The largest database of trusted experimental protocols

Zorbax rrhd eclipse plus c18 column

Manufactured by Agilent Technologies
Sourced in United States, Germany

The ZORBAX RRHD Eclipse Plus C18 column is a reversed-phase high-performance liquid chromatography (HPLC) column. It is designed for the separation and analysis of a wide range of compounds in various applications. The column features a C18 stationary phase and is suitable for use in rapid resolution high-definition (RRHD) HPLC systems.

Automatically generated - may contain errors

45 protocols using zorbax rrhd eclipse plus c18 column

1

Quantitative Analysis of Pseudomonas Metabolites

Check if the same lab product or an alternative is used in the 5 most similar protocols
This experiment was also performed by LC/Q-ToF (Agilent) using ZORBAX RRHD Eclipse Plus C18 column (Agilent) in MS mode, with 1 μL injection of experimental samples and standards for pyochelin, phenazine-1-carboxylic acid, and rhamnolipids. The mobile phase was water (A) and acetonitrile (B), both with 0.1% (v/v) formic acid. The elution procedure was 5% B between 0 and 1 min; 5−95% B between 1 and 11 min; 95% B between 11 and 13 min; 95−5% B between 13 and 14 min; and 95% B between 14 and 16 min with a flow rate of 0.4 mL/min.
The standards were dissolved in methanol to make stock solutions at a concentration of 1 g/L for pyochelin, 0.1 g/L for phenazine-1-carboxylic acid, and 10 g/L for rhamnolipids. The stock solutions were diluted to appropriate concentrations for sample preparation.
+ Open protocol
+ Expand
2

Quantitative Analysis of Infant Fecal Monosaccharides

Check if the same lab product or an alternative is used in the 5 most similar protocols
The monosaccharides analyzed in this method are those that are unconjugated and not constituents of di-, oligo-, or polysaccharides. The separation of monosaccharides used a binary solvent system, solvent A contained 5% acetonitrile (ACN) and 25 mM ammonium acetate (NH4Ac) pH adjusted with ammonium hydroxide (NH4OH) to 8.2. Solvent B consisted of 95% ACN. A flow rate of 0.500 mL/min was used with the following gradient: 0–7 min, 12–15% B; 7–7.10 min, 15–99% B; 7.10–8.50 min, 99% B; 8.50–8.60 min, 99–12% B. Quantitation of monosaccharides in infant feces was conducted using a calibration curve constructed with a range from 1 µg to 100 mg monosaccharide per mg solution. The column used for separation was an Agilent ZORBAX RRHD ECLIPSE PLUS C18 column (150 × 2.1 mm) with 1.8 µm particle size. The UHPLC-QqQ system was an Agilent 1290 Infinity II UHPLC coupled to an Agilent 6495B QqQ mass spectrometer and 1 µL of each sample was injected for a run time of 10 min. Further details on instrument parameters and monosaccharide standards have been described by Xu et al. (26 ).
+ Open protocol
+ Expand
3

Broccoli Phytosterol Extraction and HPLC Quantification

Check if the same lab product or an alternative is used in the 5 most similar protocols
Ground broccoli head samples were placed into an Eppendorf tube (50 mL) containing 30 mL of ethanol. After vortexing and shaking for 3 min, the samples were ultrasonicated for 60 min using an ultrasonic cleaner (Xianou Instrument Co., Nanjing, China) and then centrifuged at 8000× g for 10 min when the samples cooled down. The supernatant passed through a 0.22 μm PVDF membrane.
The phytosterol was analyzed by an Agilent 1260 Series HPLC system (Agilent Technologies, Wilmington, DE, USA). The analytes were separated by an Agilent ZORBAX RRHD Eclipse Plus C18 column (50 × 3 mm, 1.8 μm) held at 30 °C. The mobile phase ratio of methanol/water (90:10% v/v) maintains a flow rate of 0.5 mL/min with a temperature of 350 °C. The injection volume was 10 μL. The identification and quantification parameters of mass spectrometry are shown in Table S2. The mass spectrum detector was operated in positive electrospray ionization (ESI).
+ Open protocol
+ Expand
4

