The largest database of trusted experimental protocols

18 protocols using dxrxi raman imaging microscope

1

Characterization of LiNi₁/₃Co₁/₃Mn₁/₃O₂ and Hard Carbon

Check if the same lab product or an alternative is used in the 5 most similar protocols
The LiNi1/3Co1/3Mn1/3O2 powder was characterized by X-ray diffraction using a Huber G670 Guinier diffractometer (CoKα1 radiation (λ = 1.78892 Å), curved Ge(111) monochromator, image plate detector). To determine the lattice parameters, the Le Bail decomposition was carried out using the JANA2006 software. The XRD spectra is presented in Supplementary Fig. 4.
The hard carbon powder was characterized by scanning electron microscopy using Quattro S ESEM (FEI) and Raman spectroscopy using DXRxi Raman Imaging Microscope (Thermo Fisher Scientific). The results presented in Supplementary Fig. 5 illustrate a characteristic round particle of hard carbon and its Raman spectra with D-band (defect-induced) and G-band (crystalline graphite) peaks at ~1353 cm−1 and ~1585 cm−1 respectively.
Raman spectroscopy on the HOPG, pristine powders, and cross-sections was performed using DXRxi Raman Imaging Microscope (Thermo Fisher Scientific) using the 532 nm laser.
+ Open protocol
+ Expand
2

Characterization of Novel Nanomaterials

Check if the same lab product or an alternative is used in the 5 most similar protocols
1H and 13C-NMR spectra were measured on a Bruker Avance DRX600 spectrometer (Bruker, Billerica, MA, USA). Spot absorption spectra were obtained by use of a DXRxi Raman Imaging Microscope (Thermo Fisher Scientific, Waltham, MA, USA) with an excitation wavelength of 532 nm and 50 μm confocal pinhole diaphragm. Fourier transform infrared spectroscopy (FT-IR) was measured on a Bruker Equinox 55 Fourier transform infrared spectrometer (Bruker, Billerica, MA, USA). Scanning electron microscope (SEM) images were carried out using a HITACHI S-4300 scanning electron microscope (Hitachi, Tokyo, Japan). Thermo gravimetric analysis (TGA) was carried out with a Pyris-Diamond TGA (PerkinElmer, Boston, MA, USA). Ultra high performance liquid chromatography (UHPLC) was carried out with an American Waters Acquity UHPLC Class (Waters, Milford, MA, USA).
+ Open protocol
+ Expand
3

Raman Spectroscopy of BBQ650 Solutions

Check if the same lab product or an alternative is used in the 5 most similar protocols
Raman spectroscopic measurements were performed by using a confocal Raman microspectrometer based on the SpectraPro 2500i (Acton Research, Trenton, NJ) monochromator equipped with the Spec 10–100B CCD camera (Princeton Instruments, Trenton, NJ) for Raman signal detection .The three laser sources used for excitation are: −532 nm Verdi V-6, 633 nm He-Ne laser (both Coherent, Santa Clara, CA) and 780 nm OPO Levante (APE, Germany) pumped by a High-Q Laser (High Q Laser GmbH, Austria). This configuration enables the measurements within the range of Raman shift of 600–3000 cm−1. The spectral resolution for a fixed diffraction grating position (wave number interval of 1210 cm−1) was ~1.5 cm−1. The Raman spectra of BBQ650 solutions were measured using the following excitation powers: 633 nm–25 mW, 532 nm–25 mW, 785 nm–200 mW. Raman imaging was performed with the DXRxi Raman Imaging Microscope (Thermo Fisher Scientific, Madison, WI). For cell imaging, we used a 60x water immersion objective lens (Olympus, Japan).
The laser power at the sample plane during Raman imaging was 8 mW for 633 nm and 10 mW for 532 nm.
+ Open protocol
+ Expand
4

Structural and Optical Characterization of NaYF4:Yb3+/Er3+ Phosphor

Check if the same lab product or an alternative is used in the 5 most similar protocols
For structural analysis, powder samples were characterized using Bruker D8 Advance X-ray diffractometer operating at 40 kV tube voltage and 40 mA current in the 2θ range 10 to 90°. The Tecnai G2 TWIN was used to take TEM pictures of the NaYF4:Yb3+/Er3+ phosphor while operating at 200 kV acceleration voltages. The Raman spectra of NaYF4:Yb3+/Er3+ phosphor were recorded using a Thermo Scientific DXRxi Raman imaging microscope instrument equipped with a 532 nm laser. To capture the photoluminescence (PL) excitation and emission spectra of the powder sample, a fluoromax-plus spectrofluorometer with a 150 W xenon flash lamp was employed (without using an integrating sphere). PL decay measurements were carried out at the same set-up using a pulsed xenon lamp (25 W). The 976 nm tunable continuous-wave diode laser was used as an external excitation source in the same setup to record the UC emissions. A homemade heater equipped with a k-type thermocouple was utilized for the temperature-dependent UC emission measurements, and it was placed right next to the laser's focal point. A variac was used to control the temperature of the (by controlling the voltage of the heater) sample.
+ Open protocol
+ Expand
5

