The largest database of trusted experimental protocols

Vct500

Manufactured by Leica camera

The VCT500 is a versatile laboratory equipment designed for a range of applications. It features a high-quality optical system and precise controls, enabling accurate measurements and analysis. The core function of the VCT500 is to provide reliable and consistent performance in laboratory settings.

Automatically generated - may contain errors

8 protocols using vct500

1

Cryo-SEM Imaging of Hydrogel Structure

Check if the same lab product or an alternative is used in the 5 most similar protocols
Hydrogel samples were sliced into pieces ∼3 mm wide and ∼0.5 mm thick, blotted on filter paper and mounted onto aluminium pins using optimal cutting temperature (OCT) compound, before being plunged frozen in liquid ethane (−180 °C; Leica, EM GP2). Frozen samples on pins were then loaded into a cryostage under liquid nitrogen and transferred under vacuum to the cryopreparation system (Leica, EM ACE600 + VCT500). Frozen samples were sublimated at –90 °C for 105 seconds to remove surface ice and reveal gel structure and then coated with 4 nm platinum. After coating, the samples were transferred under vacuum to the cryo-SEM (JEOL, IT800 field emission SEM + Leica, VCT500). Imaging was conducted at −155 °C and 10 kV using both secondary and back-scattered electron signals. Qualitative element analysis and mapping of silver was performed at 10 kV using an x-ray detector interfaced to AZtec software (Oxford Instruments, 170 mm2 Ultim® Max SDD EDS detector).
+ Open protocol
+ Expand
2

Electron Microscopy Characterization of SEI Films

Check if the same lab product or an alternative is used in the 5 most similar protocols
The same samples were coated with 4 nm Pt and then transferred to a scanning electron microscope (SEM) equipped with a field emission gun (Merlin, Carl Zeiss) using a Leica VCT500 air-tight transfer module. The operation voltage was 3 kV, and the beam current was 200 pA. With the same instrument, elemental analysis was also performed using an energy dispersive X-ray spectroscopy (EDS) detector. Additionally, the same samples were transferred similarly to a FIB/SEM instrument (XEIA3 Triglav, Xe-Plasma FIB, Tescan Orsay Holding). For minimal damage of the SEI film, a low current (0.25 nA) was used for FIB etching (at 90° angle) and the sample was cooled to –130 °C.
+ Open protocol
+ Expand
3

Freeze-Etching for Cryogenic Imaging

Check if the same lab product or an alternative is used in the 5 most similar protocols
As reported in previous work15 (link), a solution droplet was dipped with a micro pipette on a precooled tungsten post, which is located within a liquid nitrogen bath to create droplets on top (60–250 µm). Subsequently, the sample holder is transferred as fast as possible into the cooled body of the modified transfer shuttle VCT500 from Leica (T =  − 184 °C) and pumped to a pressure of 6 × 10−1 mbar. Shortly thereafter, the shuttle is attached to the high vacuum-coater (Leica EM ACE600), to carry out a freeze-etching process to remove ice crystals that were formed by the contact of the sample with air. By heating up the sample very precisely to a temperature of – 90 °C and a pressure of 9 × 10−7 mbar for 30 min, a sublimation process from solid ice to vapor occurs, which is necessary for the controlled removal of condensed ice from the sample. In addition, the Leica EM ACE600 high vacuum coater is useful to improve the vacuum condition inside the shuttle (10−4 mbar), which is necessary to enable the transfer of the sample into the Focused-Ion-Beam (FIB) Microscope.
+ Open protocol
+ Expand
4

Cryogenic Honey Droplet Preparation

Check if the same lab product or an alternative is used in the 5 most similar protocols
As sample material, natural honey from the company Langnese was used. Tungsten wires were used as substrate material and their rough surfaces were produced by applying a tensile force under cryogenic conditions43 (link),44 (link),66 (link). It’s well known that honey has a high viscosity and, therefore, sticks to different surfaces. In the case of honey, the rough tungsten wire, was initially dipped into the honey and afterwards quenched to cryogenic temperature into a liquid N2 bath. The obtained honey droplet with a diameter of 50 µm can be seen in Fig. 9B. This method cannot be used for low viscous liquids; therefore, a dedicated method was developed as reported in43 (link),44 (link),66 (link). After that, the sample holder was transferred as fast as possible into the body of the cooled transfer shuttle VCT500 from Leica (T = − 184 °C). This system allowed to transfer of samples at cryo-temperatures (T = − 184 °C) and under a low-pressure atmosphere (10–4 mbar) between the machines without contaminations from the environment.