Metabolite Profiling by HPLC-QTOF MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
Metabolites were identified and quantified by HPLC/Q-TOF (Agilent 1260 Infinity II/Agilent G6545B) in MS mode using positive ionization. For analysis, 1 μL of each culture sample was injected and separated in the ZORBAX RRHD Eclipse Plus C18 column (2.1 × 50 mm, 1.8 μm) (Agilent) with water with 0.1% formic acid (A) and acetonitrile with 0.1% formic acid (B) as the mobile phase. The gradient program was set as 0–1 min, 95% A; 1–11 min, 95%–5% A; 11–13 min, 5% A; 13–14 min, 5%–95% A; and 14–16 min, 95% A at a flow rate of 0.4 mL/min. The m/z value of the [M+H]+ adduct was then used to extract the ion chromatogram (with a mass error below 20 ppm) for compound identification and quantification (see Supplementary Fig. S1 for extracted ion chromatograms of DDK, DMY, and Y standard). For compounds with available chemical standards, including cinnamic acid, hydrocinnamic acid, p-coumaric acid, p-hydrocoumaric acid, DDK, DMY, and Y, concentrations were quantified by comparing the integrated peak areas to standard curves (Supplementary Fig. S2 and Fig. S3). For identification of compounds of which no standards are available, target MS/MS mode with three different levels of collision energy (10, 20 and 40 eV) was used in addition to high-resolution MS analysis.
+ Open protocol
+ Expand
5

Purification and Analysis of Disaccharide-Depsipentapeptide from L. plantarum

Check if the same lab product or an alternative is used in the 5 most similar protocols
Disaccharide-depsipentapeptide (GlcNAc-MurNAc-L-Ala-D-Gln-mDAP-L-Ala-L-Lac) was purified from PG of dacA1dacA2 double mutant of L. plantarumNC8. PG was digested with mutanolysin (Sigma-Aldrich), and the resulting soluble muropeptides were reduced by NaBH4 as described previously (Bernard et al., 2011 (link)). The reduced muropeptides were separated by reverse-phase ultra-high-pressure liquid chromatography (RP-UHPLC) with a 1290 chromatography system (Agilent Technologies) and a Zorbax RRHD Eclipse Plus C18 column (100 by 2.1 mm; particle size, 1.8 µm; Agilent Technologies) at 50°C using ammonium phosphate buffer and methanol linear gradient. The eluted muropeptides were detected by UV absorbance at 202 nm. The peak corresponding to the disaccharide-depsipentapeptide was collected. Purified muropeptide was then incubated with 10 µg of purified DltEextra or DacA1 in 50 mM Tris-HCl, 100 mM NaCl buffer overnight at 37°C. The reaction mixtures were analyzed by RP-UHPLC as described above and by LC-MS with an UHPLC instrument (Vanquish Flex, Thermo Scientific) connected to a Q Exactive Focus mass spectrometer (Thermo Scientific). Mass spectra were collected over the range m/z = 380–1400. Data were processed using Xcalibur QualBrowser v2.0 (Thermo Scientific).
+ Open protocol
+ Expand
6

UPLC-QTOF-MS Analysis of Samples

Check if the same lab product or an alternative is used in the 5 most similar protocols
An Agilent 1290 Infinity II UPLC system connected to a G6545B Q-TOF MS system with a dual ESI source (Agilent Technologies, USA) was used to assess the samples. Using 0.1% formic acid-deionized water (A) and acetonitrile (B), all samples went through separation on an Agilent ZORBAX RRHD Eclipse Plus C18 column (50 × 2.1 mm, 1.8 μm) linked to an Agilent A-Line Quick Connect LC Fitting. Temperature: 20 °C; injection volume: 1 μL; data rate: 10 Hz; flow rate: 0.3 mL/min; split ratio: 1:1; wavelength: 210, 254, and 315 nm were the parameters of the UPLC. 0–0.3 min, 10% B; 0.3–10 min, 10%–100% B; 10–12 min, 100% B; 12–12.1 min, 100%–10% B; 12.1–15 min, 10% B, was the optimal gradient elution program.
+ Open protocol
+ Expand
7

Quantitative Analysis of PFOA and PFBA

Check if the same lab product or an alternative is used in the 5 most similar protocols
The PFBA samples were analyzed at wavelength of 262 nm using ultraviolet−visible spectrophotometry (Unico, model 2008A). PFOA was analyzed directly via LC-MS/MS by two methods. First, The PFOA samples without SDS employed a Thermo Ultimate 3000 HPLC (Thermo Fisher Scientific, Breme, Germany) and AB-Sciex Qtrap 4500 mass spectrometer (AB Sciex Pte. Ltd., Singapore). A Waters column (C18, 2.1 × 50 mm, 1.7 μm) was maintained at 35 °C. The mobile phase of water and 100% acetonitrile applied in the ratio of 40:60 was set at a flow rate of 0.25 mL/min. Second, PFOA samples with SDS employed an Agilent HPLC 1260 and mass spectrometer 6150. An Agilent ZORBAX RRHD Eclipse Plus C18 column (2.1 × 50 mm, 1.8 μm) was maintained at 40 °C. The mobile phase of 0.01M ammonium acetate and 100% acetonitrile applied in the ratio of 60:40 was set at a flow rate of 0.3 mL/min. All the samples were passed through a 0.45 μm filter to remove suspended particulate impurities. Mass-labeled internal standard 13C4-PFOA was added to monitor recovery. The R2 of the calibration curves were >0.999 for each measurement. The quantifiable detection limit of the two systems is ~ 0.1 μg/L. These methods have been used successfully in our prior studies.26 (link),33 (link),34 (link)
+ Open protocol
+ Expand
8

