For sample-specific details, refer to Table S6.
Movies were collected in compressed tiff format on a Titan Krios G2 (Thermo Fisher) operating at 300 kV with a K3 detector (Gatan) in super resolution counting mode using a custom version of EPU 2.5 (Thermo Fisher). A defocus range of 0.8-2.6 μm was applied with a nominal magnification of x105,000, corresponding to a calibrated pixel size of 0.83 Å/pixel and with a total dose of 43-47 e/ Å2, see Table S6.
Two-times binned movies were then motion corrected and aligned on the fly using Relion(3.1) scheduler (Zivanov et al., 2018 (link)) with a 5 × 5 patch based alignment. CTF-estimation of full-frame non-weighted micrographs was performed with the GCTF (1.06) (Zhang, 2016 (link)) module in cryoSPARC(v2.14.1-live) (Punjani et al., 2017 (link)).