Because of the high chromatographic resolution, OH-PCBs are typically extracted from biological or environmental matrices, derivatized to MeO-PCBs, and analyzed by GC-MS/MS.4 (link), 14 (link), 15 (link), 58 (link) We measured the RRTs and MS/MS profiles [expressed as the relative intensities of five multiple reaction monitering (MRM) transitions] of two MeO-PCB standard solutions (Solution 1 containing 72 MeO-PCBs and Solution 2 containing 52 MeO-PCBs; see Supporting Information for additional information) using an Agilent 7890B gas chromatograph equipped with an SPB-Octyl capillary column (30 m length, 250 μm inner diameter, 0.25 μm film thickness; Supelco, Bellefonte, PA, USA), an Agilent 7000D Triple Quad and an Agilent 7693 sampler. These data were used as dependent variables for the model development. For additional details, see the Supporting Information.