The largest database of trusted experimental protocols

Conflo 4

Manufactured by Thermo Fisher Scientific
Sourced in Germany, United States

The Conflo IV is an analytical instrument manufactured by Thermo Fisher Scientific. It is designed to be used in conjunction with isotope ratio mass spectrometers (IRMS) for the analysis of stable isotopes. The Conflo IV facilitates the introduction of gaseous samples into the IRMS for precise measurement of their isotopic composition.

Automatically generated - may contain errors

36 protocols using conflo 4

1

Deuterium Oxide for Total Body Water

Check if the same lab product or an alternative is used in the 5 most similar protocols
At 2230 (∼30 min after the last food/water ingestion and ∼20 min after tooth brushing), subjects voided, whereafter a saliva sample was collected. Subsequently, ∼4 g (the exact weight was recorded for each subject) of deuterium oxide (D2O, 99.88%, Cambridge Isotope Laboratories, Tewksbury, MA), diluted in 70 mL of tap water, were ingested. The bottle was then rinsed with 130 mL of tap water, which was also ingested (to preserve the daily water intake, the water portion preceding the D2O ingestion was reduced by 200 mL). At 0830 on the next morning, a second saliva sample was collected. Saliva isotope enrichment relative to standard mean ocean water was determined in quadruplicate by isotope-ratio mass spectrometry (Gasbench II, Conflo IV, Delta V advantage, Thermo Scientific, Bremen, Germany). TBW was calculated as the D2O dilution space divided by 1.04 to correct for nonaqueous exchange using the formula by Schoeller et al. (29 (link)).
+ Open protocol
+ Expand
2

Carbon Isotopes in Porewater DIC Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
Pore-water for analysis of dissolved inorganic carbon (DIC) was transferred to glass vials (Zinsser) filled up to the brim (ca. 2 mL), closed without headspace and stored at 4°C until analysis. Subsamples were transferred to 10 mL helium-flushed exetainers and the DIC was transferred to the headspace as CO2 by reaction with phosphoric acid. The CO2 content of the headspace was then analyzed by gas chromatography using a GC-IRMS with helium as carrier gas (CTC Analytics GC-pal autosampler, Thermo scientific GasBench II, Thermo scientific ConFlo IV, Thermo scientific Finnigan Delta V plus IRMS). The carbon isotopic composition (δ13C) of the CO2 was determined relative to the VPDB standard using LSVEC (δ13C: -46.4‰ VPDB) and NBS 19 (δ13C: +1.95‰ VPDB) for calibration.
+ Open protocol
+ Expand
3

Isotopic Analysis of Organic Acids

Check if the same lab product or an alternative is used in the 5 most similar protocols
The isotopic signatures of formate and acetate were determined with the method of52 (link),53 (link). Briefly, the method consisted of the following steps: (1) Apolar material is removed from 5–15 mL samples by passing them over a C18 SPE cartridge; (2) Sample pH is adjusted to > 9 with NaOH and concentrated by freeze-drying; (3) Organic acids are separated by HPLC; (4) Fractions are collected in sealed 12 mL Exetainer® screw capped vials that had been previously spiked with a chemical oxidant and purged with helium; (5) oxidation of the organic acids to CO2 is achieved by heating the samples to 90 °C for 15 minutes; (6) the δ13C value of the CO2 is determined by injection into a gas chromatograph coupled to an isotope ratio mass spectrometer (Agilent 6890 gas chromatograph equipped with a CP Poraplot Q column (27.5 m x 0.32 mm, 10 µm, Varian) maintained at 100 °C, interfaced with a ConFlo IV to a Delta V Plus Mass Spectrometer, both ThermoFisher Scientific); and (7) the Δ14C value of the remaining CO2 is determined by Accelerator Mass Spectrometry (AMS) at the Institute of Particle Physics of the ETH Zürich using the gas ion source of the 200 kV using a Mini Carbon Dating System (MICADAS)54 (link).
+ Open protocol
+ Expand
4

