The largest database of trusted experimental protocols

30 protocols using particle size analyzer

1

Physicochemical Characterization of Solid Lipid Nanoparticles

Check if the same lab product or an alternative is used in the 5 most similar protocols
Particle size analysis was conducted by dynamic light scattering (DLS), using a particle size analyzer (Brookhaven Instruments, Holtsville, NY, USA). Mean hydrodynamic diameter (Z-average), the size distribution and polydispersity index (PDI) of the nanoparticles in suspension were assessed by this technique. Zeta potential was determined by electrophoretic light scattering (ELS) using a ZetaPALS zeta potential analyzer (Brookhaven Instruments, Holtsville, NY, USA). Prior to the measurements, all samples were diluted (1:200) in PBS (pH 7.4) to yield a suitable scattering intensity. The average count rate was always between 100 and 500 kcps, showing that the dilution applied to the formulations was appropriate. The Z-average, PDI and zeta potential were obtained by calculating the average of ten runs of three independent batches of SLNs.
+ Open protocol
+ Expand
2

Particle Size and Zeta Potential Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
The mean size and PDI of the produced NPs were determined using a Particle Size Analyzer (Brookhaven Instruments Corporation; Software: Particle Sizing v.5 Brookhaven Instruments; Holtsville, NY, USA) operating at a scattering angle of 90°, at room temperature, with dust cut-off set to 30 and refractive index of 1.33. ζ potential was measured using an electrode and zeta potential analyzer (ZetaPALS, Brookhaven Instruments Corporation, Software: PALS Zeta Potential Analyser v.5, Brookhaven Instruments; Holtsville, NY, USA).
+ Open protocol
+ Expand
3

Characterization of PDAPEI/pDNA Polyplexes

Check if the same lab product or an alternative is used in the 5 most similar protocols
Particle-Size Analyzer (Brookhaven Instruments) was utilized to measure the particle size and PDI of PDAPEI/pDNA polyplexes at various ratios of w/w, whereas zeta potential was measured by Zeta Potential Analyzer (Zetasizer Nano, Malvern Instruments). All these measurements were under the temperature of 37°C. The morphology of PDAPEI/pDNA polyplex at 40 w/w ratio was observed by Transmission Electron Microscope (TEM, JEOL JEM 2010 system).
+ Open protocol
+ Expand
4

Optimizing PDAPEI/pDNA Complexes for Gene Delivery

Check if the same lab product or an alternative is used in the 5 most similar protocols
PDAPEI/pDNA complexes were prepared at different weight/weight (w/w) ratios of 1–50. First, PDAPEI was dissolved with deionized water (2 mg/ml) and pDNA (2 mg/ml). Then, the PDAPEI solution was diluted to different concentrations and added rapidly into the pDNA solution. Next, the solutions were incubated at room temperature for 30 min, and PEI/pDNA complexes and naked pDNA were simultaneously prepared as controls.
Agarose gel electrophoresis assay was used to evaluate the condensation ability of PDAPEI/pDNA complexes at different w/w. The particle size and zeta potential of PDAPEI/pDNA complexes were measured using a particle size analyzer (Brookhaven Instruments) and a zeta potential analyzer (Zetasizer Nano, Malvern Instruments) at different w/w ratios. Zeta potential across a range of pH (5.4, 6.4, 7.4, 8.4, 9.4) was measured to analysis its stability and degradation. The morphology of the PDAPEI/pDNA complexes was observed by Transmission electron microscopy (JEOL JEM 2010 system).
+ Open protocol
+ Expand
5

Particle Size Characterization of Nanomaterials

Check if the same lab product or an alternative is used in the 5 most similar protocols
The particle sizes of the HANPs and AgNPs used in this study were measured using a particle-size analyzer (Brookhaven Instruments). In addition, an FEI Sirion 200 scanning electron microscope (SEM; United States) was used to obtain images of HANP, AgNP, PM, SR-PM, SR-PM-Ag, and SR-PM-Ag-HA samples. All specimens were coated by gold evaporation at 5 keV under an argon atmosphere before SEM tests.
+ Open protocol
+ Expand
6

