The largest database of trusted experimental protocols

Thermo syncronis c18 column

Manufactured by Thermo Fisher Scientific
Sourced in United States

The Thermo Syncronis C18 column is a high-performance liquid chromatography (HPLC) column designed for the separation and analysis of a wide range of compounds. The column features a C18 stationary phase, which is a common choice for reversed-phase chromatography, and is suitable for use in various analytical applications.

Automatically generated - may contain errors

8 protocols using thermo syncronis c18 column

1

Methylglyoxal Trapping Assay Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
The methylglyoxal trapping assay followed a published method with some modifications [5 (link),8 (link)]. Methylglyoxal (10 mM), o-phenylenediamine (50 mM) were freshly prepared in phosphate buffer (50 mM, pH 7.4). The final concentration of test samples (the five fractions, compounds 13 and kaempferol) and aminoguanidine was 5 mg/mL. Three mL methylglyoxal were incubated with 0.75 mL test samples or phosphate buffer in pH 7.4 at 37 °C water bath for 0, 0.25, 0.5, 0.75, 1, 1.5, 2, 3, 4, 6, 8, 12 and 24 h. After incubation, 200 μL of o-phenylenediamine were added with 200 μL of each test solution. Before HPLC analysis, the mixtures were kept in dark at room temperature for 1 h. The remaining methylglyoxal was detected on an Thermo Dionex Ultimate 3000 HPLC system (Thermo Fisher Scientific Inc., Massachusetts, USA) equipped with a diode array detector, and a Thermo Syncronis-C18 column (250 mm × 4.6 mm i.d., 5 μm; Thermo Fisher Scientific Inc., Massachusetts, USA). Mobile phases were composed of 0.1% formic acid in water (A) and methanol (B). The elution started at 50% B and then increased to 55% in 10 min, and then increased to 100% in 11 min and lasted for 5 min, then returned to 50% in 16.5 min and lasted for 5 min. The flow rate was 1 mL/min and the injection volume was 20 μL. Detection wavelength was monitored at 315 nm.
+ Open protocol
+ Expand
2

High-Performance Liquid Chromatography Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
Liquid Chromatography (LC) was conducted using the Dionex UltiMate 3000 Ultra High-performance Liquid Chromatography system (Thermo Fisher Scientific, Massachusetts, USA) equipped with a Thermo Syncronis C18 column (2.1 mm × 100 mm × 1.7 µm) (Thermo Fisher Scientific). During the analysis, the temperature was retained at 55 °C at a flow rate of 450 µL/min. Solvent A (water mixed with 0.1% formic acid) and solvent B (acetonitrile mixed with 0.1% of formic acid) were used as the mobile phases. Beginning at 0.5% of solvent B, the gradient elution program was set for 1 min, followed by 15 min from 0.5% to 99.5% of solvent B before being maintained for 4 min, and returned to 0.5% of solvent B in 2 min. The injection volume was set at 2 µL.
+ Open protocol
+ Expand
3

Extraction and Analysis of Green Tea Catechins

Check if the same lab product or an alternative is used in the 5 most similar protocols
Fresh green tea (Camellia sinensis) leaves were collected in the spring from the Osulloc Tea Garden in Jeju, Korea and dried. Dried C. sinensis leaves were subjected to extraction twice with 50% aqueous ethanol and incubation at 100°C (1.2 atm) under aqueous conditions for 5 h to obtain GCG-GTE. The catechin contents in GCG-GTE were measured and analyzed using high-performance liquid chromatography (HPLC) with a photodiode array (PDA) detector (Alliance 2695 system, Waters) and a Thermo Syncronis C18 column (250 × 4.6 mm, I.D., 5 μm; Thermo Fisher Scientific Inc.) GCG-GTE contained almost the same amount of epicatechins and epicatechin epimers (Table 1).
+ Open protocol
+ Expand
4

HPLC-PDA Analysis of HTP-GTE

Check if the same lab product or an alternative is used in the 5 most similar protocols
HTP-GTE was analyzed by HPLC-PDA (Alliance 2695 system, Waters, USA) using a ThermoSyncronis C18 column (250 × 4.6 mm, I.D., 5 μm; Thermo Fisher Scientific Inc.). The mobile phases were 0.1% acetic acid in water for solvent A and acetonitrile for solvent B. The gradient elution was 90% A + 10% B at 0–10 min, 85% A + 15% B at 10–30 min, 80% A + 20% B at 30–53 min, 5% A + 95% B at 53–55 min, 90% A + 10% B at 55–60 min with a flow rate of 1.0 mL/min. The sample volume for injection was 10 μl, and the measurement was carried at the UV wavelength of 280 nm.
+ Open protocol
+ Expand
5

