The axon was followed from the soma into stratum lucidum using frame mode scanning (256 × 256 pixels, 500 Hz; the number of taken frames was kept small to minimize phototoxic damage) and the system was focused on a giant MFB identified by its distinct morphology, at the maximum optical resolution (~0.2 μm, digital capture 70 nm per pixel). Recording started when the baseline fluorescence in both channels was stable (approximately one hour later; see Results and
Fluorescence responses were recorded in line-scan mode at 500 Hz (500 or 1000 ms sweeps, inter-sweep interval 30 s or 1 min) and stored for off-line analysis. The Ca2+-dependent fluorescence response ΔF/F (integrated over the visible MFB width) was routinely calculated as (Fpost-Fpre)/(Fpre-F0). The values of Fpre and Fpost stand for the line scan fluorescence averaged over, respectively, 100 ms prior to the first spike and either 50 ms in the case of single-response amplitude measurements or 250 ms in the case of five-response amplitude measurements (20 Hz train of five APs) after the first spike onset. F0 denotes the background fluorescence measured outside any cell structures filled with the indicator. Because special care was taken to avoid escape of the indicator from the pipette and because the site of imaging was hundreds of microns away from the pipette tip, F0 was likely to represent the photomultiplier tube dark current. Image analyses were performed on stacks of stored line-scan images using a set of custom NIH Image macros. False color tables and averaged images were used for illustration purposes but the original (gray level) pixel brightness values in each line-scan image were used for the quantitative analysis. In most experiments, we reconstructed the axon trajectory using a collage of high-resolution Kalman-filtered z-stacks 15-20 μm deep. In total, we obtained full reconstruction of 43 axons, with an average distance between the recorded MFB and the soma of 686 ± 38 μm. Throughout the experiments, we observed no failures of spike-driven Ca2+ signals propagating along the main axonal trunk including giant MFBs. This, however, does not rule the possibility that propagation could fail at higher spiking frequencies and/or in thin axon collaterals.
The two-photon excitation probability profile is proportional to the squared illumination light intensity (Zipfel and Webb, 2001 ):
where the canonical co-ordinates and v = ksin(α)r represent the axial distance z, radial distance r, the objective’s numerical aperture NA = sin(α) = 0.9 and wave number (n = 1.33 is the medium refraction index and λ = 810 nm is the wavelength); J0 denotes zero-order Bessel’s function of the first kind. This theoretical function, however, represents the lower limit estimate of the excitation profile: in reality, optical aberrations and imperfect alignment of the experimental optical system are likely to increase the spread of excitation. Similar considerations apply to the emission path. We therefore obtained an estimate of the excitation-emission profile by recording the point-spread function (PSF) of the system using 0.17 μm fluorescent beads (PS-Speck Microscope Point Source Kit, Invitrogen) as illustrated below.