The largest database of trusted experimental protocols

Rxi 624sil ms column

Manufactured by Restek
Sourced in United States

The Rxi-624Sil MS column is a gas chromatography (GC) column designed for the analysis of a wide range of volatile and semi-volatile compounds. It features a 624-type stationary phase with a silphenylene polymer coating, providing selectivity for the separation of a diverse set of compounds.

Automatically generated - may contain errors

7 protocols using rxi 624sil ms column

1

Quantification of 1,3-Dichlorobenzene in Whole Blood

Check if the same lab product or an alternative is used in the 5 most similar protocols
1 ml of whole blood was transferred into 20-ml headspace vials with Teflon-laminated butyl rubber septa. 1,3-DCB was determined with a TurboMatrix HS 40 Trap headspace injector (Perkin-Elmer, Rodgau, Germany) working in multiple headspace extraction mode (Kolb and Ettre 1991 (link)). This headspace injector is directly coupled to a Thermo Trace GC Ultra/Thermo DSQ II GC/MS system (Thermo Fischer, Waltham, MA, USA) working in single ion monitoring mode (SIM mode) using m/z 146 as specific mass. The temperature of the headspace vial was 80 °C for 50 min with a fivefold multiple headspace extraction. The starting temperature of the GC was 150 °C for 1 min, increased by 10 °C/min until a final temperature of 230 °C was held for 0.01 min. Chromatography was performed on a Rxi-624 Sil MS column (60 m × 0,31 mm ID; 1,8 µm film thickness; Restek, Bad Homburg, Germany) using Helium 5.0 as a carrier gas with a constant pressure of 137.9 kPa.
Quantification was performed using an external calibration in sheep blood in the range of 0.5–20 µg/L. The limit of quantification–based on a signal-to-noise ratio of 6 was determined to be 0.1 µg/L.
+ Open protocol
+ Expand
2

Characterization of Crumpled Graphene

Check if the same lab product or an alternative is used in the 5 most similar protocols
Surface morphology of the crumpled GO or rGO samples was investigated using a field emission scanning electron microscope (SEM) (LEO 1530 VP) operating at 10.0 kV for low-, medium- and high-resolution imaging. Before the SEM imaging, the crumpled graphene structures were coated with a layer of AuPd (~2 nm). Transmission electron microscopy (TEM) was performed using a JEOL 2100F TEM/STEM at an acceleration voltage of 200 kV with GO nanosheets on lacey carbon grids. TCE containing aqueous samples were analyzed using a Shimadzu GC-2010 with a Restek Rxi-624Sil MS column following the US EPA 551.1 method. The interlayer spacing before and after compression were identified by X-ray diffraction spectrometry (XRD) on a Bruker AXS D8 Advance instrument with Cu KR radiation (λ = 1.5418 Å). The change of resistance of rGO-PDMS samples in response to chemical exposure was measured by using a portable standard multimeter (Fluke).
+ Open protocol
+ Expand
3

GC/MS Analysis of E-Liquid Flavors

Check if the same lab product or an alternative is used in the 5 most similar protocols
The extracted e-liquid was analyzed using previously described GC/MS methods4 (link). Each sample (50 µL) was dissolved in 0.95 mL of isopropyl alcohol and shipped overnight on dry ice to Portland State University. A 20 µL aliquot of internal standard solution (2000 ng/µL of 1,2,3- trichlorobenzene dissolved in isopropyl alcohol) was added to each diluted sample prior to analysis. Using internal-standard-based calibration procedures described elsewhere36 (link) analyses for 178 flavor-related target analytes, two synthetic coolants, and nicotine were performed with an Agilent 5975C GC/ MS system (Santa Clara, CA). A Restek Rxi-624Sil MS column (Bellefonte, PA) was used (30 m long, 0.25 mm id, and 1.4 µm film thickness). A 1.0 µL aliquot of the diluted sample was injected into the GC with a 10:1 split. The injector temperature was 235 °C. The GC temperature program for analyses was 40 °C hold for 2 min, 10 °C/min to 100 °C, then 12 °C/min to 280 °C and hold for 8 min at 280 °C, and then 10 °C/min to 230 °C. The MS was operated in the electron impact ionization mode at 70 eV in the positive-ion mode. The ion source temperature was 220 cc, and the quadrupole temperature was 150 cc. The scan range was 34 to 400 amu. Each of the 181 (178 flavor chemicals, 2 synthetic coolants, and nicotine) target analytes were quantitated using the authentic standard material.
+ Open protocol
+ Expand
4

