Lipid samples were analyzed at the Harvard Center for Mass Spectrometry. The LC–MS analyses were modified from Miraldi81 (link) and were performed on an Orbitrap Exactive plus (Thermo Scientific) in line with an Ultimate 3000 LC (Thermo Scientific). Each sample was analyzed in positive and negative modes, in top 5 automatic data-dependent MSMS mode. Column hardware consisted of a Biobond C4 column (4.6 × 50 mm, 5 μm, Dikma Technologies). Flow rate was set 100 μl min−1 for 5 min with 0% mobile phase B (MB), then switch to 400 μl min−1 for 50 min, with a linear gradient of MB from 20 to 100%. The column was then washed at 500 μl min−1 for 8 min at 100% MB before being reequilibrated for 7 min at 0% MB and 500 μl min−1. For positive mode runs, buffers consisted for mobile phase A (MA) of 5 mM ammonium formate, 0.1% formic acid and 5% methanol in water, and for mobile phase B (MB) of 5 mM ammonium formate, 0.1% formic acid, 5% water, 35% methanol in isopropanol. For negative runs, buffers consisted for MA of 0.03% ammonium hydroxide, 5% methanol in water, and for MB of 0.03% ammonium hydroxide, 5% water, 35% methanol in isopropanol. Lipids were identified and quantified using the Lipidsearch© software (version 4.2.27, Mitsui Knowledge Industry, University of Tokyo). Integrations and peak quality were curated manually before exporting and analyzing the data in Microsoft Excel.