MR examinations were performed with a 3-T imager (GE Healthcare, Milwaukee, WI). The PDFF was measured using the Dixon method with advanced processing, called IDEAL-IQ (Iterative decomposition of water and fat with echo asymmetry and the least squares estimation quantification sequence, GE Healthcare). The Dixon method using 6 echoes was employed to measure the fat fraction using the following imaging parameters: repetition time/echo space: 6.6 ms/0.78 ms; maximum TE 4.8; minimum TE 0.9; NEX 0.50; location per slab 28; number of shots 2; echo train length 3; flip angle: 3 degrees; matrix: 160 x 160; axial imaging plane; section thickness: 8 mm; FOV: 44 cm; phase FOV: 0.8 cm; bandwidth: 111.11 kHz; imaging time, single breath-hold (19 seconds). After setting the ROIs in the organs of interest, the PDFF calculation was performed by a hybrid multipoint water-fat separation approach, where a T2*-corrected, spectrally modeled ‘‘complex based” water-fat separation method (T2*-IDEAL) is followed by a fitting algorithm based on magnitude multi-echo signals as reported previously [13 (link)].