GC-MS Analysis of Volatile Organic Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
GC-MS analysis relied on an Agilent 8890/7079B GC-MS (Agilent, Santa Clara, CA, USA). The GC injector was equipped with an inert SPME liner (inner diameter 0.75 mm, Supelco, Bellefonte, ON, Canada) and operated in the splitless mode (0.75 min) followed by split mode with a ratio of 1:50. Extracted compounds were separated using a Rxi-624Sil MS column (30 m × 0.32 mm, layer thickness 1.8 μm, Restek, Bellefonte, PA, USA) operated in constant helium flow at 1.4 mL min-1. The column temperature program was as follows: 37 °C for 12 min, followed by 5 °C min-1 up to 150 °C, then 10 °C min−1 up to 290 °C, and finally remaining at 290 °C for 8 min. The untargeted VOC analysis relied on the mass spectrometer working in a SCAN mode with the associated m/z ranging from 20 up to 250. The peak integration was based on extracted m/z ratio chromatograms, and such an approach allowed for a separation of the majority of peaks of interest from their neighbors. The quadrupole rods, ion source, and transfer line were kept at 150 °C, 230 °C, and 280 °C, respectively.
VOC identification was performed using a two-step process. First, the spectrum of a peak was checked against the NIST mass spectral library database. Next, the NIST identification was confirmed by comparing the retention times of peaks of interest with retention times obtained for reference standards prepared as outlined above.
+ Open protocol
+ Expand
5

GC-MS Analysis of Refill Fluids

Check if the same lab product or an alternative is used in the 5 most similar protocols
For each refill fluid, 50 µl were dissolved in 0.95 ml of isopropyl alcohol (IPA) (Fisher Scientific, Fair Lawn, NJ). All diluted samples were shipped overnight on ice to Portland State University and analyzed using GC-MS on the day they were received. A 20 µl aliquot of internal standard solution (2000 ng/µl of 1, 2, 3-trichlorobenzene dissolved in IPA) was added to each diluted sample before analysis. Using internal standard-based calibration procedures described elsewhere52 (link), analyses were performed with an Agilent 5975 C GC-MS system (Santa Clara, CA). A Restek Rxi-624Sil MS column (Bellefonte, PA) was used (30 m long, 0.25 mm id, and 1.4 µm film thickness). A 1.0 µl aliquot of diluted sample was injected into the GC with a 10:1 split. The injector temperature was 235 °C. The GC temperature program for analyses was: 40 °C hold for 2 min; 10 °C/min to 100 °C; then 12 °C/min to 280 °C and hold for 8 min at 280 °C, then 10 °C/min to 230 °C. The MS was operated in electron impact ionization mode at 70 eV in positive ion mode. The ion source temperature was 220 °C and the quadrapole temperature was 150 °C. The scan range was 34 to 400 amu. Each of the 178 target analytes was quantitated using authentic standard material and an internal standard compound normalized multipoint calibration.
+ Open protocol
+ Expand
6

GC-MS Analysis of Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
The GC-MS analysis was carried out using an Agilent 8890/7079B GC-MS system. The GC injector contained an SPME liner (with an inner diameter of 0.75 mm, supplied by Supelco in Canada) and was maintained at 290 °C and operated in splitless mode for 0.75 min, followed by split mode with a ratio of 1:50. The extracted compounds were separated using an Rxi-624Sil MS column (30 m × 0.32 mm × 1.8 μm, Restek in Bellefonte, PA, USA) under constant helium flow at a rate of 1.4 mL/min. The column temperature program started at 37 °C for 12 min, then ramped up by 5 °C per minute to 150 °C, next ramped up by 10 °C per minute to 290 °C, and finally remained at 290 °C for 8 min. The mass spectrometer operated in the SCAN mode with an associated mass to charge ratio (m/z) range from 20 to 250. The peak integration process relied on extracted m/z ratio chromatograms. By using this approach, the majority of the peaks of interest were effectively separated from neighboring compounds. Throughout the analysis, the quadrupole, ion source, and transfer line were carefully maintained at temperatures of 150 °C, 230 °C, and 280 °C.
+ Open protocol
+ Expand
7

GC-MS Analysis of Volatile Organic Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
GC-MS analysis relied on an Agilent 8890/7079B GC-MS (Agilent, Santa Clara, CA, USA). The GC injector was equipped with an inert SPME liner (inner diameter 0.75 mm, Supelco, Canada) and operated in the splitless mode (0.75 min), followed by split mode with a ratio 1:50. Extracted compounds were separated using an Rxi-624Sil MS column (30 m × 0.32 mm, layer thickness 1.8 μm, Restek, Bellefonte, PA, USA) operated in constant helium flow at 1.4 mL min−1. The column temperature programme was as follows: 37 °C for 12 min, followed by 5 °C min−1 up to 150 °C, then 10 °C min−1 up to 290 °C, and finally remaining at 290 °C for 8 min. The untargeted VOC analysis relied on the mass spectrometer working in a SCAN mode with the associated m/z ranging from 20 up to 250. The peak integration was based on extracted m/z ratio chromatograms, and such an approach allowed for a separation of the majority of peaks of interest from their neighbours. The quadrupole, ion source, and transfer line were kept at 150 °C, 230 °C and 280 °C, respectively.
VOC identification was performed using a two-step process. First, the spectrum of a peak was checked against the NIST mass spectral library database. Next, the NIST identification was confirmed by comparing the retention times of peaks of interest with retention times obtained for reference standards prepared as outlined above.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!