The largest database of trusted experimental protocols

300sb c8

Manufactured by Agilent Technologies
Sourced in Germany

The 300SB-C8 is a chromatography column designed for high-performance liquid chromatography (HPLC) analysis. It features a stationary phase composed of C8-modified silica particles, which are commonly used for the separation and analysis of a wide range of organic compounds. The column dimensions and packing material are optimized for efficient chromatographic separations.

Automatically generated - may contain errors

6 protocols using 300sb c8

1

Optimized HPLC Separation for Proteins

Check if the same lab product or an alternative is used in the 5 most similar protocols
The mobile phase A contained 2% IPA, 98% Water with 0.1% TFA. The mobile phase B contained 70% IPA, 20% ACN, 10% Water with 0.1% TFA. The sample was diluted in formulation buffer without surfactant at 1 mg/mL. Detection was UV absorbance at 280 nm.
For column assessment, the same separation condition was used for the UPLC column (ACUITY Protein BEH C4) and SPP column (HALO Protein C4 400 Å and 1000 Å). The gradient was 25%–35% B in 8 min under a flow rate of 0.6 mL/min at 75 °C. The total run time including conditioning and washing is 12 min and the injection quantity was 2.5 μL at 1 mg/mL protein. For the reference standard HPLC column (ZORBAX 300SB-C8), the gradient elution employed 25%–40% B in 20 min (0.75% B/min.) at the flow rate of 0.6 mL/min and column temperature of 75 °C.
An improved separation method for the HALO Protein C4 column used the gradient of 25% to 33% B in 7.5 min to (1.07% B/min.) achieve better resolution and throughput. All the other conditions such as flow rate, temperature and sample loading amount were varied during the optimization process to explore effects on separation qualities.
For the redox enriched samples and tagged samples analysis, the final optimized conditions were used, as described in the figures.
+ Open protocol
+ Expand
2

Quantitative UV-VIS and LC-MS Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
UV-VIS spectrophotometric assays were performed on a BioTek Synergy HT Microplate Reader. LC-ESI-MS analysis was carried out using a Waters single quadrupole mass spectrometer and Acquity H Class UPLC with photodiode array detector. LC-ESI-MS samples were run on a 10 min gradient, eluting with increasing acetonitrile containing 0.1% formic acid over a C8 column (Zorbax 300SB-C8).
+ Open protocol
+ Expand
3

Quantitative UV-VIS and LC-MS Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
UV-VIS spectrophotometric assays were performed on a BioTek Synergy HT Microplate Reader. LC-ESI-MS analysis was carried out using a Waters single quadrupole mass spectrometer and Acquity H Class UPLC with photodiode array detector. LC-ESI-MS samples were run on a 10 min gradient, eluting with increasing acetonitrile containing 0.1% formic acid over a C8 column (Zorbax 300SB-C8).
+ Open protocol
+ Expand
4

Rapid Reversed-Phase HPLC Separation

Check if the same lab product or an alternative is used in the 5 most similar protocols
Chromatographic separation was performed on an Agilent Technologies 1260 Infinity BioInert HPLC quaternary pump system equipped with a diode array detector (DAD VL+). A reverse-phase C8 column (Zorbax 300SB-C8 Rapid Resolution 2.1 x 50 mm, 3.5 μm) and C8 guard column (Zorbax 300SB-C8, 2.1 x 12.5 mm, 5μm) were used. The following gradient elution was used (time/%B): 0/30, 2/30, 3.5/65, 4/70, 5/75, 6.5/95, 9.5/95, and 10/30, with a post time of 5 minutes. The following HPLC running conditions were used: injection volume was 40 μl, flow rate was maintained at 0.4 ml/min, detection was performed at 280 nm and column temperature was 800C. To reduce the carry over, 100 μl of D-PBS 1X was injected between sample or standard injections.
+ Open protocol
+ Expand
5

Screening Antioxidant Activity via HPLC-ABTS

Check if the same lab product or an alternative is used in the 5 most similar protocols
An on-line screening method for antioxidative activity was developed based on the method of Koleva et al., (2000), as given in Fig. 1. Samples were subjected to reverse phase-HPLC (Ultimate 3000 RSLC HPLC, Dionex, Germany) to allow separation in the column (Zorbax, 3.5 um, 300SB-C8, 2.1*50 mm, Agilent). Each peak was detected by a diode array detector (DAD) at 280 nm. Then 2,2'-azino-bis(3-ethylbenzothiazoline-6-sulphonic acid) (ABTS) radical solution was pumped into the separated fractions on-line at a flow rate of 0.5 mL/min. A peak showing a negative absorbance at 734 nm after reacting with the ABTS reagent was indicative of antioxidant activity. Antioxidative activity selection was done using an on-line ABTS reaction assay by pumping an ABTS radical solution into the separated fractions on-line with a flow rate of 0.5 mL/min. The peak showing negative absorbance at 734 nm after reaction with the ABTS reagent indicates antioxidative activity. Among the strains tested, the strain Lactococcus lactis SL6 was the one showing the antioxidative activity using the ABTS method and was thus used for the next experiments.
+ Open protocol
+ Expand
6

Protein MS Analysis using Agilent Q-TOF

Check if the same lab product or an alternative is used in the 5 most similar protocols
MS analysis of proteins was performed on an Agilent 6530 ESI quadrupole time-of-flight (Q-TOF) mass spectrometer (41 (link)). Briefly, liquid chromatography (LC)–MS analysis was performed using an Agilent Q-TOF mass spectrometer in line with an Agilent 1290 HPLC system. Five microliters of protein (about 1 μg/μl) was loaded onto a reversed-phase column (300SB-C8, 2.1 mm × 50 mm, 3.5-μm particle) (Agilent Technologies, Santa Clara, CA). The sample was then eluted over a gradient (2% B for 2 min to waste, 2 to 50% B for 6 min, 50 to 90% B for 4 min, 90% B for 4 min, and then decreased to 2% B for 1.1 min, where B is 100% acetonitrile, 0.1% formic acid and A is water with 0.1% formic acid) at a flow rate of 0.2 ml/min and introduced online into the Q-TOF mass spectrometer using electrospray ionization. MS data were analyzed by MassHunter qualitative software with Biocom firm workflow.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!