FIB Preparation—above in (A) the annular milling process with a decreasing inner diameter and respective current is schematically shown. (B) Frozen honey droplet on top of the tungsten post. (AI) show the milling process with an annular pattern and decreasing inner diameter until a final tip with a radius < 100–200 nm (J) is obtained.

+ Open protocol
+ Expand
5

Microscopy Analysis of R. vilicus

Check if the same lab product or an alternative is used in the 5 most similar protocols
Trophozoites and cyst-like forms from R. vilicus were observed by microscopy (Olympus IX73 inverted microscope, 1,000x magnification) using a differential interference contrast. All sizes were measured using ImageJ (Schneider et al., 2012 (link)). The morphological comparison between trophozoite and cyst-like forms was also assessed by electron microscopy. For scanning electron microscopy (SEM), cells were cryofixed in nitrogen with VCM (Leica), sublimed and 4 nm wolfram coated in ACE 600 (Leica). Each step was performed under vacuum and samples were transferred from one station to another using VCT 500 (Leica). Samples were observed at −100°C in scanning electron microscope Teneo VolumeScope (Thermo Fisher) to 3.5 kV and 10 kV.
For transmission electron microscopy (TEM), amoebae were fixed with 2.5% glutaraldehyde and 2% osmium tetroxide. Dehydration was carried out using different concentrations of acetone. Epon resin were used for impregnation and embedding process. After cutting with ultramicrotome UC6 (Leica), samples were contrasted with uranyl acetate and lead citrate. Samples observations were proceeded on transmission electron microscope (JEOL 1010) at 80 keV. Images were acquired with Quemesa camera (Olympus).
+ Open protocol
+ Expand
6

Cryogenic Atom Probe Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
Using a custom-made atom probe56 (link) equipped with NOPA to continuously change wavelength in the range between 350 and 900 nm, measurements were conducted with a laser wavelength of 355 nm. The pulse length was chosen to 250 fs. The spot size in the focus point amounts to a diameter of 50 microns. As detector, a 120 mm diameter delay line detector with chevron MCP setup with an open area ratio (OAR) of 50% is used. The flight length amounts to 130 mm and the effective half angle of the field of view to 38°. The system is equipped with a custom-made cryo-transfer port to accept a standard VCT500 from Leica for the transfer of cryogenic samples. The exchange of the sample in the buffer chamber uses a PEEK isolated storage position to avoid melting of the sample during transfer. The obtained datasets were analysed and reconstructed using the Scito57 software package. Calculated mass spectra were exported into csv files and plotted using OriginPro58 for publication.
+ Open protocol
+ Expand
7

High-Pressure Freezing and Cryo-SEM Analysis of Cellulosic Biofilms

Check if the same lab product or an alternative is used in the 5 most similar protocols
High-pressure freezing, fracture, and cryogenic SEM imaging were performed to examine the inner structure of cellulosic biofilms formed at the various interfaces. The cellulosic biofilm samples were placed in specimen carriers and mounted on a sample holder. The samples were then frozen under high pressure using a high-pressure freezing system (Leica EM HPM100, Leica Microsystems Inc., Austria). This was followed by fracturing at -110 °C using a flat edge knife and sublimation at -95 °C for 5 min under the vacuum of 10-7 mbar in a freeze-fracture and etching machine (Leica EM ACE900). The samples were transferred to the SEM via a cryo-transfer system (Leica EM VCT500). High-vacuum secondary electron imaging was performed using an Apreo VS SEM at 500 V. During SEM imaging, the chamber temperature was maintained at -110 °C.
+ Open protocol
+ Expand
8

Cryogenic Atom Probe Microscopy

Check if the same lab product or an alternative is used in the 5 most similar protocols
The presented results were obtained by using a custom-made atom probe68 (link) operating at a laser wavelength of 355 nm. The pulse length accounts for 250 fs and the spot size diameter to 50 microns. The system is equipped with a 120 mm diameter delay line detector with an open area ratio (OAR) of 50%. The system was additionally equipped with a custom-made cryo-transfer port to accept a standard VCT500 from Leica for the transfer of cryogenic samples.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!