Raman Spectroscopy of Materials

Check if the same lab product or an alternative is used in the 5 most similar protocols
Raman spectra are collected in ambient air and at room temperature with a DXRxi Raman imaging microscope (Thermo Scientific). An excitation laser with a wavelength of 532 nm and a mapped area of 100 × 100 µm2 are used. An incident power of 10 mW is used to avoid sample damage or laser-induced heating.
+ Open protocol
+ Expand
6

SERS Microscopy Imaging Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
All samples were brought into focus using either 20×/0.40NA or 100×/0.9NA microscope objective lenses using a DXRxi Raman imaging microscope (Thermo Fisher Scientific Inc.) with a 780 nm (λ) laser line for SERS analysis. Each spectrum was generated based on 0.1 s collection‐exposures using a 3 mW laser line through a 50 μm slit‐aperture in the range of 600‐1800 cm−1. Chemical imaging was achieved by integrating at least 4000 spectra. The laser spot size was approximately 2.38 and 1.06 μm (ø), respective to the use of 20× and 100× magnification lenses. Single‐cell SERS microscopy was conducted using a 2 μm pixel step‐size and lower magnification images were constructed using a 40 μm pixel step size. Images were analyzed using the OMNICxi Raman SW software (Thermo‐Nicolet, Madison, Wisconsin, USA). Spectra were comparatively analyzed based on discriminatory features that were identified using the TQ Analyst 9.0 software (Thermo‐Nicolet).
+ Open protocol
+ Expand
7

Raman Imaging of Polymer Composites

Check if the same lab product or an alternative is used in the 5 most similar protocols
The printed pure polymer and BN-filled samples were examined to trace the specific peaks attributed to different components using a Thermo Scientific DXR xi Raman imaging microscope with Thermo Scientific OMNIC xi Raman imaging software. The measurements were conducted at room temperature and at the excitation wavelength of 532 nm from a laser that focused on the samples using a 50× objective for the 100–3500 cm−1 zone. The obtained spectra were exported to the Origin Pro data analysis software for data processing. The baseline was corrected, and the peak heights were normalized to 1.0.
+ Open protocol
+ Expand
8

Multimodal Characterization of Advanced Materials

Check if the same lab product or an alternative is used in the 5 most similar protocols
SEM and EDS were conducted on a Zeiss Supra 55VP field emission SEM equipped with a Thermo Fisher Scientific UltraDry EDS detector. The accelerating voltage for SEM and EDS were 3 and 15 kV, respectively. Transmission electron spectroscopy images and elemental mappings were collected using a JEM-2100F microscope equipped with an Oxford energy-dispersive X-ray analysis system, with the accelerating voltage of 200 kV. PXRD was performed on a Bruker D8 Advance powder X-ray diffractometer using Cu Kα radiation. XPS was performed on a Thermo Scientific K-Alpha XPS system with an Al Kα X-ray source. UPS was collected on a Thermo ESCALAB 250Xi XPS system with a He I source gun. The Raman spectra were collected on a Thermo Fisher Scientific DXRxi Raman imaging microscope with a 532 nm laser. The ICP-MS analysis was carried out on a Shimadzu ICPMS-2030 spectrometer. The XAS were collected in the transmission mode at the Advanced Photon Source Beamline 10-BM-B at the Argonne National laboratory. To collect the Co K-edge in the energy window from 7.450 to 8.650 keV, a 71/29 N2/He gas mixture was used in the I0 ion chamber to achieve 10% absorption, while a 68/32 N2/Ar gas mixture was used in the It ion chamber to achieve 70% absorption (calculated using Hephaestus at an energy of 7.709 keV). The Co foil standard was used for the energy calibration.
+ Open protocol
+ Expand
9

Raman Imaging and UPLC-MS/MS Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
Separated compounds on TLC were located under 254 nm by an Ultraviolet analyzer (YOKO-2F; Wuhan YOKO technology Ltd., Wuhan, China). Raman spectra and its imaging were obtained by use of a DXR™ xi Raman Imaging Microscope (Thermo Fisher Scientific, Waltham, MA, USA) with an excitation wavelength of 532 nm, a resolution of 5.0 cm−1, and a 10× long working distance microscope objective. The excitation power was 10 mW and the integration time was 0.5 s; number of scans was 20. The scan range was 3300~100 cm−1, with a 50 µm confocal pinhole DXR532 full range grating (400 line/mm). Detector was TE-cooled electron-multiplying CCD (EMCCD). Area scanning was chosen as the scanning mode, scanning area was more than 150 μm × 150 μm, and total scanning time was 20 min.
Ultra high-performance liquid chromatography–tandem mass spectrometry (UPLC–MS/MS) was operated on a Dionex UltiMate 3000 ultra-performance liquid chromatography–TSQ quantum mass spectrometer system (Thermo Fisher Scientific, Waltham, MA, USA).
+ Open protocol
+ Expand
10

Raman Mapping Microscopy Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
Raman mapping was performed using a DXRxi Raman imaging microscope (Thermo Fisher Scientific Inc., Hudson, USA) equipped with a 532 nm excitation laser. The laser power, exposure time, number of scans, and image pixel size equaled 6.0 mW, 0.00286 s, 11, and 5.0 μm, respectively. System control and spectrum acquisition were conducted using the ThermoScientific OMNIC software.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!