Riesling Juice Stable Isotope Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
A 2 mL aliquot of the frozen Riesling juice (no [U-13C]-glucose added) was submitted to the Cornell Stable Isotope Laboratory (COIL) for EA-IRMS analysis. Five microliters of juice taken from a 2 mL frozen aliquot was used for bulk stable isotope analysis using a continuous flow system.26 (link) The sample was combusted at 900 °C in a Carlo Erba NC2500 elemental analyzer, and resultant CO2 was transferred via a Conflo IV interface (Thermo Scientific) to a Delta V isotope ratio mass spectrometer (Thermo Scientific) for carbon isotope analysis. Isotope ratios are reported in delta notation, as described above.
+ Open protocol
+ Expand
5

Measuring Kelp Nutrient Status via Isotopes

Check if the same lab product or an alternative is used in the 5 most similar protocols
Stable isotope ratios were measured to provide evidence of inorganic carbon uptake strategies and C and N content were measured to provide indications on the nutrient status of the kelp. Samples for determination of δ13C isotopic ratios and carbon and nitrogen content were destructively sampled and frozen at –20°C until all treatments had been measured. Following this, all samples were freeze dried (FreezeZone 4.5, Labconco) and kept at –20°C until later analysis. δ13C isotopic ratios and carbon and nitrogen content were determined by weighing approximately 5 mg of dried tissue into tin cups (Sercon, UK) and analysed using an elemental analyser (NA1500, Fisons Instruments) coupled to an isotope ratio mass spectrometer (Delta V Plus, ThermoFisher Scientific) via a Universal Continuous Flow Interface (Conflo IV, ThermoFisher Scientific). Combustion and reduction were achieved at 1020°C and 650°C, respectively. Isotope values were normalized to the Vienna Pee Dee Belemnite scale via a three-point calibration using certified reference material and both precision and accuracy were ±0.1% (1 s.d.).
+ Open protocol
+ Expand
6

Isotopic Analysis of DNA Fractions

Check if the same lab product or an alternative is used in the 5 most similar protocols
The δ13C values of DNA samples before centrifugation and the fractionated DNA of at least one sample of each triplicate were measured through isotope ratio mass spectrometry (IRMS; Delta V advantage) in conjunction with a Flash 2000 HT elemental analyzer (EA) connected to a ConFlo IV (all Thermo Fisher Scientific, USA). Two microliters of the total DNA or 5 μL of the fractionated DNA was placed in a 48-μL tin capsule and dried for 3 h at 55 °C. Then, the tin capsules were placed in an autosampler linked with an EA-IRMS (Thermo Fisher Scientific, USA). Urea was the control sample, and it was determined in triplicate before the DNA samples and after every ten samples, and the δ13C value of urea was − 40 ± 0.32‰. The corrected CO2 was adopted as a reference, and the δ13C content (‰) was determined using the following equation: δ13C()=Rsample-Rstandard/Rstandard×1000 where R = 13C/12C. Rstandard is the Vienna Pee Dee Belemnite standard (V-PDB = 0.0111802).
+ Open protocol
+ Expand
7

Feather δ²H Isotope Analysis Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
We sent feather samples for δ2H analysis at the Cornell Stable Isotope Laboratory in Ithaca, NY. After washing in a 2:1 chloroform:methanol solution, feather samples were air dried under a fume hood and a subsample (0.35 mg) of vane material was loaded into silver capsules, crushed, and placed with internal lab standards into a desiccator for a minimum of three days prior to analysis. Samples were then loaded into a Zero Blank carousel under helium flow. Pyrolysis combustion on glassy carbon was at 1350°C in a Thermo Scientific Temperature Conversion Elemental Analyzer (TC/EA; Bremen, Germany) coupled via a Conflo IV (Thermo Scientific) to a Thermo Scientific Delta V Advantage isotope ratio mass spectrometer. Analysis of δ2H was conducted using the comparative equilibration method of Wassenaar and Hobson [52 (link)] with 2 calibrated keratin reference materials (CBS, δ2H = -197‰; KHS, δ2H = -54.1‰; SPK, δ2H = -121.6‰) corrected for linear instrumental drift. Based on within-run analyses of a third keratin standard, measurement error was approximately ± 3‰ for hydrogen isotopes in feather (δ2Hf). All δ2H values are reported relative to the Vienna Standard Mean Ocean Water–Standard Light Antarctic Precipitation scale.
+ Open protocol
+ Expand
8

