The largest database of trusted experimental protocols

Picotip emitter

Manufactured by New Objective
Sourced in United States

The PicoTip emitter is a laboratory equipment designed for sample preparation. It is used to produce a fine spray of liquid samples for analysis. The device features a small-diameter tip that generates a controlled, consistent flow of the sample material.

Automatically generated - may contain errors

62 protocols using picotip emitter

1

Phosphopeptide Analysis by LC-MS/MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
Phosphopeptide-enriched samples were analyzed by LC-MS/MS on an Easy-nLC 1000 (ThermoFisher) coupled to an LTQ-Orbitrap Fusion mass spectrometer (ThermoFisher) operated in positive ion mode. The LC system, configured in a vented format consisted of a fused-silica nanospray needle (PicoTip emitter, 50μm ID x 20cm, New Objective) packed in-house with ReproSil-Pur C18-AQ, 3μm and a trap (IntegraFrit Capillary, 100μm ID x 2 cm, New Objective) containing the same resin as in the analytical column with mobile phases of 0.1% FA in water (A) and 0.1% FA in MeCN (B). The peptide sample was diluted in 20 µL of 0.1% FA, 3% MeCN, and 8.5μL was loaded onto the column and separated over 210 minutes at a flow rate of 300 nL/min with a gradient from 5 to 7% B for 2 minutes, 7 to 35% B for 150 minutes, 35 to 50% B for 1 minute, hold 50% B for 9 minutes, 50 to 95% B for 2 minutes, hold 95% B for 7 minutes, 95 to 5% B for 1 minute, re-equilibrate at 5% B for 38 minutes. A spray voltage of 2000 V was applied to the nanospray tip. MS/MS analysis occurred over a 3 second cycle time consisting of 1 full scan MS from 350–1500 m/z at resolution 120,000 followed by data dependent MS/MS scans using HCD activation with 27% normalized collision energy of the most abundant ions. Selected ions were dynamically excluded for 45 seconds after a repeat count of 1.
+ Open protocol
+ Expand
2

LC-MS/MS Analysis of Peptide Samples

Check if the same lab product or an alternative is used in the 5 most similar protocols
LC-MS/MS analysis was performed with an Easy-nLC 1000 (Thermo Scientific) coupled to an Orbitrap Elite mass spectrometer (Thermo Scientific). The LC system configured in a vented format102 (link) consisted of a fused-silica nanospray needle (PicoTip™ emitter, 75 µm ID, New Objective) packed in-house with Magic C18 AQ 100 Å reverse-phase media (Michrom Bioresources Inc.) (25 cm), and a trap (IntegraFrit™ Capillary, 100 µm ID, New Objective) containing Magic C18 AQ 200 Å (2 cm). The peptide sample was diluted in 10 µL of 2% acetonitrile and 0.1% formic acid in water and 8 µL was loaded onto the column and separated using a two-mobile-phase system consisting of 0.1% formic acid in water (A) and 0.1% acetic acid in acetonitrile (B). A 90 min gradient from 7% to 35% acetonitrile in 0.1% formic acid at a flow rate of 400 nL/minute was used for chromatographic separations. The mass spectrometer was operated in a data-dependent MS/MS mode over the m/z range of 400–1800. The mass resolution was set at 240,000. For each cycle, the 20 most abundant ions with +2 and +3 charges states from the scan were selected for MS/MS analysis using 35% normalized collision energy. Selected ions were dynamically excluded for 15 seconds.
+ Open protocol
+ Expand
3

Mass Spectrometric Analysis of Peptides

Check if the same lab product or an alternative is used in the 5 most similar protocols
Mass spectrometric analysis was basically carried out as described in Schwenk et al. (2014) (link). Briefly, peptides obtained from tryptic in-gel digests were dissolved in 0.5% trifluoroacetic acid and aliquots were loaded onto a C18 PepMap100 precolumn (particle size 5 mm; Dionex/Thermo Scientific, Germany) with 0.5% (v/v) acetic acid using a split-based UltiMate 3000 HPLC (Dionex/Thermo Scientific, Germany). Bound peptides were eluted and separated with an aqueous-organic gradient from 0.5% (v/v) acetic acid to 0.5% (v/v) acetic acid in 80% (v/v) acetonitrile (1 h 20′ total) in a PicoTip emitter (i.d. 75 mm; tip 8 mm; New Objective, United States) manually packed with ReproSil-Pur 120 ODS-3 (C18; particle size 3 mm; Dr. Maisch HPLC, Germany) and electrosprayed (2.3 kV; transfer capillary temperature 250°C) into an LTQ Orbitrap XL tandem mass spectrometer with the described settings.
+ Open protocol
+ Expand
4

