The largest database of trusted experimental protocols

331 protocols using q2000

1

Thermal Analysis of Extruded Materials

Check if the same lab product or an alternative is used in the 5 most similar protocols
Standard DSC (Q2000; TA Instruments, New Castle, Delaware) analysis was performed at a heating rate of 10°C/min. Modulated temperature DSC (Q2000; TA Instruments) analysis was conducted using a heating rate of 2°C/min, amplitude ±0.318°C and a period of 60 s. Scans were carried out within the temperature range 0°C–220°C and TA standard crimped pans were used in all experiments. Nitrogen purge gas was used with a flow rate of 50 mL/min. Calibration was performed using n-octadecane, benzoic acid, indium, and tin. For each sample, measurements were repeated at least in triplicate.
Thermogravimetric analysis (TGA Q5000; TA Instruments) was used to measure the degradation temperature (Tdeg) of the raw materials and the water content of the fresh extrudates. Analyses were conducted at 10°C/min from room temperature to 300°C. Aluminum open pans were used.
+ Open protocol
+ Expand
2

Differential Scanning Calorimetry Analysis of Liquid Crystal Elastomers

Check if the same lab product or an alternative is used in the 5 most similar protocols

Example 4

Differential scanning calorimetry (“DSC”) was performed using a TA Instruments Q2000 (Texas Instruments, Inc., Dallas, Tex.) with an aluminum hermetic crucible. All tests were performed under nitrogen. All heating and cooling rates were set to 10° C./min. Samples that were tested before polymerization were heated from room temperature to 120° C., cooled to −50° C., and then heated to 200° C. Samples tested after crosslinking were heated from room temperature to 175° C., cooled to −50° C., and then heated to 200° C. Data shown are of the second heating cycle. Both the polymer and oligomer were analyzed using DSC.

The oligomer exhibits a glass transition at −12° C. and was not observed to crystallize over several months of storage at 4° C. A nematic-isotropic transition temperature was observed starting at 100° C. and was confirmed by POM. An unaligned sample, with a characteristic polydomain texture was cross-linked and then tested utilizing DSC. The resulting polymer network was viscoelastic at room temperature, with a glass transition of 19° C. (FIG. 30). A slight endothermic shift is observed at 155° C. This shift is characteristic of supercritical behavior of liquid crystal elastomers incapable of exhibiting complete clearing. This result is supported by POM observation of remnant order at 200° C.

+ Open protocol
+ Expand
3

Calorimetric Analysis of PCL-ES Structures

Check if the same lab product or an alternative is used in the 5 most similar protocols
The calorimetric behavior of PCL-ES structures was measured by differential scanning calorimetry (DSC) using TA Instruments Q2000 (TA Instruments; New Castle, DE, USA). Samples (4.71 mg for 10% PCL; 5.42 mg for 15% PCL) were contained in an aluminum pan under a dynamic nitrogen atmosphere (50 mL/min) at a heating rate of 10 °C/min from 30 to 100 °C.
+ Open protocol
+ Expand
4

Differential Scanning Calorimetry Analysis of Liquid Crystal Elastomers

Check if the same lab product or an alternative is used in the 5 most similar protocols

Example 4

Differential scanning calorimetry (“DSC”) was performed using a TA Instruments Q2000 (Texas Instruments, Inc., Dallas, Tex.) with an aluminum hermetic crucible. All tests were performed under nitrogen. All heating and cooling rates were set to 10° C./min. Samples that were tested before polymerization were heated from room temperature to 120° C., cooled to −50° C., and then heated to 200° C. Samples tested after crosslinking were heated from room temperature to 175° C., cooled to −50° C., and then heated to 200° C. Data shown are of the second heating cycle. Both the polymer and oligomer were analyzed using DSC.

The oligomer exhibits a glass transition at −12° C. and was not observed to crystallize over several months of storage at 4° C. A nematic-isotropic transition temperature was observed starting at 100° C. and was confirmed by POM. An unaligned sample, with a characteristic polydomain texture was cross-linked and then tested utilizing DSC. The resulting polymer network was viscoelastic at room temperature, with a glass transition of 19° C. (FIG. 30). A slight endothermic shift is observed at 155° C. This shift is characteristic of supercritical behavior of liquid crystal elastomers incapable of exhibiting complete clearing. This result is supported by POM observation of remnant order at 200° C.

+ Open protocol
+ Expand
5

Differential Scanning Calorimetry Analysis of Liquid Crystal Elastomers

Check if the same lab product or an alternative is used in the 5 most similar protocols

Example 4

Differential scanning calorimetry (“DSC”) was performed using a TA Instruments Q2000 (Texas Instruments, Inc., Dallas, Tex.) with an aluminum hermetic crucible. All tests were performed under nitrogen. All heating and cooling rates were set to 10° C./min. Samples that were tested before polymerization were heated from room temperature to 120° C., cooled to −50° C., and then heated to 200° C. Samples tested after crosslinking were heated from room temperature to 175° C., cooled to −50° C., and then heated to 200° C. Data shown are of the second heating cycle. Both the polymer and oligomer were analyzed using DSC.

