RP-LC-MS/MS for Ceramide Profiling
Corresponding Organization : Medical University of Białystok
Other organizations : University of Aveiro
Variable analysis
- Chromatographic separation of CERs was performed on RP-C18 column (Acquity BEH Shield 2.1 × 100 mm; 1.7 μm; Waters, Milford, MA, USA) in gradient elution with the mixture of water (20 mM ammonium formate, pH 5) and methanol.
- The RP-LC-MS/MS was applied to obtain the CER profiles.
- The same Agilent UPLC-ESI-QTOF-MS system (Agilent 1290; Agilent 6540; Agilent Technologies, Santa Clara, CA, USA) was applied for analysis.
- The QTOF mass spectrometer was operated in positive (electrospray voltage 3.5 kV) ion mode with a capillary temperature of 300 °C and a sheath gas flow of 8 L/min, as previously described in details [27 (link)].
- No positive or negative controls were explicitly mentioned in the provided information.
Annotations
Based on most similar protocols
As authors may omit details in methods from publication, our AI will look for missing critical information across the 5 most similar protocols.
About PubCompare
Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.
We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.
However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.
Ready to get started?
Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required
Revolutionizing how scientists
search and build protocols!