Quantitative Mass Spectrometry of Analytes

Check if the same lab product or an alternative is used in the 5 most similar protocols
Desalted samples reconstituted in 0.1% formic acid were delivered into a QTRAP 6500 (Sciex) via a Dionex U3000 nano-LC system (Thermo) mounted with a NanoAcquity 5μm, 180 μm x 20 mm C18 trap and 1.7 μm, 75 μm X 100 mm analytical column (Waters)maintained at 40°C. A gradient of 2−50% acetonitrile/0.1% formic acid (v/v) over 45 minutes was applied to the columns at a flow rate of 300 nL/minute.The NanoSpray III source of the mass spectrometer was fitted with a 10 μm inner diameter PicoTip emitter (New Objective). The mass spectrometer was operated in positive ion mode using Analyst TF1.6 software (Sciex) and the MIDAS approach (MRM-initiated detection and sequencing) was used to quantify and confirm the identity of the analytes of interest. The optimized transitions are shown in Supplementary Table 1; dwell time for each transition was 20ms. The charge status of each precursor ion was determined using an enhanced resolution scan at 250Da/s and up to 3 MS/MS scans at 10,000Da/s were triggered with dynamic fill time. This gave a total cycle time of 3.9s.
+ Open protocol
+ Expand
5

Phosphopeptide Profiling by LC-MS/MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
Phosphopeptide-enriched samples were analyzed by LC-MS/MS on an Easy-nLC 1000 (ThermoFisher) coupled to an LTQ-Orbitrap Fusion mass spectrometer (ThermoFisher) operated in positive ion mode. The LC system, configured in a vented format consisted of a fused-silica nanospray needle (PicoTip emitter, 50μm ID × 20cm, New Objective) packed in-house with ReproSil-Pur C18-AQ, 3μm and a trap (IntegraFrit Capillary, 100μm ID × 2 cm, New Objective) containing the same resin as in the analytical column with mobile phases of 0.1% FA in water (A) and 0.1% FA in MeCN (B). The peptide sample was diluted in 20 μL of 0.1% FA, 3% MeCN, and 8.5 μL was loaded onto the column and separated over 210 minutes at a flow rate of 300 nL/min with a gradient from 5 to 7% B for 2 minutes, 7 to 35% B for 150 minutes, 35 to 50% B for 1 minute, hold 50% B for 9 minutes, 50 to 95% B for 2 minutes, hold 95% B for 7 minutes, 95 to 5% B for 1 minute, re-equilibrate at 5% B for 38 minutes. A spray voltage of 2000 V was applied to the nanospray tip. MS/MS analysis occurred over a 3 second cycle time consisting of 1 full scan MS from 350–1500 m/z at resolution 120,000 followed by data dependent MS/MS scans using HCD activation with 27% normalized collision energy of the most abundant ions. Selected ions were dynamically excluded for 45 seconds after a repeat count of 1.
+ Open protocol
+ Expand
6

LC-MS/MS Analysis of Peptides

Check if the same lab product or an alternative is used in the 5 most similar protocols
LC–MS/MS was performed in a Dionex UltiMateTM 3000 RSLCnano coupled to a LTQ Orbitrap XL mass spectrometer via a nanospray ion source (Thermo Fisher Scientific, Auckland, New Zealand). Peptides were fractioned on a PepMap C18 column (3 μm, 300 Å, 75 μm × 15 cm; ThermoFisher Scientific, Auckland, New Zealand) on a 350 min gradient from 0 to 80% ACN in 0.1% FA at a constant flow rate of 300 nL/min and eluted into the Orbitrap via a PicoTip emitter (360 × 20 μm; New Objective, Littleton, MA, USA) at a voltage set to 1.8 kV through a transfer tube of 25 μm inner diameter [31 (link),32 (link),33 (link)]. The six most intense peptide ions from the MS scan were selected and fragmented using collision-induced dissociation (normalised collision energy, 35%; activation Q, 0.250; and activation time, 30 ms) for MS/MS scans. Dynamic exclusion was used with the following settings: repeat count, 2; repeat duration, 30 s; exclusion list size, 500; exclusion duration, 90 s [32 (link)]. The spectra was acquired using Xcalibur (version 2.1.0 SP1, Thermo Fisher Scientific, Auckland, New Zealand).
+ Open protocol
+ Expand
7