UPLC-MS/MS Quantification of R-/S-HFBA

Check if the same lab product or an alternative is used in the 5 most similar protocols
An ACQUITY UPLC™ liquid chromatography system equipped with a quaternary pump, automatic injector, oven and a Xevo TQS MS/MS system (Waters Corporation, Milford, MA, USA) was used for excretion analyses. Chromatographic separation was performed on a Thermo Syncronis C18 column (50 mm × 2.1 mm, 1.7 μm; Thermo, USA). The analytes were eluted using a gradient method at a flow rate of 0.4 mL min−1. Mobile phase A was 0.1% formic acid in water and mobile phase B was acetonitrile according to the following gradient: 0–1.5 min, 15–46% B; 1.5–2.5 min, 46% B; 2.5–2.6 min, 46–15% B; 2.6–3.5 min, 15% B. The injection volume was 2 μL. The column and automatic injector temperature were set at 35 °C and 4 °C, respectively.
The mass spectrometer was operated in positive electrospray mode using multiple reaction monitoring (MRM). The desolvation temperature and source temperature were 350 °C and 150 °C, respectively. The capillary voltage was 3.0 kV. The desolvation gas flow rate and the cone gas flow rate were 700 L h−1 and 150 L h−1, respectively. The ion transitions (m/z) were 278.0 > 187.9 for R-/S-HFBA, 151.9 > 109.9 for IS. The collision energy were 18 eV for R-/S-HFBA and 14 eV for IS, respectively. The cone voltage were 21 V for R-/S-HFBA and 22 V for IS, respectively. The product ion spectra for R-/S-HFBA and IS were shown in our previous report.23 (link)
+ Open protocol
+ Expand
6

UHPLC-Based Analytical Method for Identification

Check if the same lab product or an alternative is used in the 5 most similar protocols
Liquid chromatography (LC) was performed using a Dionex UltiMate 3000 UHPLC system (Thermo Fisher Scientific, Waltham, MA, USA) coupled to a Thermo Syncronis C18 column (2.1 mm × 100 mm × 1.7 μm; Thermo Fisher Scientific, Waltham, MA, USA), which was maintained at 55 °C and a flow rate of 450 µL/min during analysis. The mobile phases consisted of solvent A (water added with 0.1% formic acid) and solvent B (acetonitrile added with 0.1% formic acid). The gradient elution program was started at 0.5% of solvent B for 1 min, then from 0.5% to 99.5% of solvent B for 15 min and held for 4 min. The columns were then conditioned for 2 min before the next injection. The injection volume was set to 2 µL.
+ Open protocol
+ Expand
7

HPLC Analysis of Antifungal Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
The active fraction collected from SPE separation was analyzed further by a model of Agilent 1100 series high-performance liquid chromatography (Agilent Technologies). Compounds were separated in a Thermo Syncronis C-18 column (5 μm, 250 × 4.6 mm, Thermo Fisher Scientific) using a gradient solvent system composed of water-methanol at a flow rate of 1 mL/min as the mobile phase. The concentrations of methanol were programmed as follows: 5–42% at 0–5 min, 42–44% at 5–10 min, 44–45% at 10–25 min, 45–95% at 25–30 min, and hold in 95% at 30–40 min. Compounds were detected by a UV detector with an optical absorbance at 254 nm. Each peak was collected, dried, and tested for antifungal activity.
+ Open protocol
+ Expand
8

UHPLC Method for Compound Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
Liquid chromatography (LC) was performed using the Dionex UltiMate 3000 UHPLC system (Thermo Fisher Scientific, Waltham, MA, USA) coupled with a Thermo Syncronis C18 column (2.1 mm × 100 mm × 1.7 μm; Thermo Fisher Scientific, Waltham, MA, USA), which was maintained at 55 °C at a flow rate of 450 µL/min during analysis. The mobile phases were composed of solvent A (water added with 0.1% formic acid) and solvent B (acetonitrile added with 0.1% of formic acid). The gradient elution program was initiated at 0.5% of solvent B for 1 min, then from 0.5% to 99.5% of solvent B for 15 min and maintained for 4 min. Later, the columns were conditioned as initial for 2 min before the next injection. The injection volume was set at 2 µL.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!