The oligomer exhibits a glass transition at −12° C. and was not observed to crystallize over several months of storage at 4° C. A nematic-isotropic transition temperature was observed starting at 100° C. and was confirmed by POM. An unaligned sample, with a characteristic polydomain texture was cross-linked and then tested utilizing DSC. The resulting polymer network was viscoelastic at room temperature, with a glass transition of 19° C. (FIG. 30). A slight endothermic shift is observed at 155° C. This shift is characteristic of supercritical behavior of liquid crystal elastomers incapable of exhibiting complete clearing. This result is supported by POM observation of remnant order at 200° C.

+ Open protocol
+ Expand
6

Synthesis and Characterization of Styrene-Vinyl Phenoxy Benzocyclobutene Copolymer

Check if the same lab product or an alternative is used in the 5 most similar protocols

Example 8

Styrene (4.77 g) and Vinyl Phenoxy Benzocyclobutene (1.13 g) were dissolved in THF (3.98 g) along with V601™ initiator (70 mg) in an EZ Max™ 100 ml jacketed reactor (Mettler Toledo, Columbia, Md.) equipped with overhead stirring and nitrogen atmosphere. The solution was purged with nitrogen gas for 30 minutes, then heated to an internal temperature of 60° C. overnight. The resulting viscous solution was diluted with THF (20 ml) then precipitated into methanol (250 ml), filtered and dried overnight in vacuo to give the copolymer (4.23 g, 72% yield). Mn 36.6 k, Mw 79.1 k. The polymer curing kinetics were evaluated via differential scanning calorimetry (DSC, TA Instruments Q2000, TA instruments, New Castle, Del.) at a ramp rate of 2, 5, 10 and 20° C./min. The Kissinger method was used to determine a ring opening activation barrier of 24.2 kcal/mol. Thermal stability was evaluated using thermogravimetric analysis (TA Instruments Q5000) under a nitrogen atmosphere, wherein a solid polymer sample was placed in a TGA pan and run out to 400° C. at a rate of 10° C./min.

The TGA of the resulting copolymer exhibited a five percent weight loss value at 300° C.

+ Open protocol
+ Expand
7

Thermal Characterization of Materials

Check if the same lab product or an alternative is used in the 5 most similar protocols
Thermal characterization was carried out by a differential scanning calorimeter (DSC; TA instruments Q2000; TA instruments, New Castle, New Castle County, Delaware, USA) and by thermogravimetric analysis (TGA; TA instruments Q500; TA instruments, New Castle, New Castle County, Delaware, USA). DSC was performed using a heat–cool–heat procedure, with a heating rate of 10 °C/min until 200 °C and a cooling rate of 5 °C/min until −100 °C, under a nitrogen flow of 50 mL/min. All samples were analysed in Aluminium hermetic pans and sample weight varied between 6 and 10 mg. The second heating scan was used to evaluate the thermal properties for each sample. TGA measurements were performed in Platinum pans at a heat rate of 10 °C/min, from room temperature to 800 °C, under a nitrogen flow of 50 mL/min. TGA sample weight varied between 17 and 23 mg.
+ Open protocol
+ Expand
8

DSC Analysis of Thermal Events

Check if the same lab product or an alternative is used in the 5 most similar protocols

Example 10

Differential scanning calorimetry (DSC) study was conducted using a TA Instruments, Q2000 at a heating rate 10° C./min. The instrument was calibrated for temperature and energy using the melting point and enthalpy of fusion of an indium standard. Thermal events (desolvation, melting, etc.) were evaluated using Universal Analysis 2000 software, version 4.1D, Build 4.1.0.16. Samples were weighed into A1 pans and scans ran from ˜25° C. to ˜270° C. at a rate of 10° C./min.

+ Open protocol
+ Expand
9

Thermodynamic Analysis of PCL Powder

Check if the same lab product or an alternative is used in the 5 most similar protocols
The thermodynamic properties of the PCL powder are determined by using differential scanning calorimetry (DSC, TA, Q2000, United States) and thermogravimetric analysis (TGA, TA, Q5000, United States) with a heating rate of 10°C/min.
+ Open protocol
+ Expand
10

Thermal Analysis of Nanofiber Membranes

Check if the same lab product or an alternative is used in the 5 most similar protocols
The thermal behavior of nanofiber membranes was investigated using DSC (Q2000, TA Instruments, USA) under a nitrogen atmosphere. The temperature region was 20°C to 250°C. The heating rate was 10°C/min. The cooling rate was 5°C/min.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!