Oxylipin Analysis by UPLC-MS/MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
Oxylipins were analyzed on an Agilent 1290 Infinity UPLC system coupled to an Agilent 6460 triple-quadrupole tandem mass spectrometer (UPLC-MS/MS) with electrospray ionization (ESI) (Agilent, Palo Alto, CA, USA). Analyte separation was achieved using an Agilent ZORBAX RRHD Eclipse Plus C18 column (2.1 × 150 mm, 1.8 μm Agilent Corporation). Mobile phase A consisted of MilliQ water containing 0.1% acetic acid, and mobile phase B consisted acetonitrile/methanol (v/v, 80/15) containing 0.1% acetic acid. Samples were maintained at 4 °C during the analysis. The column temperature was 45 °C. The injection volume was 10 μL and the run time was 20 min. The mobile phase gradient and flow rate were set as follows: 1) 0–2 min, 65% A, 0.25 mL/min; 2) 2–12 min, 65–15% A, 0.25 mL/min; 3) 12–15 min, 15% A, 0.25 mL/min; 4) 15.1–17 min, 0% A, 0.4 mL/min; 5) 17.1–19 min, 65% A, 0.4 mL/min; and 6) 19–20 min, 65% A, 0.3 mL/min. The MS ion source parameters were set as follows: gas temperature 300 °C, gas flow 10 L/min, nebulizer gas 35 psi, sheath gas heater temperature 350 °C, and capillary voltage 4000 V. Oxylipins were captured using optimized dynamic multiple reaction monitoring parameters shown in Supplementary Table 1.
+ Open protocol
+ Expand
9

Ultra-sensitive Glycopeptide Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
An ultra performance liquid chromatography system Nexera UPLC LC-30A (Shimadzu Corporation, Kyoto, Japan) equipped with a ZORBAX RRHD Eclipse Plus C18 column (1.8 μm, 2.1 mm × 100 mm, Agilent Technologies, Santa Clara, CA) was coupled online to a 6500 plus Qtrap Mass Spectrometer (AB Sciex, CA, USA). Mobile phase A was 0.1% FA and 3% acetonitrile in nanopure water (v/v/v), and mobile phase B was 0.1% FA in 90% acetonitrile. Peptides and glycopeptides were separated by a binary gradient at 40°C with 0.5 mL/min flow rate consisting of (1) 0 min at 2.0% B; 0.5-1.0 min at 5.0% B; 6.0 min at 30.0% B; (2) 6.1-8.0 min at 100.0% B), (3) 8.1-10.0 min at 2.0% B. The MS was operated in the positive mode. Curtain gas: 30.0 psi, collision gas: high, ion spray voltage: 5500.0 V, temperature: 300.0°C, ion source gas 1: 55.0 psi, ion source gas 2: 60.0 psi, declustering potential: 20.0 V, entrance potential: 10.0 V, and collision cell exit potential: 14 V. The scheduled MRM mode was used. Q1 and Q3 were set to unit resolution. The cycle time was fixed to 500 ms, the MRM detection window was set to 30.0 s, and the target scan time was 0.5 s. CE for each MRM transition was optimized by a 5 V step followed by a 2 V step fine-tuning. The analysis was controlled using the AB Sciex Analyst software (version 1.6.3).
+ Open protocol
+ Expand
10

UHPLC-MS/MS Analysis of Substrates and Metabolites

Check if the same lab product or an alternative is used in the 5 most similar protocols
Chromatographic separations were carried out on an Agilent 1290 UHPLC system (Agilent Technologies, Santa Clara, CA, USA) using an Agilent ZORBAX RRHD Eclipse Plus C18 column (2.1 × 50 mm inner diameter,1.8 μm particle size) (Agilent Technologies, Santa Clara, CA, USA). The elution profile was composed of an initial isocratic step with water (0.1% formic acid): acetonitrile (ACN) (95:5) for 1.5 min and then increasing ACN to 30% over 4 min to separate the substrates and metabolites at 30 °C. The column was washed with 90% of ACN for 2.5 min followed by the equilibration of the column for 5 min with 5% ACN. The flow rate was 0.4 mL/min and the injection volume was 5 μL. An Agilent 6460 triple quadrupole-mass spectrometer with an electrospray ion source (Agilent Technologies, Santa Clara, CA, USA) was used for flow injection analysis with optimized fragmentor and source parameters. The optimized source parameters for MS analysis were drying gas temperature, 350 °C; gas flow, 10 L/min; nebulizer gas flow pressure, 35 psi; and capillary voltage, 4500 V.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!