Acquisition and processing of [18F]flortaucipir PET images is described in detail elsewhere [17 (link), 25 , 36 (link)]. In short, dynamic 130-min [18F]flortaucipir PET scans were acquired on a Philips Ingenuity TF-64 PET/CT scanner. The scanning protocol consisted of two dynamic PET scans of 60 and 50 min, respectively, with an in-between 20-min break [36 (link)]. The first 60-min dynamic scan started simultaneously with a bolus injection of approximately 234 ± 14 MBq [18F]flortaucipir (injected mass 1.2 ± 0.9 μg). The second PET scan was co-registered to the first dynamic PET scan using Vinci software (Max Planck Institute, Germany). PET data were acquired in list mode and subsequently reconstructed using 3D-RAMLA including standard corrections for dead time, decay, attenuation, random, and scatter. Patients also underwent 3DT1-weighted MRI (Ingenuity TF PET/MR, Philips Medical Systems, The Netherlands) for anatomical and tissue segmentation purposes.