used in the 5 most similar protocols
At‐line MALS measurements for the determination of the LS intensity were performed using an Ultimate 3000 HPLC system (Thermo Fisher) with a quaternary LPG‐3400SD pump, a WPS‐3000TSL autosampler, and a DAD 3000 UV‐detector. Mobile phase consisted of 50 mM HEPES, 100 mM NaCl, pH 7.2. A sample volume of 50 µL was injected in bypass mode using a flow rate of 0.3 mL/min. All samples were measured in duplicates. MALS signals were acquired by the DAWN HELEOS 18‐angle detector (Wyatt, Santa Barbara, CA, USA). For HPLC programming, Chromeleon 7 software (Thermo Fisher) was used. MALS data collection and analysis were performed with ASTRA software, version 6.1.2 (Wyatt).
+ Open protocol
+ Expand
9

UPLC-Based Methanol Quantification in Lactic Acid Samples

Check if the same lab product or an alternative is used in the 5 most similar protocols
Methanol in samples producing lactic acid were quantified via UPLC (UPLC Ultimate 3000) equipped with a Rezex ROA UPLC column (Rezex ROA-Organic Acid H+ (8%): 300 × 7.8 mm; Phenomenex) coupled to a refractive index detector (RefractoMax 521, Thermo Fisher Scientific). As the solvent, 2.5 mM sulfuric acid in ultrapure water was used. To separate the metabolites, a flow rate of 600 μl min−1 was used. Peaks were identified and the peak areas were quantified using Chromeleon 7 software (Thermo Fisher Scientific) on the basis of an external standard curve of pure methanol (Sigma-Aldrich).
+ Open protocol
+ Expand
10

Antibody Purity Estimation by SEC-HPLC

Check if the same lab product or an alternative is used in the 5 most similar protocols
To estimate product purity, size exclusion chromatography was performed. For HPLC analytics, we used a Dionex UltiMate 3000 HPLC system equipped with a diode array detector (Thermo Fisher Scientific). The running buffer was a 50 mM potassium phosphate buffer with 150 mM NaCl, pH 7.0 (Merck KGaA). The buffer was filtered through 0.22‐μm filters (Merck KGaA) and degassed. We applied 20 μl of the filtered sample to a TSKgel® G3000SWXL HPLC column (5 μm, 7.8 × 300 mm2) in combination with a TSKgel SWXL Guard column (7 μm, 6.0 × 40 mm2; Tosoh). The absorbance at 280 nm was recorded, and the results were evaluated with the Chromeleon™ 7 software (Thermo Fisher Scientific). The antibody purity was calculated as the ratio of the monomer peak area (retention time: 21.2 min) to the sum of all peak areas, based on the 280 nm signal.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!