The imaging system selected was a research whole-body PCD-CT scanner (Somatom CounT, Siemens Healthcare, Forchheim, Germany) with a cadmium telluride semiconductor5 (link),6 (link). The UHR mode was chosen for all scans performed5 (link). The iodinated intravascular contrast agent used in this study was Omnipaque® 350 (GE Healthcare, Inc.). A vascular CT imaging protocol was used for all scans reported in this study. The scans were performed using the following parameters: 140 kV, 341 mAs, and energy threshold settings of 30 and 70 keV. Images were reconstructed using an 80 mm × 80 mm FOV, 1024×1024 matrix size, 1 mm slice thickness (yielding voxel size of 0.0781mm × 0.0781mm × 1 mm), and an iterative reconstruction algorithm with a sharp quantitative kernel (Q65f, SAFIRE).