Automated Isotopic Analysis of Amino Acids

Check if the same lab product or an alternative is used in the 5 most similar protocols
Automated sampling of amino acid derivatives was conducted using a Trace 1310 GC (Thermo Scientific) in combination with a TriPlus RSH autosampler (Thermo Scientific). A sample volume of 0.1 μl was injected into an injection port held at 240°C in splitless mode onto a capillary column (DB‐35, 30 m × 0.320 mm ID × 0.50 μm film thickness, Agilent J&W GC Columns). The GC oven temperature program for a single injection is detailed in Appendix S2.
All eluting compounds off the column were oxidized inside a GC Isolink II (Thermo Scientific) combustion interface containing a reactor held at 1,000°C. Water was removed through a Nafion membrane downstream of the reactor while CO2 was cryogenically trapped in tubing submerged in liquid nitrogen before transfer to the IRMS (DELTA V; Thermo Scientific) through a Conflo IV (Thermo Scientific) universal interface.
High purity N2 (>99.9997% N2, Airgas) was used as reference gas to initially calculate the isotopic composition. Raw data were drift‐corrected (for drift correction procedure see Appendix S3). Average precision across all amino acids was 0.33‰. Individual precision of each amino acid is given in Appendix S4.
+ Open protocol
+ Expand
9

Stable Isotope Analysis of Pteropods and POM

Check if the same lab product or an alternative is used in the 5 most similar protocols
Bulk carbon and nitrogen stable isotope ratios were obtained using an automated Elementar vario PYRO cube analyser interfaced with a continuous flow IsoPrime100 isotope ratio mass spectrometer for pteropods, and a Thermo Scientific Flash 2000HT analyser coupled with a Thermo Fisher Delta V Plus mass spectrometer through a Thermo Fisher Conflo IV for POM. For pteropods, SIA was performed at the Central Science Laboratory (CSL), University of Tasmania (Sandy Bay, Tasmania), and for POM, at the Australian Nuclear Science and Technology Organisation (ANSTO, Lucas Heights, Sydney). Isotopic ratios were expressed in delta (δ) notation and reported as parts per thousand (‰) deviations from conventional certified isotopic reference standards, Vienna Pee Dee Belemnite (for carbon) and atmospheric air (for nitrogen) (DeNiro & Epstein, 1978). At CSL, laboratory working standards of sulfanilamide were repeated every 6th sample for both isotopes to measure instrument stability and precision. Average standard deviations on triplicate measurements were 0.15‰ and 0.19‰ for δ13C and δ15N values of pteropods, respectively. At ANSTO, POM isotopic data are reported relative to IAEA secondary certified standards with a standard error of analysis to 1 standard deviation (SD) measured at ±0.3 mil. The carbon and nitrogen percentages, by weight, were converted to atomic C:N ratios.
+ Open protocol
+ Expand
10

Plasma Protein Nitrogen Isotope Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
Plasma samples were thawed at 4 °C overnight and their protein fraction was isolated by precipitation with sulfosalicylic acid (200 µL of 1 g/mL into 2 mL of sample). After 1 h of incubation at 4 °C and centrifugation (4500× g for 20 min at 4 °C), the supernatant and pellet were separated [19 (link)]. The pellet was rinsed three times with MilliQ water and then freeze-dried. The N stable isotopic composition (δ15N, i.e., natural relative abundance of the rare stable isotope of N) of plasma protein was determined using a Carlo Erba NA1500 elemental analyzer coupled to a Delta V plus isotope-ratio mass spectrometer via a Conflo IV (Thermo Scientific, Waltham, MA, USA). Stable isotope results were corrected via a three-point calibration using international primary reference standards. A secondary reference was included after every 12 samples to monitor any instrument drift. Nitrogen was quantified by comparison of peak areas against a response calibration curve.
Results are expressed using the delta notation according to the following equation:
where Rsample and Rstandard are the N isotope ratio between the heavier isotope and the lighter isotope (15N:14N) for the sample being analyzed and the internationally defined standard (atmospheric N2, Rstandard = 0.0036765), respectively, and δ is the delta notation in parts per 1000 (‰) relative to the standard.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!