GC-MS Analysis of Metabolites
Corresponding Organization :
Other organizations : Hubei University of Chinese Medicine, Ministry of Education of the People's Republic of China, Xinjiang Uygur Autonomous Region Uygur Medicine Hospital
Variable analysis
- Injection volume (1 µL)
- Mass spectrum (m/z 50.0–600.0)
- Carrier gas (helium, 99.999%)
- Flow rate (1.0 ml/min)
- Split ratio (10:1)
- Column temperature (60°C for 1 min, then increased at 10°C/min up to 250°C and maintained at 250°C for 10 min)
- Injection temperature (250°C)
- Electron bombardment ion source temperature (230°C)
- Solvent delay (5.9 min)
Annotations
Based on most similar protocols
As authors may omit details in methods from publication, our AI will look for missing critical information across the 5 most similar protocols.
About PubCompare
Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.
We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.
However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.