Multimodal MRI Acquisition Protocol
Corresponding Organization : Kessler Foundation
Other organizations : Rutgers, The State University of New Jersey, New Jersey Institute of Technology, Northeastern University
Variable analysis
- Magnetic field strength (3-T)
- Signal intensity in T1-weighted images
- Signal intensity in T2-weighted images
- Siemens Allegra head-only scanner
- Siemens standard head coil
- Magnetization-Prepared Rapid Gradient-Echo sequence parameters (TR = 2 s, TE = 4.38 ms, voxel size 0.938 × 0.938 × 1, 176 slices, and 1 mm slice thickness)
- T2-weighted image sequence parameters (TR = 2s, TE = 30 ms, FOV 220 mm × 220 mm, voxel size = 3.4 mm × 3.4 mm × 4 mm, 32 slices, and 62 ms inter-slice time)