Phosphopeptide Enrichment and LC-MS/MS Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
Phosphopeptide enrichment and LC-MS/MS were carried out as described
previously (Juvvadi et al., 2013 (link); Juvvadi et al., 2015 (link); Shwab et al., 2017 (link)). GFP-Trap® affinity
purified protein was processed for TiO2 phosphopeptide enrichment and
mass spectrometry. Proteolytic digestion was accomplished by the addition of 500
ng sequencing grade trypsin (Promega, Madison, WI) directly to the resin, with
incubation at 37°C for 18 h. Peptides were subjected to phosphopeptide
enrichment using a 10 μl GL Sciences TiO2 Spin Tip. The dried
phosphopeptide enriched samples were resuspended and subjected to
chromatographic separation on a Waters NanoAquity UPLC equipped with a 1.7
μm HSS T3 C18 75 μm I.D. x250 mm reversed-phase column. The
analytical column was connected to a fused silica PicoTip emitter (New
Objective, Cambridge, MA) with a 10 µm tip orifice.
Phosphopeptide-enriched samples were analyzed on a QExactive Plus mass
spectrometer using a data-dependent mode of acquisition. MS/MS spectra of the 10
most abundant precursor ions were acquired with a CID energy setting of 27 and a
dynamic exclusion of 20 s was employed for previously fragmented precursor
ions.
+ Open protocol
+ Expand
8

LC-MS/MS Analysis of Peptide Samples

Check if the same lab product or an alternative is used in the 5 most similar protocols
Samples were resuspended in 10 μl 0.1% FA and 6 μl were loaded onto an EASY-nano LC system (Thermo Fisher Scientific, Germany). The analytical column was a silica capillary (75 μm* 16 cm Pico Tip Emitter, New Objective, USA) packed in house with C18 ReproSil-Pur 1.9 μm (Dr. Maisch GmbH, Germany). Peptides were separated using a 60 min LC gradient from 5% to 25% solvent B (80% ACN, 0.1% FA) and continuously sampled by a Q-Exactive HF-X Mass Spectrometer (Thermo Fisher Scientific, Germany) through an electrospray interface. Data were acquired using data-dependent acquisition (DDA) in positive ion mode. Precursor spectra (375 to 1500 m/z) were acquired at 120,000 resolution with automatic gain control (AGC, MS1 target 3x106) and a maximum injection time of 50 ms. The 20 most abundant ion peptides were continuously selected for fragmentation. Fragmentation spectra were acquired at 15,000 resolution with an AGC target of 1x105 ions and a maximum injection time of 20 ms. Isolation width for fragmentation was set to 1.2 m/z.
+ Open protocol
+ Expand
9

Mass Spectrometry Proteomics Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
Mass spectrometry analyses were carried in triplicates for each sample with a nano-HPLC Proxeon (Thermo Scientific, San Jose, CA, USA) which allows in-line liquid chromatography with the capillary column, 60 μm × 100 mm (Pico Tip™ EMITTER, New Objective, Woburn, MA) filled with C18 resin of 5 mm diameter and 200Ǻ pores sizes (Michrom BioResources, Auburn, CA) linked to the mass spectrometer (LTQ-Velos, Thermo Scientific, San Jose, CA, USA) using an electrospray ionization in a survey scan in the range of m/z values 390–2000 tandem MS/MS.
The equivalent of 20 μg of each sample already dried by rotary evaporator was re-suspended in 20 μg of 0.1% formic acid and then subjected to reversed-phase LC-ESI-MS/MS. The nano-flow reversed-phase HPLC was developed with linear gradient of 85 min ranging from 0 to 100% of solvent B (97.5% acetonitrile, 0.1% formic acid) at a flow rate of 200 nl/min with a maximum pressure of 280 bar. Electrospray voltage and the temperature of the ion transfer capillary were 1.8 kV and 250 °C respectively.
+ Open protocol
+ Expand
10

Nano-scale Liquid Chromatography-Mass Spectrometry

Check if the same lab product or an alternative is used in the 5 most similar protocols
Desalted fractions were reconstituted in 40 microliters of 0.1% formic acid, and 5 microliter aliquots were delivered into a TripleTOF 5600 system (AB SCIEX) via an Eksigent NanoUltra cHiPLC system (AB SCIEX) mounted with a microfluidic trap and an analytical column (15 cm × 75 micrometer) packed with ChromXP C18-CL (3 μm). A NanoSpray III source was fitted with a 10 μm inner diameter PicoTip emitter (New Objective, Woburn, MA). The trap column was washed with 2% ACN and 0.1% formic acid for 10 min at 2 microliter/min before switching in-line with the analytical column. A gradient of 2–50% ACN and 0.1% (v/v) formic acid over 90 min was applied to the column at a flow rate of 300 nl/min. Spectra were acquired automatically in positive ion mode using information-dependent acquisition powered by Analyst TF 1.5.1 software (AB SCIEX). Up to 25 MS/MS spectra were acquired per cycle (10 Hz) using a threshold of 100 counts/s and with dynamic exclusion for 12 s. The rolling collision energy was increased automatically by selecting the iTRAQ check box in Analyst and manually by increasing the collision energy intercepts by 5.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!