Characterization of Lipid Nanoparticles

Check if the same lab product or an alternative is used in the 5 most similar protocols
The hydrodynamic diameter and PDI were measured using a Particle Size Analyzer (Brookhaven Instruments Corporation, Software: Particle Sizing v.5 Brookhaven Instruments, Holtsville, New York, USA). The zeta potential of the developed LNPs was determined using the Zeta Potential Analyzer (ZetaPALS, Brookhaven Instruments Corporation, Software: PALS Zeta Potential Analyzer v.5 Brookhaven Instruments, Holtsville, New York, USA). The system operated with an incidence light angle of 90˚, at room temperature. Prior to the measurements, the LNPs were diluted (1:100) in type I water.
The morphology of the produced LNPs was assessed by TEM. For this purpose, the formulations were diluted (1:100) in type I water. Then, 10 µL of the diluted LNPs were placed onto a copper-mesh grid for 2 min. The excess of LNPs was then removed, and 10 µL of 0.75% (w/v) uranyl acetate solution were added to the grid for 30 s (negative staining). The samples were observed in a JEM-1400 transmission electron microscope (JEOL Ltd., Tokyo, Japan) with an acceleration voltage of 80 kV.
+ Open protocol
+ Expand
7

Nanoparticle Size and Charge Characterization

Check if the same lab product or an alternative is used in the 5 most similar protocols
The mean particle size and polydispersity index (PdI) were evaluated using dynamic light scattering (DLS) at 659 nm, with a detection angle of 90°, on a particle size analyzer (Brookhaven Instruments, Long Island, NY, USA), at 25 °C. The zeta potential measurements were performed on the same equipment, applying a field strength applied around 5.9 V cm−1 and using Zeta Plus® software Particle Sizing version 3.95. Samples were diluted 1:50 (v/v) with purified water using ten measurements in triplicate. Data were expressed as mean ± standard deviation (SD).
+ Open protocol
+ Expand
8

Liposome Particle Characterization by DLS

Check if the same lab product or an alternative is used in the 5 most similar protocols
The particle size distribution and polydispersity index (PDI) of the liposome particles was determined at 25˚C by dynamic light scattering (DLS) using a ZetaPALS zeta potential and particle size analyzer (Brookhaven Instruments, Holtsville, NY, USA). Each parameter was measured three times, after which we calculated average values and standard deviations. L-DOX morphology was investigated by cryogenic transmission electron microscopy (Cryo-TEM, FEI Tecnai 20, type Sphera, Hillsboro, OR, USA).
+ Open protocol
+ Expand
9

Polyplex Characterization by Particle Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
The particle size and zeta potential of PDAPEI/pDNA polyplexes were measured at different w/w ratios by Particle-Size Analyzer (Brookhaven Instruments) and Zeta Potential Analyzer (Zetasizer Nano, Malvern Instruments), respectively at 25 °C. The morphology of PDAPEI/pDNA polyplexes was imaged by transmission electron microscopy (JEOL JEM 2010 system) at the w/w ratio of 2.
+ Open protocol
+ Expand
10

Synthesis and Characterization of PVO Nanoparticles

Check if the same lab product or an alternative is used in the 5 most similar protocols
PVO was synthesized from the reaction of oxalyl chloride, and acid-cleavable vanillin derivative as previously reported [23 (link)]. In brief, and acid-cleavable vanillin derivative (3.941 mmol) was placed in a flask containing 25 mL dried dichloromethane and pyridine (9.8 mmol). The flask was placed in ice bath and added with oxalyl chloride (3.941 mmol). The reaction was continued for 6 h at room temperature. PVO was obtained from extraction using dichloromethane/water and precipitated in cold hexane. After purification, the chemical structure of PVO was verified by NMR. PVO nanoparticles were prepared by a single emulsion method as previously reported [23 (link)]. Indocyanine green (ICG)-loaded PVO nanoparticles were also prepared for fluorescence imaging. Unloaded ICG was removed by centrifugation twice. The PVO nanoparticles suspended in water were characterized using a scanning electron microscopy (SUPRA40VP, Carl Zeiss, Jena, Germany) and particle size analyzer (Brookhaven Instruments, Holtsville, NY, USA). UV-vis absorption of the ICG-loaded PVO nanoparticles was investigated using a spectrometer (S-3100, Scinco, Seoul, Korea).
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!